Literature DB >> 30324225

Noninvasive tissue adhesive for cardiac implantable electronic device pocket closure: the TAPE pilot study.

S M Koerber1, T Loethen2, M Turagam3, J Payne2, R Weachter2, G Flaker2, M R Gold4, S Gautam2.   

Abstract

PURPOSE: Device infection is a serious complication of cardiac implantable electronic devices (CIED). Ensuring complete pocket closure can be time consuming, but remains vital to prevent infection. The Zip® Surgical Skin Closure (ZIP) is a noninvasive adhesive device applied to the skin as an alternative to subcuticular sutures for skin closure. We hypothesized that using this device would decrease pocket closure times without increasing the risk of pocket infections. This is a single center, retrospective cohort study to compare pocket closure times and infection rates between ZIP and standard suture for CIED pocket closure.
METHODS: Two separate groups of consecutive new intravenous implants, upgrades, and pulse generator replacements from October 2015 to April 2017 were included. A total of 175 patients were included, using either ZIP (n = 80) or suture (n = 95). Total procedure time (local anesthetic to dressing application) and pocket closure time (fascial suture to dressing application) were compared. Pocket infections were defined as infections leading to CIED extraction or wound dehiscence requiring repeat procedure. Statistical analysis was performed using chi square test and Student's t test.
RESULTS: Pocket closure time and procedure time were significantly shorter for the ZIP group (14.9 ± 6.8 vs 20.1 ± 11.09 min, p = 0.0003) and (65.02 ± 30.4 vs 83.83 ± 40.3 min, p = 0.0008), respectively. No pocket infections occurred in the Zip group, while the suture group had 2:1 wound dehiscence and 1 pocket infection.
CONCLUSION: The ZIP device resulted in significantly shorter pocket closure and procedure times without increasing device pocket infections.

Entities:  

Keywords:  Cardiac device implant; Device pocket closure

Mesh:

Substances:

Year:  2018        PMID: 30324225     DOI: 10.1007/s10840-018-0457-5

Source DB:  PubMed          Journal:  J Interv Card Electrophysiol        ISSN: 1383-875X            Impact factor:   1.900


  17 in total

1.  Comparison of skin adhesive and absorbable intracutaneous suture for the implantation of cardiac rhythm devices.

Authors:  Sebastian Spencker; Nalan Coban; Lydia Koch; Alexander Schirdewan; Dirk Mueller
Journal:  Europace       Date:  2010-11-11       Impact factor: 5.214

Review 2.  Commonly used suturing techniques in skin surgery.

Authors:  R L Moy; A Lee; A Zalka
Journal:  Am Fam Physician       Date:  1991-11       Impact factor: 3.292

Review 3.  Sutures versus staples for the management of surgical wounds: a meta-analysis of randomized controlled trials.

Authors:  Christos Iavazzo; Ioannis D Gkegkes; Evridiki K Vouloumanou; Ioannis Mamais; George Peppas; Matthew E Falagas
Journal:  Am Surg       Date:  2011-09       Impact factor: 0.688

4.  16-year trends in the infection burden for pacemakers and implantable cardioverter-defibrillators in the United States 1993 to 2008.

Authors:  Arnold J Greenspon; Jasmine D Patel; Edmund Lau; Jorge A Ochoa; Daniel R Frisch; Reginald T Ho; Behzad B Pavri; Steven M Kurtz
Journal:  J Am Coll Cardiol       Date:  2011-08-30       Impact factor: 24.094

5.  Standardized assessment of breast cancer surgical scars integrating the Vancouver Scar Scale, Short-Form McGill Pain Questionnaire, and patients' perspectives.

Authors:  Pauline T Truong; Freddy Abnousi; Celina M Yong; Allen Hayashi; James A Runkel; Theressa Phillips; Ivo A Olivotto
Journal:  Plast Reconstr Surg       Date:  2005-10       Impact factor: 4.730

6.  A review of scar scales and scar measuring devices.

Authors:  Regina Fearmonti; Jennifer Bond; Detlev Erdmann; Howard Levinson
Journal:  Eplasty       Date:  2010-06-21

7.  Management and outcome of permanent pacemaker and implantable cardioverter-defibrillator infections.

Authors:  Muhammad R Sohail; Daniel Z Uslan; Akbar H Khan; Paul A Friedman; David L Hayes; Walter R Wilson; James M Steckelberg; Sarah Stoner; Larry M Baddour
Journal:  J Am Coll Cardiol       Date:  2007-04-23       Impact factor: 24.094

8.  Renal insufficiency and the risk of infection from pacemaker or defibrillator surgery.

Authors:  Heather Bloom; Brian Heeke; Angel Leon; Fernando Mera; David Delurgio; John Beshai; Jonathan Langberg
Journal:  Pacing Clin Electrophysiol       Date:  2006-02       Impact factor: 1.976

9.  Risk factors related to infections of implanted pacemakers and cardioverter-defibrillators: results of a large prospective study.

Authors:  Didier Klug; Mamadou Balde; Dominique Pavin; Françoise Hidden-Lucet; Jacques Clementy; Nicolas Sadoul; Jean Luc Rey; Gilles Lande; Arnaud Lazarus; Jacques Victor; Claude Barnay; Bruno Grandbastien; Salem Kacet
Journal:  Circulation       Date:  2007-08-27       Impact factor: 29.690

10.  Risk factors and time delay associated with cardiac device infections: Leiden device registry.

Authors:  J C Lekkerkerker; C van Nieuwkoop; S A Trines; J G van der Bom; A Bernards; E T van de Velde; M Bootsma; K Zeppenfeld; J W Jukema; J-W Borleffs; M J Schalij; L van Erven
Journal:  Heart       Date:  2008-11-26       Impact factor: 5.994

View more
  1 in total

1.  A novel zipper device versus sutures for wound closure after surgery: a systematic review and meta-analysis.

Authors:  Cheng-Xin Xie; Cheng-Qiang Yu; Wei Wang; Cheng-Long Wang; Dong Yin
Journal:  Int Wound J       Date:  2020-08-26       Impact factor: 3.315

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.