Performance of extracorporeally adjustable ventricular assist device inflow cannula.

Akira Shiose1, Hyun-Il Kim, Tohru Takaseya, Mariko Kobayashi, Masako Fujiki, Ji-Feng Chen, Ryan Klatte, Shengqiang Gao, Alex L Massiello, Diyar Saeed, Kiyotaka Fukamachi.   

Abstract

PURPOSE: This study evaluated the feasibility and efficacy of a newly developed adjustable left ventricular assist device inflow cannula in a short-term calf model. DESCRIPTION: In this inflow cannula, the angle between the cannula body and the inflow cannula tip can be altered extracorporeally by manipulating 2 externalized cables connected to the cannula. The cannula tip is adjustable in any plane to a maximum of ±15 degrees. EVALUATION: After initial prototyping in 4 calf cadavers, a Cleveland Heart left ventricular assist device was implanted with the adjustable inflow cannula placed in the left ventricular apex and the outlet to the descending aorta. Under hypovolemic conditions, the angle of the cannula tip could be changed to induce varying degrees of ventricular suction and then eliminate it, as evidenced by recorded pump and native left ventricular flows. Epicardial echocardiography and fluoroscopy in the closed-chest condition documented extracorporeal adjustments of the inflow cannula position.
CONCLUSIONS: This extracorporeally adjustable inflow cannula was effective in preventing or controlling left ventricular suction.
Copyright © 2010 The Society of Thoracic Surgeons. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2010        PMID: 20971290     DOI: 10.1016/j.athoracsur.2010.06.081

Source DB:  PubMed          Journal:  Ann Thorac Surg        ISSN: 0003-4975            Impact factor:   4.330


  1 in total

1.  Left Ventricular Assist Device Inflow Cannula Position May Contribute to the Development of HeartMate II Left Ventricular Assist Device Pump Thrombosis.

Authors:  Jay K Bhama; Aditya Bansal
Journal:  Ochsner J       Date:  2018
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.