Reliability of transcutaneous bilirubin devices in preterm infants: a systematic review.

Gaurav Nagar1, Ben Vandermeer, Sandra Campbell, Manoj Kumar.   

Abstract

BACKGROUND AND
OBJECTIVE: Transcutaneous bilirubin (TcB) devices are widely used for the estimation of serum bilirubin levels in term and near-term infants. Our objective was to review the diagnostic accuracy of TcB devices in preterm infants.
METHODS: Medline, Embase, Cochrane library, Cumulative Index to Nursing and Allied Health Literature, and Scopus were searched (from database inception date until December 2012). Additional citations were identified by using the bibliographies of selected articles and from conference proceedings. The studies were included if they compared TcB with total serum bilirubin in preterm infants before phototherapy and presented data as correlation coefficients or as Bland-Altman difference plots. Data were extracted by 1 reviewer and checked for accuracy by the second reviewer. An assessment tool (quality assessment of diagnostic accuracy studies) was used for risk of bias assessments.
RESULTS: Twenty-two studies met the inclusion criteria; 21 studies reported results as correlation coefficients, with pooled estimates of r = 0.83 for each site of measurement. Pooled estimates in infants <32 weeks' gestation were similar to the overall preterm population (r = 0.89 [95% confidence interval: 0.82-0.93]). For the 2 commonly used TcB devices (ie, JM103 and BiliCheck), the results were comparable at the forehead site, although the JM103 device exhibited better correlation at the sternum. Analysis of the Bland-Altman plots (13 studies) revealed negligible bias in measurement at the forehead or sternum site by using either the JM-103 or BiliCheck device; however, the JM-103 device exhibited better precision than the BiliCheck (SD for TcB - total serum bilirubin differences: 24.3 and 31.98 µmol/L, respectively).
CONCLUSIONS: The TcB devices reliably estimated bilirubin levels in preterm infants and could be used in clinical practice to reduce blood sampling.

Entities:  

Keywords:  Bland-Altman difference plots; correlation coefficient; meta-analysis; systematic review; transcutaneous bilirubin

Mesh:

Substances:

Year:  2013        PMID: 24127472     DOI: 10.1542/peds.2013-1713

Source DB:  PubMed          Journal:  Pediatrics        ISSN: 0031-4005            Impact factor:   7.124


  23 in total

1.  Screening neonatal jaundice based on the sclera color of the eye using digital photography.

Authors:  Terence S Leung; Karan Kapur; Ashley Guilliam; Jade Okell; Bee Lim; Lindsay W MacDonald; Judith Meek
Journal:  Biomed Opt Express       Date:  2015-10-23       Impact factor: 3.732

Review 2.  Transcutaneous Bilirubin for Screening of Significant Jaundice in Very Preterm Neonates.

Authors:  Deepak Chawla
Journal:  Indian J Pediatr       Date:  2018-10-29       Impact factor: 1.967

3.  Reliability of transcutaneous bilirubin determination based on skin color determined by a neonatal skin color scale of our own.

Authors:  Silvia Maya-Enero; Júlia Candel-Pau; Jordi Garcia-Garcia; Xavier Duran-Jordà; María Ángeles López-Vílchez
Journal:  Eur J Pediatr       Date:  2021-01-06       Impact factor: 3.183

4.  Transcutaneous bilirubin nomogram for evaluating the risk of hyperbilirubinemia in Iranian healthy newborns.

Authors:  Robabe Seyedi; Mojgan Mirghafourvand; Abdollah Jannat Dost; Sakineh Mohammad-Alizadeh-Charandabi; Mohammad Asghari Jafarabadi; Shirin Osouli Tabrizi
Journal:  World J Pediatr       Date:  2018-11-21       Impact factor: 2.764

5.  Reliability of transcutaneous bilirubinometry from shielded skin in neonates receiving phototherapy: a prospective cohort study.

Authors:  L Murli; A Thukral; M J Sankar; S Vishnubhatla; A K Deorari; V K Paul; A Sakariah; R Agarwal
Journal:  J Perinatol       Date:  2016-10-20       Impact factor: 2.521

6.  Transcutaneous bilirubin levels in newborns <35 weeks' gestation.

Authors:  M J Maisels; M P Coffey; E Kring
Journal:  J Perinatol       Date:  2015-06-25       Impact factor: 2.521

7.  Defining the limitations of transcutaneous bilirubin measurement in late preterm newborns.

Authors:  K L Fine; W A Carey; J A W Schuster; S C Bryant; W J Cook; B S Karon
Journal:  J Perinatol       Date:  2017-02-16       Impact factor: 2.521

8.  Minimizing blood loss and the need for transfusions in very premature infants.

Authors:  Brigitte Lemyre; Megan Sample; Thierry Lacaze-Masmonteil
Journal:  Paediatr Child Health       Date:  2015 Nov-Dec       Impact factor: 2.253

9.  Efficacy of transcutaneous bilirubinometry as compared to serum bilirubin in preterm newborn during phototherapy.

Authors:  Sunil Gothwal; Neelam Singh; Sadasivam Sitaraman; Ramesh Choudhary; Kailash Kumar Meena; Ghan Shyam Bairwa; Mohan Bairwa; Amrit Jeevan
Journal:  Eur J Pediatr       Date:  2021-06-12       Impact factor: 3.183

10.  Neonatal Jaundice Detection System.

Authors:  Mustafa Aydın; Fırat Hardalaç; Berkan Ural; Serhat Karap
Journal:  J Med Syst       Date:  2016-05-26       Impact factor: 4.460

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.