Literature DB >> 19092682

Validity and reliability of short-term heart-rate variability from the Polar S810.

David Nunan1, Gay Donovan, Djordje G Jakovljevic, Lynette D Hodges, Gavin R H Sandercock, David A Brodie.   

Abstract

PURPOSE: : To assess the validity and the reliability of short-term resting heart-rate variability (HRV) measures obtained using the Polar S810 heart-rate monitor and accompanying software.
METHODS: : Measures of HRV were obtained from 5-min R to R wave (RR) interval data for 19 males and 14 females during 10 min of quiet rest on three separate occasions at 1-wk intervals using the Polar S810. Criterion measures of HRV were obtained simultaneously using the CardioPerfect (CP; Medical Graphics Corporation, St Paul, MN) 12-lead ECG module. Measures of validity of the Polar S810 were estimated by regression analysis, and measures of reliability of both devices were estimated by analysis of change scores. Measures of the SD of normal-to-normal intervals (SDNN), the root mean square of successive differences (RMSSD), and the low-frequency (LF) and the high-frequency (HF) spectral power and their ratio (LF/HF) were analyzed after log transformation, whereas mean RR and LF and HF in normalized units were analyzed without transformation.
RESULTS: : There were marginal differences between the Polar and the CP mean measures of HRV, and the uncertainty in the differences was small. The Polar S810 demonstrated high correlations (0.85-0.99) with CP for all measures of HRV indicating good to near-perfect validity. Except for the low- and the high-frequency normalized units, Polar S810 did not add any substantial technical error to the within-subject variability in the repeated measurements of HRV.
CONCLUSION: : HRV measures obtained with the Polar S810 and accompanying software have no appreciable bias or additional random error in comparison with criterion measures, but the measures are inherently unreliable over a 1-wk interval. Reliability of HRV from longer (e.g., 10 min) and/or consecutive 5-min RR recordings needs to be investigated with the Polar and criterion instruments.

Mesh:

Year:  2009        PMID: 19092682     DOI: 10.1249/MSS.0b013e318184a4b1

Source DB:  PubMed          Journal:  Med Sci Sports Exerc        ISSN: 0195-9131            Impact factor:   5.411


  70 in total

1.  Effect of cold water immersion on 100-m sprint performance in well-trained swimmers.

Authors:  Jonathan Parouty; Hani Al Haddad; Marc Quod; Pierre Marie Leprêtre; Said Ahmaidi; Martin Buchheit
Journal:  Eur J Appl Physiol       Date:  2010-02-17       Impact factor: 3.078

2.  Trans-fatty acid consumption and heart rate variability in 2 separate cohorts of older and younger adults.

Authors:  Luisa Soares-Miranda; Phyllis K Stein; Fumiaki Imamura; Jacob Sattelmair; Rozenn N Lemaitre; David S Siscovick; Jorge Mota; Dariush Mozaffarian
Journal:  Circ Arrhythm Electrophysiol       Date:  2012-07-06

3.  Assessment of heart rate variability by application of central tendency measure.

Authors:  Laurita dos Santos; Joaquim J Barroso; Elbert E N Macau; Moacir F de Godoy
Journal:  Med Biol Eng Comput       Date:  2015-09-22       Impact factor: 2.602

Review 4.  [Options and limitations of heart rate measurement and analysis of heart rate variability by mobile devices: A systematic review].

Authors:  Stefan Sammito; Irina Böckelmann
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2016-02-10

5.  Heart rate variability before and after cycle exercise in relation to different body positions.

Authors:  Otto F Barak; Djordje G Jakovljevic; Jelena Z Popadic Gacesa; Zoran B Ovcin; David A Brodie; Nikola G Grujic
Journal:  J Sports Sci Med       Date:  2010-06-01       Impact factor: 2.988

6.  Bioharness(™) Multivariable Monitoring Device: Part. II: Reliability.

Authors:  James A Johnstone; Paul A Ford; Gerwyn Hughes; Tim Watson; Andrew T Garrett
Journal:  J Sports Sci Med       Date:  2012-09-01       Impact factor: 2.988

7.  Monitoring endurance running performance using cardiac parasympathetic function.

Authors:  Martin Buchheit; A Chivot; J Parouty; D Mercier; H Al Haddad; P B Laursen; S Ahmaidi
Journal:  Eur J Appl Physiol       Date:  2009-12-22       Impact factor: 3.078

8.  Cardiac autonomic modulation impairments in advanced breast cancer patients.

Authors:  Claudia Arab; Luiz Carlos Marques Vanderlei; Laércio da Silva Paiva; Kyle Levi Fulghum; Carlos Elias Fristachi; Afonso Celso Pinto Nazario; Simone Elias; Luiz Henrique Gebrim; Celso Ferreira Filho; Yori Gidron; Celso Ferreira
Journal:  Clin Res Cardiol       Date:  2018-05-02       Impact factor: 5.460

9.  A church-based intervention to change attitudes about physical activity among Black adolescent girls: a feasibility study.

Authors:  Wanda M Thompson; Diane Berry; Jie Hu
Journal:  Public Health Nurs       Date:  2012-10-05       Impact factor: 1.462

10.  Atypical pupillary light reflex and heart rate variability in children with autism spectrum disorder.

Authors:  Chathuri Daluwatte; Judith H Miles; Shawn E Christ; David Q Beversdorf; T Nicole Takahashi; Gang Yao
Journal:  J Autism Dev Disord       Date:  2013-08
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.