Determination of hemolysis thresholds by the use of data loggers in pneumatic tube systems.

Thomas Streichert1, Benjamin Otto, Claudia Schnabel, Gerhard Nordholt, Munif Haddad, Mario Maric, Astrid Petersmann, Roman Jung, Christoph Wagener.   

Abstract

BACKGROUND: Pneumatic tube systems (PTSs) for the transport of blood samples are regaining popularity in medical centers after earlier reports that their use could introduce preanalytical distortions such as hemolysis and changes in blood gases.
METHODS: We drew duplicate blood samples from 30 volunteers. One sample was hand transported, and the other sample was transported through a PTS together with a mini-data logger that provided continuous measurements of temperature, humidity, pressure, and acceleration. After transport the samples were analyzed at the same time. We looked for possible relationships of the transport method and the parameters measured by the data loggers with differences in hematological parameters, standard clinical chemistry analyses, blood coagulation, erythrocyte sedimentation rate, and blood gas analysis.
RESULTS: There were no significant differences in temperature, humidity, and pressure between the methods of transport, but we observed significant differences in 3-axis accelerations. The combined effect of these forces could be described by the right-tailed area under the vector sum acceleration distribution. Our data show that this area correlated with PTS speed and that PTS speed and the area under the curve exhibited a direct relation to the degree of hemolysis.
CONCLUSIONS: Assessment of 3-axis acceleration by use of data loggers can be used to identify preanalytical deviations that result from the transportation of blood samples in PTSs. Our approach could be used for the evaluation and regular control of PTSs without the need for repeated blood drawing and laboratory analyses.

Entities:  

Mesh:

Year:  2011        PMID: 21836074     DOI: 10.1373/clinchem.2011.167932

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  9 in total

1.  Pseudohyperkalaemia associated with leukaemic cell lysis during pneumatic tube transport of blood samples.

Authors:  Helen Dickinson; Nicholas J A Webb; Christopher Chaloner; Robert F Wynn; Denise K Bonney
Journal:  Pediatr Nephrol       Date:  2012-02-26       Impact factor: 3.714

2.  Stability of Routine Biochemical Analytes in Whole Blood and Plasma From Lithium Heparin Gel Tubes During 6-hr Storage.

Authors:  Denis Monneret; Alexandre Godmer; Ronan Le Guen; Clotilde Bravetti; Cecile Emeraud; Anthony Marteau; Rana Alkouri; Fouzi Mestari; Sylvie Dever; Françoise Imbert-Bismut; Dominique Bonnefont-Rousselot
Journal:  J Clin Lab Anal       Date:  2016-02-18       Impact factor: 2.352

3.  Effect of Pneumatic Tubing System Transport on Platelet Apheresis Units.

Authors:  Jevgenia Zilberman-Rudenko; Frank Z Zhao; Stephanie E Reitsma; Annachiara Mitrugno; Jiaqing Pang; Joseph J Shatzel; Beth Rick; Christina Tyrrell; Wohaib Hasan; Owen J T McCarty; Martin A Schreiber
Journal:  Cardiovasc Eng Technol       Date:  2018-05-21       Impact factor: 2.495

4.  Pathologic Blood Samples Tolerate Exposure to Vibration and High Turbulence in Simulated Drone Flights, but Plasma Samples Should be Centrifuged After Flight.

Authors:  Karl Arne Johannessen; Nina Kristin Steen Wear; Karin Toska; Morten Hansbo; Jens Petter Berg; Erik Fosse
Journal:  IEEE J Transl Eng Health Med       Date:  2021-01-20       Impact factor: 3.316

5.  Effects of one directional pneumatic tube system on routine hematology and chemistry parameters; A validation study at a tertiary care hospital.

Authors:  A-Jin Lee; Hun Suk Suh; Chang-Ho Jeon; Sang-Gyung Kim
Journal:  Pract Lab Med       Date:  2017-07-19

Review 6.  Hemolyzed Specimens: Major Challenge for Identifying and Rejecting Specimens in Clinical Laboratories.

Authors:  Wan Norlina Wan Azman; Julia Omar; Tan Say Koon; Tuan Salwani Tuan Ismail
Journal:  Oman Med J       Date:  2019-03

Review 7.  Preanalytical considerations in blood gas analysis.

Authors:  Geoffrey Baird
Journal:  Biochem Med (Zagreb)       Date:  2013       Impact factor: 2.313

8.  The Effects of Sample Transport by Pneumatic Tube System on Routine Hematology and Coagulation Tests.

Authors:  Devi Subbarayan; Chidambharam Choccalingam; Chittode Kodumudi Anantha Lakshmi
Journal:  Adv Hematol       Date:  2018-07-02

9.  The incidence rate and influence factors of hemolysis, lipemia, icterus in fasting serum biochemistry specimens.

Authors:  Gang Tian; Yu Wu; Xinrui Jin; Zhangrui Zeng; Xiujuan Gu; Tao Li; Xiu Chen; Guangrong Li; Jinbo Liu
Journal:  PLoS One       Date:  2022-01-19       Impact factor: 3.240

  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.