Ismail AlDahlawi1,2,3, Dheerendra Prasad1,4,5,6, Matthew B Podgorsak1,2,5.
Abstract
INTRODUCTION: The Gamma Knife® Icon™ comes with an image guidance system for tracking patient motion and correcting for inter- and intrafractional shifts, mainly used with frameless thermoplastic immobilization. The system consists of a cone-beam CT (CBCT) and a couch-mounted infrared camera (IFMM). We report our quality assurance program for Icon's image guidance system.Entities:
Keywords: Gamma knife; Icon; QA tool; motion management; stereotactic radiosurgery
Mesh:
Year: 2018 PMID: 30076672 PMCID: PMC6123129 DOI: 10.1002/acm2.12417
Source DB: PubMed Journal: J Appl Clin Med Phys ISSN: 1526-9914 Impact factor: 2.102
Figure 1CBCT‐based and Frame‐based stereotactic space agreement tool setup at (a) Gamma Knife Icon for defining the CBCT‐based stereotactic coordinates of five BBs hanged diagonally on the stereotactic frame; and (b) CT‐simulator with the indicator box for defining the Frame‐based stereotactic coordinates for the same stereotactic frame with BBs.
Figure 2(a) A picture of the linear translational stage tool with the added rotary component. A Lucite platform is attached to the transitional stage with a post with the IR reflective marker to check the IFMM accuracy, and two film holders to check delivery‐after‐shift accuracy; and (b) The same tool setup on Icon's couch. The head mask adapter which holds reference markers is attached to the couch to activate the IFMM and registration with plan adaptive functions in the Icon system.
Figure 3Maximum deviation (mm) in image volume of the daily CBCT precision test over 2‐yr period. The red line represents the manufacturer limit for the test to pass (0.4 mm).
Summary results of monthly image quality tests (spatial resolutions, CNR, and uniformity) with the manufacturer specification for each test
| [Range] or average ± SD | |||
|---|---|---|---|
| CTDI 2.5 mGy (low‐dose preset) | CTDI 6.3 mGy (high‐quality preset) | Manufacturer specification | |
| Spatial resolution | [6,7] lp/cm | [7,8] lp/cm | Minimum 6 lp/cm |
| CNR | 1.07 ± 0.08 | 1.69 ± 0.10 |
>0.5 (CTDI 2.5) |
| Uniformity | 12.84% ± 0.70 | 13.01% ± 0.69 | <14% |
Coordinates of each marker in Frame‐based stereotactic space and CBCT‐based stereotactic space and the differences between each definition as established in the agreement of CBCT‐based with Frame‐based stereotactic space test. The magnitude of vector difference is calculated as
| Displacement (mm) in | ||||||||
|---|---|---|---|---|---|---|---|---|
| Marker ID | Coordinates definition |
|
|
|
|
|
|
|
| BB1 | Frame‐based | 78.2 | 141.4 | 68.2 | 0.4 | 0 | ‐0.4 | 0.57 |
| CBCT‐based | 77.8 | 141.4 | 68.6 | |||||
| BB2 | Frame‐based | 101 | 107.1 | 83.2 | 0.3 | 0 | ‐0.3 | 0.42 |
| CBCT‐based | 100.7 | 107.1 | 83.5 | |||||
| BB3 | Frame‐based | 120 | 78 | 97.4 | 0.3 | −0.1 | 0.1 | 0.33 |
| CBCT‐based | 119.7 | 78.1 | 97.3 | |||||
| BB4 | Frame‐based | 136.8 | 53.1 | 109.6 | 0.3 | −0.3 | 0.3 | 0.52 |
| CBCT‐based | 136.5 | 53.4 | 109.3 | |||||
| BB5 | Frame‐based | 154.1 | 24.5 | 122.8 | 0.3 | −0.3 | 0.5 | 0.66 |
| CBCT‐based | 153.8 | 24.8 | 122.3 | |||||
Summary results of accuracy tests of the Intra‐fraction Motion Management system (IFMM), registration algorithm, and delivery‐after‐shift to a film. The results for IFMM and registration (3D stage method) and the shot measurements are averaged over small (2 mm in each of x, y, and z direction) and large (5 mm in x and y directions, 20 mm in z direction, and 1 degree rotation along y) shift test scenarios. The difference is calculated as measured value minus expected value
| Test | Average difference ± SD (mm) | [Range] (mm) | |
|---|---|---|---|
| IFMM accuracy | (3D stage method) | 0.00 ± 0.12 | [−0.27, 0.23] |
| (Transformation matrix method) | 0.08 ± 0.08 | [−0.07, 0.15] | |
| Registration accuracy | (3D stage method) | 0.03 ± 0.06 | [−0.06, 0.23] |
| (Transformation matrix method) | −0.01 ± 0.13 | [−0.48, 0.21] | |
| Delivery‐after‐shift accuracy | 0.24 ± 0.09 | [0.08, 0.42] | |