Comparison between transcutaneous bilirubinometry and total serum bilirubin measurements in preterm infants <35 weeks gestation.

M Ahmed1, S Mostafa, G Fisher, T M Reynolds.   

Abstract

BACKGROUND: Neonatal hyperbilirubinaemia is a common treatable cause of brain injury. The treatment for this condition is phototherapy. The decision whether to use phototherapy is currently dependent upon serum bilirubin assay results. However, repeated blood sampling is not only traumatic but may also be a cause of anaemia in neonates. We evaluated a transcutaneous bilirubin assay method to determine whether it was suitable for routine use in preterm infants.
METHODS: One hundred and eighty-three transcutaneous bilirubin measurements were taken contemporaneously with blood samples for laboratory measurement of serum bilirubin. The study was carried out with informed parental consent and approval by the local research ethics committee.
RESULTS: The transcutaneous bilirubin method (BiliChek) exhibited a consistent positive bias compared with the laboratory bilirubin assay. Consequently, for a given detection rate, the transcutaneous method had a higher screen positive rate, i.e. more neonates would be given phototherapy if transcutaneous bilirubin results were used to decide. There was a margin of safety in the transcutaneous bilirubin assay calibration.
CONCLUSION: The BiliChek transcutaneous bilirubin assay is a safe alternative to laboratory bilirubin assay in deciding whether to give preterm neonates phototherapy.

Entities:  

Mesh:

Substances:

Year:  2009        PMID: 19940207     DOI: 10.1258/acb.2009.009072

Source DB:  PubMed          Journal:  Ann Clin Biochem        ISSN: 0004-5632            Impact factor:   2.057


  6 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

2.  Interscapular site for transcutaneous bilirubin measurement in preterm infants: a better and safer screening site.

Authors:  A Yaser; L Tooke; N Rhoda
Journal:  J Perinatol       Date:  2014-01-09       Impact factor: 2.521

3.  Defining the limitations of transcutaneous bilirubin measurement in late preterm newborns.

Authors:  K L Fine; W A Carey; J A W Schuster; S C Bryant; W J Cook; B S Karon
Journal:  J Perinatol       Date:  2017-02-16       Impact factor: 2.521

4.  Transcutaneous bilirubinometry during and after phototherapy in preterm infants: a prospective observational study.

Authors:  Ali Ahmed Raba; Anne O'Sullivan; Jan Miletin
Journal:  BMJ Paediatr Open       Date:  2020-07-16

5.  BiliCheck vs JM-103 in identifying neonates not at risk of hyperbilirubinaemia.

Authors:  Costantino Romagnoli; Piero Catenazzi; Giovanni Barone; Lucia Giordano; Riccardo Riccardi; Antonio Alberto Zuppa; Enrico Zecca
Journal:  Ital J Pediatr       Date:  2013-07-23       Impact factor: 2.638

6.  Agreement test of transcutaneous bilirubin and bilistick with serum bilirubin in preterm infants receiving phototherapy.

Authors:  Rinawati Rohsiswatmo; Hanifah Oswari; Radhian Amandito; Hikari Ambara Sjakti; Endang Windiastuti; Rosalina Dewi Roeslani; Indrayady Barchia
Journal:  BMC Pediatr       Date:  2018-09-29       Impact factor: 2.125

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.