Can Realistic Liver Tissue Surrogates Accurately Quantify the Impact of Reduced-kV Imaging on Attenuation and Contrast of Parenchyma and Lesions?

Andre Euler1, Justin Solomon2, Paul F FitzGerald3, Ehsan Samei2, Rendon C Nelson2.   

Abstract

RATIONALE AND
OBJECTIVES: To assess if a liquid tissue surrogate for the liver (LTSL) can emulate contrast-enhanced liver parenchyma and lesions and quantify the impact of reduced-kV imaging as a function of lesion contrast, phase of enhancement, and phantom size.
MATERIALS AND METHODS: First, CT attenuation of LTSL- and water-iodine solutions were measured as a function of iodine concentration and tube potential. For each solution, the iodine concentration was determined to emulate liver parenchyma at 120 kV. CT attenuation for both solutions was predicted for different tube potentials and compared to published patient data. Second, liver parenchyma in late arterial phase (LA: +92 HU at 120 kV) and portal venous phase (PV: +112 HU at 120 kV) was emulated using LTSL-iodine and a two-size phantom. Fourteen setups of hyper- and hypoattenuating lesions (lesion-to-parenchyma contrast (CLP) = -50 to +50HU) were created. Each combination of CLP, phase, and size was imaged at 80, 100, 120, and 140 kV at constant radiation dose. CT attenuation, CLP, and lesion-to-parenchyma contrast-to-noise ratio (CNRLP) were assessed and compared to a theoretical model.
RESULTS: LTSL-iodine more accurately emulated the CT attenuation of liver parenchyma across different tube potentials compared to water-iodine solutions. The theoretical model was confirmed by the empirical measurements using LTSL-iodine solutions: attenuation, CLP, and CNRLP increased when the tube potential decreased (p < 0.001). This trend was independent of lesion contrast, phase, and size. The absolute improvement in CLP and CNRLP, however, was inversely related to the magnitude of CLP at 140kV.
CONCLUSION: LTSL accurately emulated the energy-dependent CT attenuation characteristics of contrast-enhanced liver parenchyma and lesions. The relative improvement in CLP and CNRLP by applying reduced-kV imaging was independent of lesion contrast, phase, and size while the absolute improvement decreased for low-contrast lesions.
Copyright © 2018 The Association of University Radiologists. Published by Elsevier Inc. All rights reserved.

Entities:  

Keywords:  Computed tomography; Lesion; Liquid tissue surrogate; Liver; Tube potential

Mesh:

Substances:

Year:  2018        PMID: 30269958      PMCID: PMC6437027          DOI: 10.1016/j.acra.2018.08.008

Source DB:  PubMed          Journal:  Acad Radiol        ISSN: 1076-6332            Impact factor:   3.173


  20 in total

1.  Dual source dual energy MDCT: comparison of 80 kVp and weighted average 120 kVp data for conspicuity of hypo-vascular liver metastases.

Authors:  Emma Robinson; James Babb; Hersh Chandarana; Michael Macari
Journal:  Invest Radiol       Date:  2010-07       Impact factor: 6.016

2.  Dual-energy CT with tin filter technology for the discrimination of renal lesion proxies containing blood, protein, and contrast-agent. An experimental phantom study.

Authors:  Christoph Karlo; Arno Lauber; Robert Paul Götti; Stephan Baumüller; Paul Stolzmann; Hans Scheffel; Lotus Desbiolles; Bernhard Schmidt; Borut Marincek; Hatem Alkadhi; Sebastian Leschka
Journal:  Eur Radiol       Date:  2010-08-15       Impact factor: 5.315

3.  CT Image Contrast of High-Z Elements: Phantom Imaging Studies and Clinical Implications.

Authors:  Paul F FitzGerald; Robert E Colborn; Peter M Edic; Jack W Lambert; Andrew S Torres; Peter J Bonitatibus; Benjamin M Yeh
Journal:  Radiology       Date:  2015-09-10       Impact factor: 11.105

4.  Hybrid iterative reconstruction technique for abdominal CT protocols in obese patients: assessment of image quality, radiation dose, and low-contrast detectability in a phantom.

Authors:  Sebastian T Schindera; Devang Odedra; Diego Mercer; Seng Thipphavong; Paul Chou; Zsolt Szucs-Farkas; Patrik Rogalla
Journal:  AJR Am J Roentgenol       Date:  2014-02       Impact factor: 3.959

5.  CT of the abdomen: degree and quality of enhancement obtained with two concentrations of the same iodinated contrast medium with fixed iodine delivery rate and total iodine load.

Authors:  Francesco Paparo; Isabella Garello; Lorenzo Bacigalupo; Andrea Marziano; Athena Galletto Pregliasco; Ludovica Rollandi; Cristina Puppo; Francesca Mattioli; Matteo Puntoni; Gian Andrea Rollandi
Journal:  Eur J Radiol       Date:  2014-08-12       Impact factor: 3.528

6.  Iodine quantification with dual-energy CT: phantom study and preliminary experience with renal masses.

Authors:  Hersh Chandarana; Alec J Megibow; Benjamin A Cohen; Ramya Srinivasan; Danny Kim; Christianne Leidecker; Michael Macari
Journal:  AJR Am J Roentgenol       Date:  2011-06       Impact factor: 3.959

7.  Effect of tumor size and tumor-to-liver contrast of hypovascular liver tumors on the diagnostic performance of hepatic CT imaging.

Authors:  Sebastian T Schindera; Lisa F Hareter; Stephan Raible; Jaled Charimo Torrente; Oliver Rusch; Andrea B Rosskopf; Daniele Marin; Peter Vock; Zsolt Szucs-Farkas
Journal:  Invest Radiol       Date:  2012-03       Impact factor: 6.016

8.  Investigation of American Association of Physicists in Medicine Report 204 size-specific dose estimates for pediatric CT implementation.

Authors:  Samuel L Brady; Robert A Kaufman
Journal:  Radiology       Date:  2012-10-23       Impact factor: 11.105

9.  Liquid tissue surrogates for X-ray and CT phantom studies.

Authors:  Paul F FitzGerald; Robert E Colborn; Peter M Edic; Jack W Lambert; Peter J Bonitatibus; Benjamin M Yeh
Journal:  Med Phys       Date:  2017-10-31       Impact factor: 4.071

10.  The characterization of small hypoattenuating renal masses on contrast-enhanced CT.

Authors:  Neesha S Patel; Liina Poder; Zhen J Wang; Benjamin M Yeh; Aliya Qayyum; Hua Jin; Fergus V Coakley
Journal:  Clin Imaging       Date:  2009 Jul-Aug       Impact factor: 1.605

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.