Agreement and repeatability of central corneal thickness measurements by four different optical devices and an ultrasound pachymeter.

Nesrin Buyuktortop Gokcinar1, Erhan Yumusak2, Nurgul Ornek2, Serap Yorubulut3, Zafer Onaran2.   

Abstract

PURPOSE: To compare the repeatability and agreement of central corneal thickness (CCT) measurements by spectral-domain optical coherence tomography (OCT), corneal topography (CT) with a combined Scheimpflug-Placido system, optical biometry (OB), specular microscopy (SM), and ultrasound pachymetry (UP).
METHODS: A single observer measured CCT twice in 150 eyes of 150 subjects with each of five devices: Nidek RS-3000 Advance OCT, CSO Sirius combined Scheimpflug-Placido disc system CT, Nidek AL-Scan partial coherence interferometry-based OB, Tomey EM-3000 SM, and Reichert iPac ultrasonic pachymeter. Pachymetry values corrected by the SM device software were also recorded. Levels of agreement between devices were evaluated by Bland-Altman plots with 95% limits of agreement, and repeatability for each device was analysed with intraclass correlation coefficients.
RESULTS: The mean CCTs measured by OCT, CT, OB, SM, corrected SM, and UP were 544.60 ± 29.56, 536.19 ± 32.14, 528.29 ± 29.45, 524.88 ± 32.38, 537.88 ± 32.38, and 545.29 ± 30.75 μm, respectively. Mean CCT differed significantly between the devices (p < 0.05) apart from between OCT and UP, and between CT and corrected SM. Mean paired differences between devices ranged between 0.68 and 20.41 μm. Repeatability with all devices was excellent (> 0.99). The range of limits of agreement was the least between OCT and UP.
CONCLUSIONS: Different CCT measurement techniques produce quite different results, so CCT evaluation and follow-up should be performed using the same device or devices with close compatibility.

Entities:  

Keywords:  Central corneal thickness; Corneal topography; Optic biometry; Optical coherence tomography; Specular microscopy; Ultrasonic pachymeter

Mesh:

Year:  2018        PMID: 29984376     DOI: 10.1007/s10792-018-0983-2

Source DB:  PubMed          Journal:  Int Ophthalmol        ISSN: 0165-5701            Impact factor:   2.031


  7 in total

1.  Impact of corneal parameters on intraocular pressure measurements in different tonometry methods.

Authors:  Aleksandra Zakrzewska; Marta P Wiącek; Anna Machalińska
Journal:  Int J Ophthalmol       Date:  2019-12-18       Impact factor: 1.779

2.  Sirius Scheimpflug-Placido versus ultrasound pachymetry for central corneal thickness: meta-analysis.

Authors:  Yili Jin; Colm McAlinden; Yong Sun; Daizong Wen; Yiran Wang; Jinjin Yu; Ke Feng; Benhao Song; Qinmei Wang; Shihao Chen; Jinhai Huang
Journal:  Eye Vis (Lond)       Date:  2021-02-18

3.  Comparison of central corneal thickness measurements with three different optical devices.

Authors:  Mehmet Barış Üçer; Erdinç Bozkurt
Journal:  Ther Adv Ophthalmol       Date:  2021-02-23

4.  Anterior Segment Optic Coherence Tomography Changes Before and After Phacoemulsification in Primary Open-Angle Glaucoma.

Authors:  Sibel Zirtiloglu; Ozen Ayranci Osmanbasoglu; Tulay Alpar Akcetin; Yeliz Acar; Mustafa Suat Alikma; Erkan Unsal; Kadir Eltutar
Journal:  Beyoglu Eye J       Date:  2019-08-09

5.  Comparison of Six Methods of Central Corneal Thickness Measurement in Healthy Eyes.

Authors:  Turgay Ucak; Erel Icel; Nurdan Gamze Tasli; Yucel Karakurt; Hayati Yilmaz; Adem Ugurlu; Mehmet Demir
Journal:  Beyoglu Eye J       Date:  2021-02-12

6.  Repeatability and Interchangeability of Topometric, Anterior Chamber and Corneal Wavefront Data Between Two Scheimpflug Camera Devices.

Authors:  Hesham Mohamed Gharieb; Hisham Samy Shalaby; Ihab Saad Othman
Journal:  Clin Ophthalmol       Date:  2020-11-05

7.  Evaluation of 6 biometers based on different optical technologies.

Authors:  Robert Montés-Micó
Journal:  J Cataract Refract Surg       Date:  2022-01-01       Impact factor: 3.351

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.