Literature DB >> 30258293

Left Ventricular Assist Device Inflow Cannula Position May Contribute to the Development of HeartMate II Left Ventricular Assist Device Pump Thrombosis.

Jay K Bhama1, Aditya Bansal2,3.   

Abstract

BACKGROUND: Pump thrombosis (PT) is a dreaded complication after left ventricular assist device (LVAD) implantation. Problems with inflow cannula (IC) position may precipitate thrombus development. We sought to determine if IC position contributes to the development of PT.
METHODS: We conducted a retrospective review of 76 HeartMate II LVAD implants. The angle of the IC (AIC) to the horizontal plane was measured on chest x-rays. Patients who developed PT (PT group) were compared to the remaining patients (control group).
RESULTS: The mean age at implantation was 56 ± 14 years, and 82% of the patients were male. Ten patients (13%) developed PT. Six (60%) required device exchange, and 4 (40%) were managed with anticoagulation and/or thrombolysis. The median AIC for all patients at implantation was 59° (range, 38°-98°; 25th-75th interquartile range, 50°-75°). In the PT group, the median AIC was larger at the time of PT diagnosis compared to implantation (70° vs 60°, P = 0.005). In the control group, the median AIC was also larger at follow-up compared to implantation (61° vs 58°, P < 0.001) although to a lesser degree than in the PT group. No difference was seen in the median AIC between the PT group and the control group at implantation (60° vs 58°, respectively; P = 0.668) or at follow-up (70° vs 61°, respectively; P = 0.309). However, the median AIC at follow-up in the PT group was significantly larger than the median AIC at implantation in the control group (70° vs 58°, respectively; P = 0.014).
CONCLUSION: The HeartMate II LVAD IC position contributes to the development of PT. Regular monitoring of cannula position may help identify patients at risk for this problem.

Entities:  

Keywords:  Cannula; chest x-ray; thrombosis; ventricular assist device

Year:  2018        PMID: 30258293      PMCID: PMC6135282          DOI: 10.31486/toj.17.0070

Source DB:  PubMed          Journal:  Ochsner J        ISSN: 1524-5012


  9 in total

1.  Diaphragmatic implantation of the HeartWare ventricular assist device.

Authors:  Igor D Gregoric; William E Cohn; O H Frazier
Journal:  J Heart Lung Transplant       Date:  2011-01-06       Impact factor: 10.247

2.  Performance of extracorporeally adjustable ventricular assist device inflow cannula.

Authors:  Akira Shiose; Hyun-Il Kim; Tohru Takaseya; Mariko Kobayashi; Masako Fujiki; Ji-Feng Chen; Ryan Klatte; Shengqiang Gao; Alex L Massiello; Diyar Saeed; Kiyotaka Fukamachi
Journal:  Ann Thorac Surg       Date:  2010-11       Impact factor: 4.330

3.  Development of an ex vivo ovine ventricular assist device model for intraventricular visualization of the inflow cannula.

Authors:  Jay K Bhama; Timothy N Bachman; Robert L Kormos; Harvey Borovetz; James F Antaki
Journal:  J Heart Lung Transplant       Date:  2009-08       Impact factor: 10.247

4.  Presentation and management of left ventricular assist device inflow cannula malposition.

Authors:  Carmelo A Milano; Alan A Simeone; Laura J Blue; Joseph G Rogers
Journal:  J Heart Lung Transplant       Date:  2011-04-13       Impact factor: 10.247

5.  Computed tomography correlates of inflow cannula malposition in a continuous-flow ventricular-assist device.

Authors:  Erik N Sorensen; Nicholas J Hiivala; Jean Jeudy; Keshava Rajagopal; Bartley P Griffith
Journal:  J Heart Lung Transplant       Date:  2013-04-10       Impact factor: 10.247

6.  Device exchange after primary left ventricular assist device implantation: indications and outcomes.

Authors:  John M Stulak; Jennifer Cowger; Jonathon W Haft; Matthew A Romano; Keith D Aaronson; Francis D Pagani
Journal:  Ann Thorac Surg       Date:  2012-10-11       Impact factor: 4.330

7.  Surgical technique influences HeartMate II left ventricular assist device thrombosis.

Authors:  Sharven Taghavi; Charisse Ward; Senthil N Jayarajan; John Gaughan; Lynn M Wilson; Abeel A Mangi
Journal:  Ann Thorac Surg       Date:  2013-08-20       Impact factor: 4.330

8.  Pump replacement for left ventricular assist device failure can be done safely and is associated with low mortality.

Authors:  Nader Moazami; Carmelo A Milano; Ranjit John; Benjamin Sun; Robert M Adamson; Francis D Pagani; Nicholas Smedira; Mark S Slaughter; David J Farrar; O Howard Frazier
Journal:  Ann Thorac Surg       Date:  2012-12-20       Impact factor: 4.330

9.  In-vitro Evaluation of Ventricular Cannulation for Rotodynamic Cardiac Assist Devices.

Authors:  Timothy N Bachman; Jay K Bhama; Josiah Verkaik; Stijn Vandenberghe; Robert L Kormos; James F Antaki
Journal:  Cardiovasc Eng Technol       Date:  2011-09       Impact factor: 2.495

  9 in total
  3 in total

Review 1.  Update on the Practical Role of Echocardiography in Selection, Implantation, and Management of Patients Requiring Left Ventricular Assist Device Therapy.

Authors:  Aashish Katapadi; Matt Umland; Bijoy K Khandheria
Journal:  Curr Cardiol Rep       Date:  2022-08-19       Impact factor: 3.955

2.  The influence of left ventricular assist device inflow cannula position on thrombosis risk.

Authors:  Mojgan Ghodrati; Alexander Maurer; Thomas Schlöglhofer; Thananya Khienwad; Daniel Zimpfer; Dietrich Beitzke; Francesco Zonta; Francesco Moscato; Heinrich Schima; Philipp Aigner
Journal:  Artif Organs       Date:  2020-05-06       Impact factor: 3.094

3.  Inflow cannula position as risk factor for stroke in patients with HeartMate 3 left ventricular assist devices.

Authors:  Thomas Schlöglhofer; Philipp Aigner; Marcel Migas; Dietrich Beitzke; Kamen Dimitrov; Franziska Wittmann; Julia Riebandt; Marcus Granegger; Dominik Wiedemann; Günther Laufer; Francesco Moscato; Heinrich Schima; Daniel Zimpfer
Journal:  Artif Organs       Date:  2022-01-09       Impact factor: 2.663

  3 in total