| Literature DB >> 30186376 |
Minchao Xiong1, Nian Zhao2, Yan Qin2, Peng Min2, Hong Peng2, Yong Zhang2, Bin Fan3.
Abstract
The safety of magnetic resonance imaging (MRI) scans in patients undergoing dual-chamber pacemaker (Medtronic's EnRhythm MRI SureScan IPG (implantable pulse generator) and CapSureFix MRI pacing electrodes) implantation were evaluated. A total of 86 patients undergoing this pacemaker implantation who were admitted to Huanggang Hospital from June 2006 to March 2017 were continuously selected. On average 6.8±2.3 months after the implantation, the first MRI scan was conducted. The mean scan time was 2.2±0.9 and mean duration was 45.6±12.3 min; mean follow-up visit period was 40.5±15.6 months, and after follow-up, all the abnormal symptoms of patients and pacemaker abnormalities after scans were recorded. Causes of receiving MRI scans included neurological diseases (27%), spinal diseases (14%), cancers (26%), joint injuries (25%) and visceral systems (8%). A total of 12 cases (14.0%) showed significant discomfort symptoms, 10 cases (11.6%) showed pacemaker abnormalities, and the incidence rate was 25.6%. Multivariate logistic regression analysis showed that the basic types of cardiovascular diseases, dependence on pacemakers, duration of education, pacing threshold, impedance and MRI scan time were related to the occurrence of adverse scan results (p<0.05). In conclusion, MRI scans show that MRI-compatible dual-chamber pacemaker was relatively safe.Entities:
Keywords: EnRhythm MRI SureScan pacemaker; MRI; safety
Year: 2018 PMID: 30186376 PMCID: PMC6122164 DOI: 10.3892/etm.2018.6392
Source DB: PubMed Journal: Exp Ther Med ISSN: 1792-0981 Impact factor: 2.447
Figure 1.Analysis of causes of receiving MRI scans.
Analysis of risk factors affecting MRI scans.
| Factors | β | Wald | P-value | OR | 95% CI |
|---|---|---|---|---|---|
| Basic types of cardiovascular diseases | 0.123 | 5.236 | 0.018 | 1.625 | 1.128–2.659 |
| Dependence on pacemakers | 0.285 | 4.625 | 0.029 | 1.147 | 1.007–3.065 |
| Duration of education | −0.547 | 4.121 | 0.033 | 0.522 | 0.147–0.863 |
| Pacemaker threshold | 0.352 | 4.554 | 0.031 | 1.085 | 1.021–2.659 |
| Impedance | 0.427 | 4.329 | 0.035 | 1.072 | 1.011–2.954 |
| MRI scan time | 0.441 | 4.859 | 0.025 | 1.326 | 1.124–2.836 |