Caval to pulmonary 3D flow distribution in patients with Fontan circulation and impact of potential 4D flow MRI error sources.

Kelly Jarvis1,2, Susanne Schnell1, Alex J Barker1, Michael Rose3, Joshua D Robinson1,3,4,5, Cynthia K Rigsby1,3,4,5, Michael Markl1,2.   

Abstract

PURPOSE: Uneven flow distribution in patients with Fontan circulation is suspected to lead to complications. 4D flow MRI offers evaluation using time-resolved pathlines; however, the potential error is not well understood. The aim of this study was to systematically assess variability in flow distribution caused by well-known sources of error.
METHODS: 4D flow MRI was acquired in 14 patients with Fontan circulation. Flow distribution was quantified by the % of caval venous flow pathlines reaching the left and right pulmonary arteries. Impact of data acquisition and data processing uncertainties were investigated by (1) probabilistic 4D blood flow tracking at varying noise levels, (2) down-sampling to mimic acquisition at different spatial resolutions, (3) pathline calculation with and without eddy current correction, and (4) varied segmentation of the Fontan geometry to mimic analysis errors.
RESULTS: Averaged among the cohort, uncertainties accounted for flow distribution errors from noise ≤3.2%, low spatial resolution ≤2.3% to 3.8%, eddy currents ≤6.4%, and inaccurate segmentation ≤3.9% to 9.1% (dilation and erosion, respectively). In a worst-case scenario (maximum additive errors for all 4 sources), flow distribution errors were as high as 22.5%.
CONCLUSION: Inaccuracies related to postprocessing (segmentation, eddy currents) resulted in the largest potential error (≤15.5% combined) whereas errors related to data acquisition (noise, low spatial resolution) had a lower impact (≤5.5%-7.0% combined). Whereas it is unlikely that these errors will be additive or affect the identification of severe asymmetry, these results illustrate the importance of eddy current correction and accurate segmentation to minimize Fontan flow distribution errors.
© 2018 International Society for Magnetic Resonance in Medicine.

Entities:  

Keywords:  4D flow MRI; Fontan circulation; background phase errors; congenital heart disease; flow distribution; probabilistic tracking; segmentation; uncertainty; velocity noise

Mesh:

Year:  2018        PMID: 30277276      PMCID: PMC6289652          DOI: 10.1002/mrm.27455

Source DB:  PubMed          Journal:  Magn Reson Med        ISSN: 0740-3194            Impact factor:   4.668


  28 in total

1.  Generalized autocalibrating partially parallel acquisitions (GRAPPA).

Authors:  Mark A Griswold; Peter M Jakob; Robin M Heidemann; Mathias Nittka; Vladimir Jellus; Jianmin Wang; Berthold Kiefer; Axel Haase
Journal:  Magn Reson Med       Date:  2002-06       Impact factor: 4.668

2.  Conversion from total cavopulmonary shunt to Fontan circulation: improved cyanosis with an 11-year interval.

Authors:  Takeshi Hiramatsu; Shigeru Komori; Yoshiharu Nishimura; Yoshitaka Okamura; Hiroyuki Suzuki; Takashi Takeuchi
Journal:  Ann Thorac Cardiovasc Surg       Date:  2008-02       Impact factor: 1.520

3.  Parallel MRI with extended and averaged GRAPPA kernels (PEAK-GRAPPA): optimized spatiotemporal dynamic imaging.

Authors:  Bernd Jung; Peter Ullmann; Matthias Honal; Simon Bauer; Jürgen Hennig; Michael Markl
Journal:  J Magn Reson Imaging       Date:  2008-11       Impact factor: 4.813

4.  Probabilistic 4D blood flow tracking and uncertainty estimation.

Authors:  Ola Friman; Anja Hennemuth; Andreas Harloff; Jelena Bock; Michael Markl; Heinz-Otto Peitgen
Journal:  Med Image Anal       Date:  2011-06-07       Impact factor: 8.545

5.  Image-based background phase error correction in 4D flow MRI revisited.

Authors:  Julia Busch; Daniel Giese; Sebastian Kozerke
Journal:  J Magn Reson Imaging       Date:  2017-02-22       Impact factor: 4.813

6.  Time-resolved three-dimensional magnetic resonance velocity mapping of cardiovascular flow paths in volunteers and patients with Fontan circulation.

Authors:  Michael Markl; Julia Geiger; Philip J Kilner; Daniela Föll; Brigitte Stiller; Friedhelm Beyersdorf; Raoul Arnold; Alex Frydrychowicz
Journal:  Eur J Cardiothorac Surg       Date:  2010-07-02       Impact factor: 4.191

7.  Surgical repair of tricuspid atresia.

Authors:  F Fontan; E Baudet
Journal:  Thorax       Date:  1971-05       Impact factor: 9.139

8.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

9.  Semiautomated method for noise reduction and background phase error correction in MR phase velocity data.

Authors:  P G Walker; G B Cranney; M B Scheidegger; G Waseleski; G M Pohost; A P Yoganathan
Journal:  J Magn Reson Imaging       Date:  1993 May-Jun       Impact factor: 4.813

10.  Hepatic venous blood and the development of pulmonary arteriovenous malformations in congenital heart disease.

Authors:  D Srivastava; T Preminger; J E Lock; V Mandell; J F Keane; J E Mayer; H Kozakewich; P J Spevak
Journal:  Circulation       Date:  1995-09-01       Impact factor: 29.690

View more
  5 in total

1.  Respiration Dependency of Caval Blood Flow in Patients with Fontan Circulation: Quantification Using 5D Flow MRI.

Authors:  Rene Bastkowski; Robert Bindermann; Konrad Brockmeier; Kilian Weiss; David Maintz; Daniel Giese
Journal:  Radiol Cardiothorac Imaging       Date:  2019-10-31

Review 2.  4D flow MRI applications in congenital heart disease.

Authors:  Judy Rizk
Journal:  Eur Radiol       Date:  2020-09-01       Impact factor: 5.315

3.  Enhancement of cerebrovascular 4D flow MRI velocity fields using machine learning and computational fluid dynamics simulation data.

Authors:  David R Rutkowski; Alejandro Roldán-Alzate; Kevin M Johnson
Journal:  Sci Rep       Date:  2021-05-13       Impact factor: 4.379

4.  Multipoint 5D flow cardiovascular magnetic resonance - accelerated cardiac- and respiratory-motion resolved mapping of mean and turbulent velocities.

Authors:  Jonas Walheim; Hannes Dillinger; Sebastian Kozerke
Journal:  J Cardiovasc Magn Reson       Date:  2019-07-22       Impact factor: 5.364

5.  Reduced scan time and superior image quality with 3D flow MRI compared to 4D flow MRI for hemodynamic evaluation of the Fontan pathway.

Authors:  Friso M Rijnberg; Hans C van Assen; Joe F Juffermans; Lucia J M Kroft; Pieter J van den Boogaard; Patrick J H de Koning; Mark G Hazekamp; Séline F S van der Woude; Evangeline G Warmerdam; Tim Leiner; Heynric B Grotenhuis; Jelle J Goeman; Hildo J Lamb; Arno A W Roest; Jos J M Westenberg
Journal:  Sci Rep       Date:  2021-03-22       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.