OBJECTIVE: To evaluate the mechanical behavior of the proximal end of the femur submitted to the X-shaped polymethylmethacrylate (PMMA) reinforcement technique.Entities:
Keywords: Femoral fractures; Femur; Intertrochanteric fractures; Osteoporosis; Polymethylmethacrylate
Year: 2018 PMID: 30210250 PMCID: PMC6131274 DOI: 10.1590/1413-785220182604187691
Source DB: PubMed Journal: Acta Ortop Bras ISSN: 1413-7852 Impact factor: 0.513
Figure 1Images illustrating the locations and parameters of the entry points of PR I and PR II.
Figure 2Radioscopic images and positioning of the guidewire, at their points of references for the preparation of PR I.
Figure 3Radioscopic images of the positioning of the guidewire, at its points of reference, for the preparation of PR II.
Figure 4Radiographic images of the synthetic model after completely filling the holes.
Figure 5Sample of the control group, fixed onto the machine, demonstrating its positioning. At the mechanical pre-assay moment.
Figure 6Sample of the DX group post-assay, showing the fracture of the femoral neck in the basicervical region.
Description of maximum load and energy to fracture, by group.
| Variable | n | mean | SD | median | IQR | minimum | maximum | p value |
|---|---|---|---|---|---|---|---|---|
|
| ||||||||
| DX | 5 | 1500 | 180 | 1553 | 1367 – 1607 | 1182 | 1627 | 0.005 |
| DP | 10 | 935 | 290 | 905 | 704 – 1174 | 555 | 1399 | |
|
| ||||||||
| DX | 5 | 9.7 | 1.4 | 9.7 | 8.4 – 11.1 | 8.1 | 11.5 | 0.050 |
| DP | 10 | 7.1 | 2.5 | 6.6 | 4.6 – 9.7 | 4.4 | 10.4 | |
SD: Standard deviation; IQR: interquartile range (Q1-Q3). Mann-Whitney test.
Figure 7Maximum load by group (p = 0.005).
Figure 8Energy to fracture, by group (p = 0.05).