| Literature DB >> 30081480 |
Maren Wierig1, Leonard P Mandtler2, Peter Rottmann3, Viktor Stroh4, Ute Müller5, Wolfgang Büscher6, Lutz Plümer7,8.
Abstract
In the last decades, there has been an increasing interest in animal protection and welfare issues. Heart rate variability (HRV) measurement with portable heart rate monitors on cows has established itself as a suitable method for assessing physiological states. However, more forward-looking technologies, already successfully applied to evaluate HRV data, are pushing the market. This study examines the validity and usability of collecting HRV data by exchanging the Polar watch V800 as a receiving unit of the data compared to a custom smartphone application on cows. Therefore, both receivers tap one signal sent by the Polar H7 transmitter simultaneously. Furthermore, there is a lack of suitable methods for the preparation and calculation of HRV parameters, especially for livestock. A method is presented for calculating more robust time domain HRV parameters via median formation. The comparisons of the respective simultaneous recordings were conducted after artifact correction for time domain HRV parameters. High correlations (r = 0.82⁻0.98) for cows as well as for control data set in human being (r = 0.98⁻0.99) were found. The utilization of smart devices and the robust method to determine time domain HRV parameters may be suitable to generate valid HRV data on cows in field-based settings.Entities:
Keywords: cardiac parasympathetic; cattle; heart rate; mobile sensing; monitoring; smart device; validation; welfare
Mesh:
Year: 2018 PMID: 30081480 PMCID: PMC6111714 DOI: 10.3390/s18082541
Source DB: PubMed Journal: Sensors (Basel) ISSN: 1424-8220 Impact factor: 3.576
Figure 1Dairy Cow I equipped with the Polar equestrian chest belt and Polar H7 receiver; red circles mark the placement of the electrodes.
Figure 2Schematic diagram of the experimental setup with Polar H7 transmitter, receiver smartphone IoTool application (left) and receiver Polar watch V800 (right).
Figure 3Exemplary comparison of RR interval (R-to-R) (ms) measurements prepared with mean vs. median over 60 s (n = 72 measured values).
Figure 4Comparison between the mean RR intervals (R-to-R) (ms) of the two signals in the raw data set derived from the smartphone IoTool application and the Polar watch V800 for Cow I over the experimental period of 27 min (n = 3630 measured values).
Figure 5Comparison between the mean RR intervals (R-to-R) (ms) of the two signals in the data set derived from the smartphone IoTool application and the Polar watch V800 for Cow I over the experimental period of 27 min after artifact correction (n = 3630 measured values).
Figure 6Extract of simultaneous RR interval (R-to-R) (ms) measurements for Cow I during 60 s of recording after outlier removal and but before interpolation acquired with the smartphone IoTool application and Polar watch V800.
Figure 7Simultaneous RR interval (R-to-R) (ms) measurements for Cow I during 27 min of recording after outlier removal and interpolation acquired with smartphone IoTool application and Polar watch V800 (r = 0.98; p < 0.001).
Figure 8Relationships between RR intervals (R-to-R) (ms) derived from the smartphone IoTool application and Polar watch V800 for measurements of Cow I and Cow II.
Figure 9Relationships between RR intervals (R-to-R) (ms) derived from the smartphone IoTool application and Polar watch V800 for measurements on human.
Summary statistics of heart rate variability (HRV) parameters (mean ± SD) and validity statistics (correlation) measured via Polar watch V800 and smartphone IoTool application.
| Smartphone IoTool | Polar V800 Mean | Spearman Correlation | |
|---|---|---|---|
| Dairy Cow I | |||
| RR interval (ms) | 819.08 ± 21.96 | 818.08 ± 22.91 | 0.98 * (<0.001) |
| SDNN (ms) | 15.36 ± 8.31 | 14.67 ± 8.54 | 0.89 † (<0.001) |
| rMSSD (ms) | 7.82 ± 1.86 | 7.41 ± 1.77 | 0.82 † (<0.001) |
| Dairy Cow II | |||
| RR interval (ms) | 899.84 ± 42.03 | 897.41 ± 44.91 | 0.98 * (<0.001) |
| SDNN (ms) | 26.86 ± 15.15 | 26.33 ± 15.30 | 0.97 * (<0.001) |
| rMSSD (ms) | 10.15 ± 2.22 | 9.34 ± 1.96 | 0.88 † (<0.001) |
| Human control data | |||
| RR interval (ms) | 1326.13 ± 123.08 | 1324.68 ± 125.86 | 0.99 * (<0.001) |
| SDNN (ms) | 88.07 ± 43.73 | 88.50 ± 44.48 | 1.00 * (<0.001) |
| rMSSD (ms) | 73.63 ± 12.73 | 73.59 ± 12.63 | 1.00 * (<0.001) |
RR interval: distance between successive R waves; SDNN: standard deviation of all RR intervals; rMSSD: Square root of the mean of the sum of all differences between adjacent RR intervals; * good validity; † moderate validity.
Percentage relative errors of heart rate variability (HRV) parameter measurements.
| Error % | Smartphone IoTool vs. Polar V800 |
|---|---|
| Dairy Cow I | |
| RR interval (ms) | 0.12 |
| SDNN (ms) | 4.70 |
| rMSSD (ms) | 5.53 |
| Dairy Cow II | |
| RR interval (ms) | 0.27 |
| SDNN (ms) | 2.01 |
| rMSSD (ms) | 8.67 |
| Human control data | |
| RR interval (ms) | 0.11 |
| SDNN (ms) | 0.49 |
| rMSSD (ms) | 0.05 |
RR interval: distance between successive R waves; SDNN: standard deviation of all RR intervals; rMSSD: Square root of the mean of the sum of all differences between adjacent RR intervals.