Ultra-low-dose computed tomography for attenuation correction of cadmium-zinc-telluride single photon emission computed tomography myocardial perfusion imaging.

Marvin Grossmann1, Andreas A Giannopoulos1, Fabiola A Bechtiger1, Michael Messerli1, Moritz Schwyzer1, Dominik C Benz1, Ken Kudura1, Catherine Gebhard1, Christoph Gräni1, Aju P Pazhenkottil1, Philipp A Kaufmann1, Ronny R Buechel2.   

Abstract

BACKGROUND: The applicability of ultra-low-dose computed tomography (CT) for attenuation correction (AC) of single-photon-emission computed tomography myocardial perfusion imaging (SPECT-MPI) remains elusive. METHODS AND
RESULTS: One-hundred patients underwent one-day 99mTc-tetrofosmin stress-rest MPI and non-contrast enhanced cardiac CT with 120, 80, and 70 kilovolt peak (kVp) tube voltage and tube current of 200 milliamperes for creation of AC maps. Normalized percent myocardial uptake from SPECT-MPI using 80 kVp scans for AC showed excellent correlation vs AC from 120 kVp scans for stress [intraclass correlation (ICC) = 0.988, 95% CI = 0.986-0.989, P < .001] and rest (ICC = 0.985, 95% CI = 0.983-0.987, P < .001) with narrow Bland-Altman limits of agreement (BA-LA) (- 5.3% to 4.5% and - 5.4% to 4.4%, respectively) and minimal bias (- 0.4% and - 0.5%, respectively). Correlation of AC SPECT-MPI based on 70 vs 120 kVp scans was excellent for stress (ICC = 0.988, 95% CI = 0.986-0.989, P < .001) and rest (ICC = 0.986, 95% CI = 0.984-0.987, P < .001) with narrow BA-LA (- 5.3% to 4.4% and - 5.2% to 4.5%, respectively) and small bias (- 0.4% and - 0.3%, respectively). Mean effective radiation dose for the 120, 80 and 70 kVp scans were 0.58 ± 0.07, 0.19 ± 0.02, and 0.12 ± 0.01 mSv, respectively.
CONCLUSIONS: Attenuation maps for MPI obtained from ultra-low radiation dose CT scans are interchangeable with attenuation maps from standard-dose CT while offering a substantial reduction in radiation dose exposure.

Entities:  

Keywords:  SPECT; attenuation correction; computed tomography; ultra-low dose

Mesh:

Substances:

Year:  2018        PMID: 29923103     DOI: 10.1007/s12350-018-1303-y

Source DB:  PubMed          Journal:  J Nucl Cardiol        ISSN: 1071-3581            Impact factor:   5.952


  37 in total

1.  Improved outcome prediction by SPECT myocardial perfusion imaging after CT attenuation correction.

Authors:  Aju P Pazhenkottil; Jelena-Rima Ghadri; Rene N Nkoulou; Mathias Wolfrum; Ronny R Buechel; Silke M Küest; Lars Husmann; Bernhard A Herzog; Oliver Gaemperli; Philipp A Kaufmann
Journal:  J Nucl Med       Date:  2011-02       Impact factor: 10.057

2.  Yield of a novel ultra-low-dose computed tomography device mounted on a dedicated cardiac SPECT system in improving the accuracy of myocardial perfusion imaging and the detection of chest abnormalities.

Authors:  Nili Zafrir; Gigeon Shafir; Gil Kovalski; Israel Mats; Jean-Paul Bouhnik; Alexander Battler; Alejandro Solodky
Journal:  J Nucl Cardiol       Date:  2011-12-28       Impact factor: 5.952

3.  Ionizing radiation risks of cardiac imaging: estimates of the immeasurable.

Authors:  Philipp A Kaufmann; Juhani Knuuti
Journal:  Eur Heart J       Date:  2010-09-09       Impact factor: 29.983

4.  Use of coronary calcium score scans from stand-alone multislice computed tomography for attenuation correction of myocardial perfusion SPECT.

Authors:  Tiziano Schepis; Oliver Gaemperli; Pascal Koepfli; Christine Rüegg; Cyrill Burger; Sebastian Leschka; Lotus Desbiolles; Lars Husmann; Hatem Alkadhi; Philipp A Kaufmann
Journal:  Eur J Nucl Med Mol Imaging       Date:  2006-08-04       Impact factor: 9.236

5.  Comparison of 80 and 120 kVp contrast-enhanced CT for attenuation correction in PET/CT, using quantitative analysis and reporter assessment of PET image quality.

Authors:  H Bernstine; V Sopov; N Yefremov; M Nidam; M Gabbai; J Sosna; D Groshar
Journal:  Clin Radiol       Date:  2013-10-21       Impact factor: 2.350

6.  The feasibility of low-dose CT protocols for coronary artery calcium scoring and PET attenuation correction in cardiac PET/CT.

Authors:  Tung-Hsin Wu; Kun-Mu Lu; Nien-Yun Wu; Shyh-Jen Wang; Greta S P Mok; Bang-Hung Yang; Ching-Ching Yang
Journal:  Nucl Med Commun       Date:  2015-04       Impact factor: 1.690

7.  A scanning line source for simultaneous emission and transmission measurements in SPECT.

Authors:  P Tan; D L Bailey; S R Meikle; S Eberl; R R Fulton; B F Hutton
Journal:  J Nucl Med       Date:  1993-10       Impact factor: 10.057

8.  Real-time respiratory triggered SPECT myocardial perfusion imaging using CZT technology: impact of respiratory phase matching between SPECT and low-dose CT for attenuation correction.

Authors:  Olivier F Clerc; Tobias A Fuchs; Mathias Possner; Jan Vontobel; Fran Mikulicic; Julia Stehli; Riccardo Liga; Dominik C Benz; Christoph Gräni; Aju P Pazhenkottil; Oliver Gaemperli; Ronny R Buechel; Philipp A Kaufmann
Journal:  Eur Heart J Cardiovasc Imaging       Date:  2016-03-16       Impact factor: 6.875

9.  Validation of CT attenuation correction for high-speed myocardial perfusion imaging using a novel cadmium-zinc-telluride detector technique.

Authors:  Bernhard A Herzog; Ronny R Buechel; Lars Husmann; Aju P Pazhenkottil; Irene A Burger; Mathias Wolfrum; Rene N Nkoulou; Ines Valenta; Jelena R Ghadri; Valerie Treyer; Philipp A Kaufmann
Journal:  J Nucl Med       Date:  2010-09-16       Impact factor: 10.057

10.  Current worldwide nuclear cardiology practices and radiation exposure: results from the 65 country IAEA Nuclear Cardiology Protocols Cross-Sectional Study (INCAPS).

Authors:  Andrew J Einstein; Thomas N B Pascual; Mathew Mercuri; Ganesan Karthikeyan; João V Vitola; John J Mahmarian; Nathan Better; Salah E Bouyoucef; Henry Hee-Seung Bom; Vikram Lele; V Peter C Magboo; Erick Alexánderson; Adel H Allam; Mouaz H Al-Mallah; Albert Flotats; Scott Jerome; Philipp A Kaufmann; Osnat Luxenburg; Leslee J Shaw; S Richard Underwood; Madan M Rehani; Ravi Kashyap; Diana Paez; Maurizio Dondi
Journal:  Eur Heart J       Date:  2015-04-21       Impact factor: 29.983

View more
  1 in total

1.  Dose reduction is good but it is image quality that matters.

Authors:  R Glenn Wells
Journal:  J Nucl Cardiol       Date:  2018-07-24       Impact factor: 5.952

  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.