Evaluation of a 3D diode array dosimeter for helical tomotherapy delivery QA.

Vladimir Feygelman1, D Opp, K Javedan, A J Saini, G Zhang.   

Abstract

The Delta4 biplanar diode array dosimeter was validated for helical tomotherapy delivery QA. The basic detector characteristics were found to be satisfactory in terms of short-term reproducibility (0.1%), linearity (<0.1%), dose rate dependence (0.4%), and absolute calibration accuracy (0.4% in the center of the phantom compared with the independently calibrated diode). Relative calibration of the arrays was verified by comparison with film and by rotating the detector 180°. The dosimeter response to rotational irradiation changed by no more than 0.2% when one of the detector boards was replaced by the homogeneous phantom material. The daily output correction factor can be derived from a Delta4 measurement in a uniform cylindrical field. The γ(3%, 3 mm) passing rate (absolute dose) was above 90% for all 9 evaluated clinical plans, and above 96% for all but one. The mean passing rate was 97 ± 2.7%. The plans varied in modulation factor, pitch, and calculation grid size. For best results, the phantom needs to be aligned carefully, preferably by megavoltage computed tomography imaging.
Copyright © 2010 American Association of Medical Dosimetrists. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20097061     DOI: 10.1016/j.meddos.2009.10.007

Source DB:  PubMed          Journal:  Med Dosim        ISSN: 1873-4022            Impact factor:   1.482


  11 in total

1.  Fidelity of dose delivery at high dose rate of volumetric modulated arc therapy in a truebeam linac with flattening filter free beams.

Authors:  Georgios Kalantzis; Jianguo Qian; Bin Han; Gary Luxton
Journal:  J Med Phys       Date:  2012-10

2.  Commissioning compensator-based IMRT on the Pinnacle treatment planning system.

Authors:  Daniel Opp; Kenneth Forster; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2011-03-08       Impact factor: 2.102

3.  Evaluation of a new VMAT QA device, or the "X" and "O" array geometries.

Authors:  Vladimir Feygelman; Geoffrey Zhang; Craig Stevens; Benjamin E Nelms
Journal:  J Appl Clin Med Phys       Date:  2011-01-31       Impact factor: 2.102

4.  Validation of an improved helical diode array and dose reconstruction software using TG-244 datasets and stringent dose comparison criteria.

Authors:  Saeed Ahmed; Benjamin Nelms; Jakub Kozelka; Geoffrey Zhang; Eduardo Moros; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2016-11-08       Impact factor: 2.102

5.  Measurement-guided volumetric dose reconstruction for helical tomotherapy.

Authors:  Cassandra Stambaugh; Benjamin Nelms; Theresa Wolf; Richard Mueller; Mark Geurts; Daniel Opp; Eduardo Moros; Geoffrey Zhang; Vladimir Feygelman
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

6.  MLC positional accuracy evaluation through the Picket Fence test on EBT2 films and a 3D volumetric phantom.

Authors:  Christos Antypas; Ioannis Floros; Maritina Rouchota; Christina Armpilia; Maria Lyra
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

7.  Gonadal-sparing total body irradiation with the use of helical tomotherapy for nonmalignant indications.

Authors:  Khaled Dibs; Austin J Sim; José A Peñagaricano; Kujtim Latifi; Genevieve A Garcia; Julia A Peters; Michael L Nieder; Sungjune Kim; Timothy J Robinson
Journal:  Rep Pract Oncol Radiother       Date:  2021-02-25

8.  Evaluation of a software system for estimating planned dose error in patients, based on planar IMRT QA measurements.

Authors:  Mohammad Bakhtiari; Ashkan Parniani; Fritz Lerma; Shannon Reynolds; James Jordan; Alireza Sedaghat; Mehrdad Sarfaraz; James Rodgers
Journal:  Radiol Oncol       Date:  2014-01-22       Impact factor: 2.991

9.  Twin machines validation for VMAT treatments using electronic portal-imaging device: a multicenter study.

Authors:  P Fenoglietto; M Khodri; D Nguyen; F Josserand-Pietri; N Aillères
Journal:  Radiat Oncol       Date:  2016-01-14       Impact factor: 3.481

10.  Clinical implementation of Dosimetry Check™ for TomoTherapy® delivery quality assurance.

Authors:  Eunah Chung; Dongyeol Kwon; Taeyang Park; Hyeri Kang; Yoonsun Chung
Journal:  J Appl Clin Med Phys       Date:  2018-10-24       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.