| Literature DB >> 30075796 |
I Kagan1, O Zusman2, I Bendavid3, M Theilla3,4, J Cohen3, P Singer3.
Abstract
BACKGROUND: Indirect calorimetry (IC) measurement is considered the gold standard for the assessment of resting energy expenditure (REE). It is based on the measurement of oxygen and carbon dioxide consumption (VO2 and VCO2, respectively). However, its use is limited by cost and technical issues. It has been proposed that, in critically ill patients, the analysis of VCO2 obtained from the ventilator alone may be used as an accurate method to assess REE in ventilated patients. This retrospective study aimed to assess the accuracy of VCO2 measurement alone in the determination of REE.Entities:
Keywords: Abbreviations; Carbon dioxide production (VCO2); IC Indirect calorimetry; Indirect calorimetry; Oxygen consumption (VO2); REE Resting energy expenditure; RQ Respiratory quotient; Resting energy expenditure (REE); VCO2 Carbon dioxide production; VO2 Oxygen consumption
Mesh:
Substances:
Year: 2018 PMID: 30075796 PMCID: PMC6091032 DOI: 10.1186/s13054-018-2108-8
Source DB: PubMed Journal: Crit Care ISSN: 1364-8535 Impact factor: 9.097
Patient characteristics
| Mean | Standard deviation | |
|---|---|---|
| Sex | 53% male, 47% female | – |
| Age, years | 55.6 | 22.5 |
| SOFA score | 6.6 | 3.5 |
| APACHE II score | 22.5 | 7.0 |
| Weight, kg | 82.4 | 21.7 |
| Height, m | 1.70 | 0.10 |
| Body mass index, kg/m2 | 28.7 | 8.7 |
| Enteral energy delivered, kcal/d | 1278 | 654 |
| Parenteral energy delivered, kcal/d | 333 | 439 |
Abbreviations: APACHE Acute Physiology and Chronic Health Evaluation; SOFA Sequential Organ Failure Assessment
The VCO2 performance
| Mean difference | Standard deviation | % Error | % Difference | Correlation | Concordance | Agreementa | Tight agreementb | |
|---|---|---|---|---|---|---|---|---|
| VCO2 with 0.89 RQc | − 122.78 | 604.92 | 0.24 | 0.16 | 0.51 (0.44–0.57) | 0.37 (0.27–0.47) | 0.37 | 0.11 |
| VCO2 with 0.85 RQ | −41.77 | 625.57 | 0.24 | 0.16 | 0.51 (0.44–0.57) | 0.37 (0.27–0.47) | 0.4 | 0.13 |
| VCO2 with 0.80 RQ | 62.54 | 652.95 | 0.25 | 0.16 | 0.51 (0.44–0.57) | 0.37 (0.26–0.47) | 0.42 | 0.16 |
| VCO2 with 0.75 RQ | 178.17 | 684.37 | 0.26 | 0.16 | 0.51(0.44–0.57) | 0.35(0.25–0.45) | 0.43 | 0.18 |
aWithin 85% and 115% of measured REE
bWithin 95% and 105% of measured REE
cEquivalent to REE calculation of 5.5*VCO2*1.44
Abbreviations: REE resting energy expenditure, RQ respiratory quotient, VCO carbon dioxide production
Fig. 1Bland–Altman graph representing resting energy expenditure derived from VCO2 (REE- VCO2) in comparison with REE obtained from indirect calorimetry (REE-IC)
Fig. 2Correlation between resting energy expenditure derived from VCO2 (REE-VCO2) and REE calculated from indirect calorimetry measurements (REE-IC) presented as a scatter plot