OBJECTIVE: To investigate the validity of an eight-contact electrode bioelectrical impedance analysis (BIA) system within a household scale for assessing whole body composition in COPD patients.Entities:
Mesh:
Year: 2018 PMID: 30328930 PMCID: PMC6326721 DOI: 10.1590/S1806-37562017000000121
Source DB: PubMed Journal: J Bras Pneumol ISSN: 1806-3713 Impact factor: 2.624
Characteristics of the study group.
| Characteristic | Mean ± SD | Range |
|---|---|---|
| Age, years | 67 ± 8 | 56-79 |
| Smoking history, pack-years | 50 (24-74)a | 21-150 |
| FEV1/FVCb | 0.45 ± 0.11 | 0.26-0.65 |
| FEV1, Lb | 1.17 ± 0.48 | 0.54-2.04 |
| FEV1, % predictedb | 38.6 ± 16.1 | 15-65 |
| FVC, Lb | 2.53 ± 0.61 | 1.51-3.49 |
| FVC, % predictedb | 64.5 ± 16.0 | 42-103 |
| BMI, kg/m2 | 24.7 ± 5.4 | 16.9-33.9 |
BMI: body mass index. aValue expressed as median (interquartile range). bPost-bronchodilator values.
Body composition of the study group.
| Variable | DEXA | BIA | p | ||
|---|---|---|---|---|---|
| Mean ± SD | Range | Mean ± SD | Range | ||
| FFM, kg | 50.7 ± 8.5 | 35.1-64.3 | 50.6 ± 8.6 | 38.9-63.0 | 0.848 |
| LM, kg | 48.3 ± 8.2 | 33.6-61.4 | 48.0 ± 8.2 | 36.9-59.9 | 0.730 |
| BMC, kg | 2.43 ± 0.46 | 1.59-3.35 | 2.56 ± 0.39 | 2.00-3.10 | 0.107 |
| FM, kg | 20.7 ± 9.4 | 5.06-33.9 | 21.3 ± 8.3 | 9.89-35.3 | 0.471 |
DEXA: dual-energy X-ray absorptiometry; BIA: bioelectrical impedance analysis; FFM: fat free mass; LM: lean mass; BMC: bone mineral content; and FM: fat mass.
Figure 1Correlation between body composition variables assessed by dual-energy X-ray absorptiometry and the eight-contact electrode bioelectrical impedance analysis system. DEXA: dual-energy X-ray absorptiometry; BIA: bioelectrical impedance analysis; FFM: fat free mass; LM: lean mass; BMC: bone mineral content; and FM: fat mass.
Figure 2Mean differences and limits of agreement between body composition variables assessed by dual-energy X-ray absorptiometry and the eight-contact electrode bioelectrical impedance analysis system. DEXA: dual-energy X-ray absorptiometry; BIA: bioelectrical impedance analysis; FFM: fat free mass; LM: lean mass; BMC: bone mineral content; and FM: fat mass.