| Literature DB >> 30046929 |
Angharad E Hobby1,2,3, Diana Kozareva1,3, Ekaterina Yonova-Doing1,3, Ibtesham T Hossain1, Mohamed Katta1, Byki Huntjens2, Christopher J Hammond1,3, Alison M Binns2, Omar A Mahroo4,5,6,7,8.
Abstract
PURPOSE: A handheld device (the RETeval system, LKC Technologies) aims to increase the ease of electroretinogram (ERG) recording by using specially designed skin electrodes, rather than corneal electrodes. We explored effects of electrode position on response parameters recorded using this device.Entities:
Keywords: Electrode; Electroretinogram; Retina; Retinal function
Mesh:
Year: 2018 PMID: 30046929 PMCID: PMC6153519 DOI: 10.1007/s10633-018-9652-z
Source DB: PubMed Journal: Doc Ophthalmol ISSN: 0012-4486 Impact factor: 2.379
Fig. 1Two different electrode positions in a study participant. The skin sensor electrodes are placed 2 mm below the lid margin (manufacturer’s recommended position) in the upper photograph, and in an inferior position 10–20 mm below lid margin (lower photograph). Permission granted for use of photograph
Parameter values compared between the two age-matched groups
| Group 1 (“comfortable” electrode position) | Group 2 (recommended electrode position) | ||||
|---|---|---|---|---|---|
| Mean (SD) | Median (LQ, UQ) | Mean (SD) | Median (LQ, UQ) | ||
| Flicker ERG amplitude (microvolts) | 21.0 (12.0) | 18.6 (10.7, 28.7) | 30.6 (9.7) | 31.7 (23.3, 37.9) | 0.0024* |
| Flicker ERG peak time (ms) | 25.9 (1.4) | 25.5 (25.0, 26.6) | 25.9 (0.9) | 26.0 (25.2, 26.7) | 0.54 |
*p < 0.05 regarded as significant
Fig. 2Boxplots comparing the distribution of ERG parameters between participants in Groups 1 and 2. In Group 1, the electrode was in a comfortable position (up to 20 mm from the lid margin); in Group 2, the electrode was placed in the recommended position of 2 mm below the lid margin. Median, upper and lower quartiles (limits of box) are shown. Squares plot means; whiskers span the 5th–95th centiles; and crosses plot minimum and maximum data points. Amplitudes are plotted in a and peak times in b
Fig. 3Sample traces from one participant. a, b Averaged light-adapted ERGs to a flash obtained from right and left eyes, respectively. In the right eye, the skin electrode was placed 2 mm from the lid margin. In the left eye, the electrode was 15 mm below the lid margin. c, d Light-adapted responses to the flicker stimulus obtained with electrodes placed as in a, b, respectively. Green and orange traces represent averaged recordings obtained from consecutive runs of stimulus presentations
Fig. 4ERG parameters for different electrode positions in participants in the third group (n = 10; data normalised to control eye). a, b, c Boxplots showing a-wave, b-wave and flicker ERG amplitudes: differences were significant between the two positions (p = 0.004, 0.002 and 0.002, respectively, Wilcoxon signed rank test). d, e, f, Boxplots showing a-wave, b-wave and flicker ERG peak times: differences were not significant (p = 0.85, 0.57 and 0.56, respectively). Boxes show median and upper and lower quartiles; squares plot mean values; whiskers extend to maximum and minimum values. Note that the y-axis scales differ in the lower panels and have been expanded considerably in F, where the actual range between maximum and minimum values is less than 0.1