Eunah Chung1, Dongyeol Kwon1, Taeyang Park1, Hyeri Kang1, Yoonsun Chung1,2.
Abstract
PURPOSE: The delivery quality assurance (DQA) of intensity-modulated radiotherapy (IMRT) plans is a prerequisite for ensuring patient treatments. This work investigated the clinical usefulness of a new DQA system, Dosimetry Check™(DC), on TomoTherapy® -based helical IMRT plans.Entities:
Keywords: Dosimetry Check™; TomoTherapy®; delivery quality assurance (DQA); pretreatment verification
Mesh:
Year: 2018 PMID: 30354001 PMCID: PMC6236814 DOI: 10.1002/acm2.12480
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1A treatment plan for TomoTherapy®‐based IMRT delivery verification in (a) axial, (b) coronal and (c) sagittal planes.
Comparison of Dosimetry Check™‐calculated, TomoTherapy® TPS‐calculated and Exradin A1SL ionization chamber‐measured doses for the IMRT verification plan shown in Fig. 1
| Field width (cm) | Machine | Absolute dose difference (%) | |
|---|---|---|---|
| DC − Exradin A1SL | TomoTherapy TPS − Exradin A1SL | ||
| 1.0 | Tomo HD | −0.5 | −0.8 |
| Tomo Hi‐Art | −1.1 | 0.4 | |
| 2.5 | Tomo HD | 0.0 | 0.2 |
| Tomo Hi‐Art | 1.6 | 1.5 | |
| 5.0 | Tomo HD | 0.7 | 1.1 |
| Tomo Hi‐Art | 0.8 | 1.5 | |
Figure 2Examples of the DQA results using Dosimetry Check™ for (a)–(c) head and neck and (e)–(g) abdominal cancer patients. For each treatment plan, the 2D planar dose distributions are compared in (a), (e) axial, (b), (f) coronal and (c), (g) sagittal planes. The magenta and green isolines refer to DC and TomoTherapy® TPS‐calculated dose distributions, respectively. Figures (d) and (h) present the dose–volume histogram of the target and normal structures calculated in Dosimetry Check™ (solid line) and TomoTherapy® TPS (dashed line).
Summary of DQA results of the 15 clinical TomoTherapy® treatment plans using the Dosimetry Check™ and cheese phantom method. The P‐value for the absolute dose difference determined by the two different DQA methods was calculated using Wilcoxon signed‐rank test
| Patient no. | Treatment machine | Treatment site | Absolute dose difference (%) | Gamma pass rate (%) | ||||||
|---|---|---|---|---|---|---|---|---|---|---|
| Dosimetry Check™ | Cheese phantom | Dosimetry Check™ | Cheese phantom | |||||||
| Axial | Coronal | Sagittal | 3D Body | Coronal | Sagittal | |||||
| 1 | Tomo HD | Prostate | 2.94 | 1.60 | 97.9 | 99.8 | 97.7 | 99.3 | – | 99.7 |
| 2 | Head and neck | 2.13 | 2.37 | 100 | 99.9 | 99.9 | 99.9 | 99.2 | – | |
| 3 | 1.06 | 1.19 | 98.2 | 99.4 | 99.9 | 99.5 | 99.9 | – | ||
| 4 | 1.73 | 0.15 | 99.2 | 98.1 | 100 | 99.5 | 99.9 | – | ||
| 5 | L‐spine | 1.89 | 1.65 | 97.6 | 99.5 | 99.8 | 99.7 | – | 99.9 | |
| 6 | T‐spine | 0.64 | −0.11 | 99.5 | 99.7 | 99.1 | 99.7 | 99.5 | – | |
| 7 | Abdomen | 4.22 | 0.38 | 46.5 | 82.2 | 87.8 | 84.2 | – | 99.6 | |
| 8 | Tomo Hi‐ART | Prostate | 1.51 | 1.20 | 99.6 | 99.8 | 99.3 | 99.4 | – | 99.6 |
| 9 | Head and neck | 1.56 | 0.90 | 100 | 100 | 100 | 99.7 | 96.5 | – | |
| 10 | −0.28 | 1.40 | 97.8 | 100 | 99.9 | 99.8 | 99.7 | – | ||
| 11 | 1.04 | 0.80 | 98.1 | 99.4 | 98.2 | 99.0 | 99.8 | – | ||
| 12 | Lung | −1.95 | −1.89 | 99.4 | 99.3 | 99.5 | 98.9 | 97.6 | – | |
| 13 | Retroperitoneum | −1.93 | 0.60 | 95.6 | 98.4 | 98.9 | 99.2 | – | 96.8 | |
| 14 | −2.05 | −1.35 | 96.0 | 98.2 | 98.1 | 98.2 | – | 96.0 | ||
| 15 | Brain | 1.74 | −1.04 | 95.6 | 99.5 | 100 | 98.8 | 99.1 | – | |
|
| 0.256 | N/A | ||||||||