Comparison of two commercial quality control sera for adrenocorticotropin (ACTH) used in Elecsys® immunoassay system.

Zhi-Qi Wu1, Hua-Guo Xu1.   

Abstract

OBJECTIVES: The purpose of our study was to investigate whether the storage time and temperature of internal quality control (IQC) material influence the result of ACTH in IQC measurements. DESIGN AND METHODS: Five levels of IQC materials from two manufacturers were tested through the precision of ACTH, the three freeze/thaw cycles, and the storage time and temperature to evaluate the stability of IQC material. All commercial control materials were simultaneously tested three times a day for five consecutive days.
RESULTS: Total precision of three levels of Bio-Rad IQC sera was 13.93%, 16.45%, and 15.98%, respectively, but repeatability was <2%. The concentration of ACTH decreased by 30%-50% after 3 freeze/thaw cycles. At room temperature, the concentration of ACTH from 3 levels decreased by 16.60%, 17.98%, and 17.20%, respectively, after 0.5 hours, and 70.54%, 74.36%, and 72.03%, respectively, after 4 hours. However, after 2 hours of storage at 4°C, the decline in the measured ACTH IQC was 8.04%, 11.84%, and 10.11%, respectively. Total precision of Roche IQC was 1.17% and 1.08%, respectively. After 3 freeze/thaw cycles, the concentration of ACTH decreased <5%. After 4 hours, the change of ACTH still steadied within 5% both at the room temperature and at 4°C.
CONCLUSION: Roche is a better choice for ACTH of IQC material in Elecsys® immunoassay system in our study. If Bio-Rad control materials be used in Elecsys® immunoassay system for ACTH IQC testing material, it should be stored at 4°C and testing should be completed within 1 hours.
© 2018 Wiley Periodicals, Inc.

Entities:  

Keywords:  adrenocorticotropin; internal quality control

Mesh:

Substances:

Year:  2018        PMID: 30006935      PMCID: PMC6430340          DOI: 10.1002/jcla.22618

Source DB:  PubMed          Journal:  J Clin Lab Anal        ISSN: 0887-8013            Impact factor:   2.352


  7 in total

1.  Preanalytical stability of adrenocorticotropic hormone depends on time to centrifugation rather than temperature.

Authors:  Nicole Reisch; Martin Reincke; Martin Bidlingmaier
Journal:  Clin Chem       Date:  2007-02       Impact factor: 8.327

Review 2.  Internal quality control: best practice.

Authors:  Helen Kinns; Sarah Pitkin; David Housley; Danielle B Freedman
Journal:  J Clin Pathol       Date:  2013-09-26       Impact factor: 3.411

3.  Comparison of two commercial quality control sera for adrenocorticotropin (ACTH) used in Elecsys® immunoassay system.

Authors:  Zhi-Qi Wu; Hua-Guo Xu
Journal:  J Clin Lab Anal       Date:  2018-07-13       Impact factor: 2.352

4.  Preanalytical stability of adrenocorticotropic hormone depends on both time to centrifugation and temperature.

Authors:  Zhi-Qi Wu; Hua-Guo Xu
Journal:  J Clin Lab Anal       Date:  2016-10-13       Impact factor: 2.352

Review 5.  IQC Laboratory management strategy for medical biology.

Authors:  Estelle Bugni; Richard Cohen; Cécile Mazellier
Journal:  Ann Biol Clin (Paris)       Date:  2017-12-01       Impact factor: 0.459

6.  Differential interferences of hemoglobin and hemolysis on insulin assay with the Abbott Architect-Ci8200 immunoassay.

Authors:  Simon Garinet; Soraya Fellahi; Ginette Marlin; Jacqueline Capeau; Guillaume Lefèvre; Jean-Philippe Bastard
Journal:  Clin Biochem       Date:  2014-01-31       Impact factor: 3.281

7.  Efficacy of dexamethasone suppression test during the diagnosis of primary pigmented nodular adrenocortical disease in Chinese adrenocorticotropic hormone-independent Cushing syndrome.

Authors:  Shi Chen; Ran Li; Lin Lu; Lian Duan; Xuebin Zhang; Anli Tong; Hui Pan; Huijuan Zhu; Zhaolin Lu
Journal:  Endocrine       Date:  2017-11-01       Impact factor: 3.633

  7 in total
  2 in total

1.  Comparison of two commercial quality control sera for adrenocorticotropin (ACTH) used in Elecsys® immunoassay system.

Authors:  Zhi-Qi Wu; Hua-Guo Xu
Journal:  J Clin Lab Anal       Date:  2018-07-13       Impact factor: 2.352

2.  Establishment and Evaluation of an Internal Quality Control Method for ESR Relays.

Authors:  Shui Fu; Liang Zhang; Qi-Lei Hu; Zuo-Jie Li
Journal:  Int J Gen Med       Date:  2022-03-22
  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.