BACKGROUND: The vestibular evoked myogenic potential triggered by galvanic vestibular stimulation (galvanic-VEMP) has been used to assess the function of the vestibulospinal motor tract and is a candidate biomarker to predict and monitor the human T-cell lymphotropic virus type 1 (HTLV-1) associated myelopathy (HAM). This study determined the agreement and reliability of this exam.Entities:
Mesh:
Year: 2018 PMID: 30261002 PMCID: PMC6160040 DOI: 10.1371/journal.pone.0204449
Source DB: PubMed Journal: PLoS One ISSN: 1932-6203 Impact factor: 3.240
Fig 1Vestibular-evoked myogenic potential triggered by galvanic vestibular stimulation procedure.
The standing position of the patient (barefoot on a hard flat surface with eyes closed, feet close together and body leaning forward in order to cause the gastrocnemius muscle contraction); the equipment used for stimulus generation (a); the electrode positions for GVS (b); the electrode position for electromyography on the gastrocnemius muscle (c); the equipment for signal processing (d); and the laptop (e) connected to (a) and (d).
Fig 2Normal, delayed, and abnormal response patterns in vestibular-evoked myogenic potential triggered by galvanic vestibular stimulation (galvanic-VEMP).
(A) Normal electromyographic (EMG) response recorded from the gastrocnemius muscle. The black line indicates the trace with the cathode on the right and the anode on the left, whereas the gray line indicates the opposite stimulation polarity. SL (~50 ms) and ML (~100 ms). (B) Delayed EMG responses. SL ~80 ms and ML ~150 ms. (C) Absent EMG response, no SL and no ML.
Galvanic-VEMP variables (SL and ML): Comparison between groups.
| Variable | HTLV-1 negative controls (n = 45) | Asymptomatic carriers (n = 27) | HAM (n = 24) | p-value | |
|---|---|---|---|---|---|
| SL | Median (IQR) | 56 (53–64) | 53 (52–63) | 63 (56–75) | 0.089 |
| normal | 29(64.4%) | 14 (51.9%) | 8 (33.3%) | 0.067 | |
| delayed | 10 (22.2%) | 4 (14.8%) | 8 (33.3%) | ||
| absent | 6 (13.4%) | 9 (33.3%) | 8 (33.3%) | ||
| ML | Median (IQR) | 114 (105–126) | 116 (101–130) | 136 (124–144) | <0.001 |
| normal | 42 (93.3%) | 17 (63%) | 8 (33.3%) | <0.001 | |
| delayed | 2 (4.4%) | 4 (14.8%) | 11 (45.8%) | ||
| absent | 1 (2.3%) | 6 (22.2%) | 5 (20.9%) | ||
Notes: SL delay: > 63ms; ML delay: > 132ms; IQR: interquartile range. Statistical tests: Kruskal-Wallis for continuous SL and ML values; Qui-square for categorized SL and ML.
* statistically different group.
Intrarater (test-retest) and interrater agreement and reliability measures of galvanic-VEMP variables (SL and ML) in the total sample (n = 96).
| Variable | SEM | CR | ICC | P value | Kappa | P value | |
|---|---|---|---|---|---|---|---|
| Intrarater | SL | 6 | 16 | 0.803 | <0.001 | 0.533 | <0.001 |
| ML | 8 | 22 | 0.913 | <0.001 | 0.829 | <0.001 | |
| Interrater | SL | 3 | 8 | 0.953 | <0.001 | 0.769 | <0.001 |
| ML | 10 | 27 | 0.863 | <0.001 | 0.884 | <0.001 |
SEM: standard error of measurement. CR: coefficient of repeatability. ICC: intraclass correlation coefficient.
Intrarater (test-retest) and interrater agreement and reliability measures of galvanic-VEMP variables (SL and ML) in individuals with HAM (n = 24).
| Variable | SEM | CR | ICC | P value | Kappa | P value | |
|---|---|---|---|---|---|---|---|
| Intrarater | SL | 7 | 19 | 0.850 | 0.001 | 0.438 | 0.002 |
| ML | 10 | 29 | 0.861 | <0.001 | 0.869 | <0.001 | |
| Interrater | SL | 3 | 8 | 0.978 | <0.001 | 0.813 | <0.001 |
| ML | 9 | 25 | 0.808 | 0.001 | 0.509 | <0.001 |
SEM: standard error of measurement. CR: coefficient of repeatability. ICC: intraclass correlation coefficient.
Intrarater (test-retest) and interrater agreement and reliability measures of galvanic-VEMP variables (SL and ML) in HTLV-1 negative controls (n = 45).
| Variable | SEM | CR | ICC | P value | Kappa | P value | |
|---|---|---|---|---|---|---|---|
| Intrarater | SL | 5 | 14 | 0.688 | <0.001 | 0.494 | <0.001 |
| ML | 7 | 19 | 0.860 | <0.001 | 0.567 | <0.001 | |
| Interrater | SL | 2 | 5 | 0.963 | <0.001 | 0.587 | <0.001 |
| ML | 12 | 33 | 0.752 | <0.001 | 0.395 | <0.001 |
SEM: standard error of measurement. RC: repeatability coefficient. ICC: intraclass correlation coefficient.
Intrarater (test-retest) and interrater agreement and reliability measures of galvanic-VEMP variables (SL and ML) in HTLV-1 asymptomatic carriers (n = 27).
| Variable | SEM | CR | ICC | P value | Kappa | P value | |
|---|---|---|---|---|---|---|---|
| Intrarater | SL | 6 | 16 | 0.694 | 0.014 | 0.567 | <0.001 |
| ML | 8 | 23 | 0.923 | <0.001 | 0.861 | <0.001 | |
| Interrater | SL | 4 | 12 | 0.861 | <0.001 | 0.878 | <0.001 |
| ML | 4 | 12 | 0.946 | <0.001 | 0.749 | <0.001 |
SEM: standard error of measurement. CR: coefficient of repeatability. ICC: intraclass correlation coefficient.