| Literature DB >> 30217160 |
Jae Pil Chung1, Young Min Seong1, Tae Yeon Kim2, Yona Choi2, Tae Hoon Kim3, Hyun Joon Choi4, Chul Hee Min4, Hamza Benmakhlouf5, Kook Jin Chun6, Hyun-Tai Chung7.
Abstract
BACKGROUND: To measure the absorbed dose rate to water and penumbra of a Gamma Knife® (GK) using a polymethyl metacrylate (PMMA) phantom.Entities:
Keywords: Absorbed dose rate to water; Dose distribution; Gamma knife; PMMA phantom; Penumbra; Quality control; Scattered photon contribution
Mesh:
Substances:
Year: 2018 PMID: 30217160 PMCID: PMC6137750 DOI: 10.1186/s13014-018-1117-8
Source DB: PubMed Journal: Radiat Oncol ISSN: 1748-717X Impact factor: 3.481
Fig. 1Schematic diagrams and photographs of the PMMA phantom. a Design of the phantom with the chamber-hosting inner phantom. b Design of the phantom with the xy-plane film-hosting inner phantom. c Manufactured phantom combined with the chamber-hosting inner phantom. d The xz-plane film-hosting (left) and xy-plane film-hosting inner phantom (right)
Fig. 2Pictures of the experimental setup for measurement of the absorbed dose to water with different phantoms. a The PMMA phantom with a PTW 31010 ionization chamber was mounted to the Leksell Gamma Knife Perfexion for dose rate measurement. b The spherical water-filled phantom with a PTW 31010 ionization chamber. c A solid water phantom with a PTW 31010 ionization chamber. d The PMMA phantom with the xz-plane film-hosting inner phantom. The film shows the dose distribution of the 16-mm collimator in the xz-plane
Chemical composition and electron density of the PMMA used as the phantom material
| Component | fi | Z/A | ρe,PMMA | ρe,PMMA /ρe,water |
|---|---|---|---|---|
| H | 0.081 | 0.992162 | 0.639 | 1.155 |
| C | 0.595 | 0.499542 | ||
| O | 0.317 | 0.500031 | ||
| unknown | 0.007 | 0.500 |
Legend: fi is the fraction by weight of the atom i, Z/A is the atomic number weight mass ratio, ρe,PMMA is the electron density of PMMA and ρe,PMMA /ρe,water is the ratio of the electron densities of PMMA and water
Fig. 3Reconstructed coronal computed tomography images of the PMMA phantom with different inner phantoms. In each figure, the crossing point of the dashed lines is the PPS calibration center of a GK. a The effective point of the PTW TN31010 ionization chamber is 0.2 mm off in the x-direction (right-left). b The xy-plane film position is well matched with the xy-plane. c The xz-plane film position is well matched with the xz-plane
Fig. 4One-dimensional dose distributions using the 16-mm collimator of a Gamma Knife Perfexion. The solid lines indicate measured distributions in the PMMA phantom, and the solid circles indicate results calculated by the GK treatment planning program at the center of a sphere with an 8.0-cm radius. The measured data are expressed a band and its half width is one standard deviation of each point
FWHM and penumbra values measured in the PMMA phantom and the solid water phantom (SW) and calculated by LGP
| FWHM (mm) | Penumbra (mm) | |||||||
|---|---|---|---|---|---|---|---|---|
| Axis | PMMA | LGP6.93a | SW | LGP8.0b | PMMA | LGP6.93a | SW | LGP8.0b |
| x | 21.83 ± 0.25 | 21.68 | 21.57 ± 0.27 | 21.75 | 9.58 ± 0.22 | 9.04 | 8.76 ± 0.19 | 9.03 |
| y | 22.10 ± 0.32 | 21.68 | 21.56 ± 0.27 | 21.75 | 9.86 ± 0.18 | 9.04 | 8.96 ± 0.23 | 9.03 |
| z | 17.46 ± 0.11 | 17.39 | 17.41 ± 0.17 | 17.44 | 2.72 ± 0.10 | 3.09 | 2.65 ± 0.14 | 2.55 |
Legend: aCalculated by Leksell Gamma Plan at the center of the PMMA phantom (LGP6.93)
bCalculated by Leksell Gamma Plan and at the center of a sphere with a radius of 8.0 cm (LGP8.0)
The uncertainty components for measurement of the absorbed dose rate to water
| Uncertainty component | Type A (%) | Type B (%) |
|---|---|---|
| Calibration factor of the ionization chamber, | 0.50 | |
| Ionization current measurement | 0.01 | 0.02 |
| Temperature and pressure measurement | 0.056 | |
| Scattered photon contribution | 0.01 | 1.28 |
| Displacement of the ionization chamber | 0.01 | 0.15 |
| Alignment of the PMMA phantom with the PPS calibration center | 0.20 | |
| Long-term stability of the ionization chamber | 0.12 | |
| Standard relative combined uncertainty | 1.40 | |
Legend: A PTW TN31010 ionization chamber was used for measurements of the FWHM and penumbra at the patient positioning system (PPS) calibration center
The uncertainty components for measurements of the FWHM and penumbra
| Uncertainty component | Type A (%) | Type B (%) |
|---|---|---|
| Optical density measurement | 0.10 | 0.02 |
| Intensity resolution | 0.002 | |
| Uniformity of the radiochromic film | 0.50 | |
| Temperature and relative humidity measurements | 0.056 | |
| Alignment of the PMMA phantom with the PPS calibration center | 0.20 | |
| Positioning of the radiochromic film inside the phantom | 0.16 | |
| Calibration curve fitting | 1.33 | |
| Measurements of FWHM and penumbra | 0.02 | 0.01 |
| Standard relative combined uncertainty | 1.45 | |
Legend: GafChromic™ MD-V3 films were used for the measurement of absorbed dose rate to water at the patient positioning system (PPS) calibration center