| Literature DB >> 30168887 |
Adrian H Heald1,2, Mark Livingston3, Anthony Fryer4, Gabriela Cortes5, Simon G Anderson6, Roger Gadsby7, Ian Laing2, Mark Lunt1, Robert J Young8, Mike Stedman9.
Abstract
AIMS/HYPOTHESIS: Our aim was to quantify the impact of Blood Glucose Monitoring Strips variability (BGMSV) at GP practice level on the variability of reported glycated haemoglobin (HbA1cV) levels.Entities:
Mesh:
Substances:
Year: 2018 PMID: 30168887 PMCID: PMC6766879 DOI: 10.1111/ijcp.13252
Source DB: PubMed Journal: Int J Clin Pract ISSN: 1368-5031 Impact factor: 2.503
Relationship between BGM meter accuracy and expected Blood Glucose taken from Breton and Kovatchev (2010)
| BGM Meter 95% results ± | @ 100 mg/dL blood glucose 95% results within ± |
|---|---|
| 5% | 12 mg/dl |
| 10% | 24 mg/dl |
| 15% | 39 mg/dl |
| 20% | 47 mg/dl |
Blood Glucose Monitoring Strips (BGMS) where accuracy data were available and used
| Strip | GMMG ISO classification | 3‐year quantity | Trials | Samples | >25% | 20%‐25% | 20%‐15% | 15%‐10% | 10%‐5% | <5% | Variability |
|---|---|---|---|---|---|---|---|---|---|---|---|
| Aviva | Yes | 380 170 055 | 19 | 2100 | 0.0% | 0.0% | 0.6% | 5.3% | 27.0% | 67.0% | 5.4% |
| Mobile | Yes | 151 500 204 | 10 | 1200 | 0.0% | 0.0% | 0.1% | 5.2% | 28.0% | 66.8% | 5.3% |
| FreeStyle Optium | Yes | 149 029 926 | 1 | 100 | 0.0% | 0.0% | 1.0% | 9.0% | 30.0% | 60.0% | 6.1% |
| GlucoRx Nexus | Yes | 146 689 898 | 4 | 400 | 0.7% | 4.0% | 4.5% | 12.8% | 33.0% | 45.0% | 9.0% |
| FreeStyle Lite | Yes | 126 726 538 | 8 | 800 | 0.0% | 0.0% | 3.4% | 14.8% | 31.1% | 50.8% | 7.4% |
| Contour Next | Yes | 116 019 267 | 8 | 800 | 0.0% | 0.0% | 0.0% | 1.8% | 24.5% | 73.8% | 4.6% |
| WaveSense JAZZ | No Info | 56 017 572 | 1 | 180 | 0.0% | 1.1% | 5.0% | 16.1% | 32.8% | 45.0% | 8.2% |
| OneTouch Verio | Yes | 35 583 668 | 12 | 1300 | 0.0% | 1.7% | 4.8% | 17.4% | 31.5% | 44.7% | 8.4% |
| TRUEresult | Not Promoted | 32 078 870 | 2 | 200 | 0.0% | 0.0% | 0.0% | 8.0% | 28.0% | 64.0% | 5.7% |
| BGStar | Yes | 26 179,688 | 2 | 360 | 0.0% | 3.1% | 3.9% | 12.5% | 31.4% | 49.2% | 8.2% |
| TRUEyou | Yes | 21 534 310 | 2 | 200 | 0.0% | 0.0% | 1.5% | 12.5% | 31.0% | 55.0% | 6.7% |
| Mylife Pura | Yes | 21 337 759 | 3 | 300 | 0.0% | 4.3% | 14.3% | 21.7% | 25.0% | 34.7% | 10.8% |
| OneTouch Vita | Not Promoted | 19 423 110 | 1 | 100 | 0.0% | 0.0% | 10.0% | 22.0% | 33.0% | 35.0% | 9.3% |
| Active | Yes | 19 314 119 | 5 | 600 | 0.0% | 0.0% | 0.0% | 0.2% | 22.0% | 77.8% | 4.2% |
| Omnitest 3 | No Info | 17 204 404 | 1 | 180 | 0.0% | 5.0% | 3.9% | 12.2% | 31.1% | 47.8% | 8.8% |
| Microdot+ | Excluded | 12 254 656 | 1 | 200 | 0.0% | 1.0% | 5.0% | 13.0% | 38.0% | 43.0% | 8.0% |
| Element | Yes | 8 936 507 | 1 | 100 | 8.0% | 9.0% | 8.0% | 21.0% | 29.0% | 25.0% | 13.5% |
| OneTouch Select Plus | Yes | 1 385 801 | 1 | 100 | 0.0% | 0.0% | 3.0% | 13.0% | 32.0% | 52.0% | 7.1% |
| GlucoRx Original | Not Promoted | 1 272 250 | 2 | 300 | 0.0% | 17.0% | 17.3% | 27.0% | 23.3% | 15.3% | 14.0% |
| Performa | Yes | 1 000 300 | 13 | 1500 | 0.0% | 0.1% | 0.3% | 6.3% | 27.5% | 65.7% | 5.6% |
Figure 1Flow chart describing GP practice selection
Figure 2Overall cumulative mix of strip used over 3 years as a percentage of total use
Figure 3BGMS variability vs % of BG results within specified bands in relation to the reference BG value for ranked by their BGM strip variability (lowest, median and highest decile)
Applying findings to ISO standard and best in class BGMS
| ISO standard accuracy | Best in class accuracy | |
|---|---|---|
| 95% results are within % of reference value | 15% | 7.5% |
| Equivalent meter variability | 7.50% | 3.75% |
| Consequent blood glucose variability | 14% | 7% |
| 95% Blood glucose within % of reference value | 27% | 14% |
| With blood glucose @10 mmol/L difference is | 2.7 mmol/L | 1.4 mmol/L |