| Literature DB >> 30122892 |
Afshin James Khodabakhsh1, John Hofbauer1.
Abstract
PURPOSE: The purpose of the study was to compare phacometrics, patient experience factors and clinical outcomes in patients undergoing bilateral laser-assisted cataract surgery using two common femtosecond laser platforms.Entities:
Keywords: Catalys; LenSx; curved contact lens; femtosecond laser-assisted cataract surgery; liquid optics interface; subconjunctival hemorrhage
Year: 2018 PMID: 30122892 PMCID: PMC6087026 DOI: 10.2147/OPTH.S164513
Source DB: PubMed Journal: Clin Ophthalmol ISSN: 1177-5467
Demographic data
| Demographic characteristics | Mean ± SD | n | % |
|---|---|---|---|
| Age, years | 69.9 ± 8.61 | 50 | 100 |
| Gender | |||
| Female | 31 | 62 | |
| Male | 19 | 38 | |
| Race | |||
| African American | 3 | 6 | |
| Caucasian | 47 | 94 | |
| Laterality | |||
| Right eyes | 29 | 58 | |
| Left eyes | 21 | 42 | |
| First surgical eye | |||
| Catalys | 42 | 84 | |
| LenSx | 8 | 16 | |
| Second surgical eye | |||
| Catalys | 8 | 16 | |
| LenSx | 42 | 84 |
Laser system measurements (mean ± SD)
| System group | Catalys | LenSx | |
|---|---|---|---|
| Patient interface preparation time (seconds) | 140.6 ± 36.13 | 148.7 ± 52.14 | 0.346 |
| Number of docking attempts | 1.3 ± 0.45 | 1.4 ± 0.61 | 0.132 |
| Total suction time (seconds) | 123.2 ± 24.82 | 108.5 ± 34.93 | 0.005 |
Figure 1Incidence of subconjunctival hemorrhage in Catalys and LenSx groups.
Figure 2The patient’s perception of pressure after femtosecond laser fragmentation in Catalys and LenSx groups.
Figure 3Patient’s experience with vision loss after femtosecond laser fragmentation in Catalys and LenSx groups.
Figure 4Completeness of capsulotomy in Catalys and LenSx groups.
Cataract density
| System group | Grade 1 | Grade 2 | Grade 3 | Grade 4 |
|---|---|---|---|---|
| Catalys | 6% | 50% | 42% | 2% |
| LenSx | 10% | 58% | 32% | 0% |
Phacometric measurements (mean ± SD)
| System group | Catalys | LenSx | |
|---|---|---|---|
| Total ultrasound time (seconds) | 50.6 ± 35.94 | 50 ± 41.09 | 0.885 |
| Average phaco power (%) | 6.2 ± 3.98 | 6.4 ± 3.13 | 0.675 |
| Nucleus removal time (seconds) | 208.8 ± 143.31 | 223.1 ± 132.27 | 0.256 |
| Cortical removal time (seconds) | 86.3 ± 36.72 | 106.7 ± 70.9 | 0.01 |
| Effective phaco time (seconds) | 21.4 ± 23.27 | 21 ± 26.19 | 0.896 |
Figure 5Surgeon’s evaluation of the ease of cortex removal during surgery in Catalys and LenSx groups.
Preoperative and postoperative (last follow-up) clinical outcomes (mean ± SD)
| Parameter | Time point | Catalys | LenSx | |
|---|---|---|---|---|
| UDVA | Preoperative | 0.68 ± 0.32 | 0.62 ± 0.30 | 0.046 |
| Postoperative | 0.16 ± 0.21 | 0.10 ± 0.12 | 0.063 | |
| CDVA | Preoperative | 0.28 ± 0.21 | 0.24 ± 0.17 | 0.232 |
| Postoperative | 0.03 ± 0.06 | 0.03 ± 0.05 | 0.687 | |
| MRSE (D) | Preoperative | −0.09 ± 3.45 | 0.03 ± 3.16 | 0.417 |
| Postoperative | −0.16 ± 0.71 | −0.03 ± 0.29 | 0.181 | |
| Flat K (D) | Preoperative | 43.1 ± 1.62 | 43.2 ± 1.48 | 0.498 |
| Postoperative | 43.1 ± 1.60 | 43.1 ± 1.42 | 0.499 | |
| Steep K (D) | Preoperative | 44 ± 1.62 | 44 ± 1.42 | 0.86 |
| Postoperative | 44.1 ± 2.2 | 43.8 ± 1.44 | 0.336 | |
| Pupil size (mm) | Preoperative | 5.02 ± 1.24 | 5.02 ± 1.36 | 1 |
| Postoperative | 4.95 ± 1.21 | 4.99 ± 1.31 | 0.438 | |
| Corneal thickness (μm) | Preoperative | 540.0 ± 35.55 | 539.8 ± 34.85 | 0.887 |
| Postoperative | 540.6 ± 35.17 | 540.6 ± 33.11 | 0.977 | |
| Axial length (mm) | Preoperative | 23.98 ± 1.42 | 23.98 ± 1.33 | 0.986 |
| Postoperative | 23.86 ± 1.36 | 23.93 ± 1.31 | 0.156 | |
| IOP (mmHg) | Preoperative | 14.8 ± 2.92 | 14.8 ± 3.08 | 0.888 |
| Postoperative | 13.7 ± 2.48 | 13.6 ± 2.21 | 0.574 | |
| ECD (cells/mm2) | Preoperative | 2,705.9 ± 286.11 | 2,727.8 ± 268.17 | 0.396 |
| Postoperative | 2,689.4 ± 266.5 | 2,711.1 ± 257.88 | 0.358 |
Abbreviations: UDVA, uncorrected distance visual acuity; CDVA, corrected distance visual acuity; MRSE, manifest refraction spherical equivalent; K, keratometry; IOP, intraocular pressure; ECD, endothelial cell density.