Literature DB >> 29922068

The safety and accuracy of ECG-guided PICC tip position verification applied in patients with atrial fibrillation.

Yufang Gao1, Yuxiu Liu1,2, Hui Zhang1, Fang Fang3, Lei Song4.   

Abstract

BACKGROUND: Tip position verification of peripherally inserted central catheters (PICCs) is essential to the use of the catheter. Postprocedural chest X-ray as the "gold standard" practice for PICC tip confirmation can lead to a significant delay for patient IV therapy, cost more, and lead to radiation exposure for both patients and staffs. Intracavitary electrocardiogram (IC-ECG)-guided PICC placement which provides real-time tip confirmation during the insertion procedure has been widely used. However, safety and accuracy of ECG for abnormal surface ECG patients, such as patients with atrial fibrillation (AF), have not been reported.
OBJECTIVE: To determine the safety and accuracy of IC-ECG technique for PICC tip position verification among the patients with AF. PATIENTS AND METHODS: A prospective cohort study was conducted in a teaching and tertiary referral hospital with more than 3,600 beds in Qingdao, People's Republic of China. Adult patients with diagnosis of AF who need a PICC for infusion from June 2015 to May 2017 were enrolled in the study. For every included patient with AF, ECG was used to detect the PICC tip position during catheterization and X-ray was done to confirm the tip position as the "gold standard" after PICC insertion. The effectiveness and accuracy of ECG-guided catheter tip positioning and chest X-ray confirmation were compared.
RESULTS: Totally, 118 AF patients with 118 PICCs were enrolled (58 male and 60 female, age range 50-89 years old). There was no catheterization-related complication. When the catheter entered the lower 1/3 of superior vena cava, the amplitude of f wave reached the maximum. There was no statistical difference between X-ray PICC tip position verification and IC-ECG PICC tip position verification among patients with AF (χ2=1.31, P=0.232). Utilizing the cutoff point of f wave change ≥0.5 cm, a sensitivity of 0.94, a specificity of 0.71, a positive predictive value of 0.98, and a negative predictive value of 0.42 were observed. The area under the receiver operating characteristic curve was 0.909 (95% CI: 0.810-1.000).
CONCLUSION: The ECG-guided technique represents a safe and accurate technique to verify the position of PICC tip in patients with AF and could potentially remove the requirement for postprocedural chest X-ray among the patients with AF.

Entities:  

Keywords:  ECG; PICC; electrocardiograph; patients with atrial fibrillation; peripherally inserted central catheter; tip position

Year:  2018        PMID: 29922068      PMCID: PMC5995413          DOI: 10.2147/TCRM.S156468

Source DB:  PubMed          Journal:  Ther Clin Risk Manag        ISSN: 1176-6336            Impact factor:   2.423


Introduction

Correct tip positioning of peripherally inserted central catheters (PICCs) is crucial to avoid catheterization-related complications such as migration, venous thrombosis, or arrhythmia.1 There are several methods for PICC tip positioning, postprocedural chest X-ray, electrocardiogram (ECG), and some reported novel techniques such as Sherlock 3CG® Tip Confirmation System (Bard Access Systems, Inc., Salt Lake City, UT, USA) which integrated magnetic tracking and ECG-based PICC tip confirmation technology2 and conductance guidewire system.3 The chest X-ray is the most frequently used method to verify PICC tip position and was recommended as the gold standard.4 One of the limitations of X-ray is that the postprocedural confirmation may lead to a significant delay in a patient intravenous (IV) therapy.5 Moreover, if incorrect PICC tip position is detected by postprocedural X-ray, repeated catheter manipulation and chest X-ray are required, which leads to a delay in patient treatment and further time use with increased cost. In addition, there is a possibility for complications, including catheter-related bloodstream infection owing to the integrity of the dressing being interrupted.6,7 The additional time, cost, and radiation exposure involved with radiographic assessment lead to PICC placement only within a hospital setting.1 ECG technique used in central venous catheter (CVC) tip positioning was first reported in 1949.8 Intracavitary ECG-guided PICC placement provides real-time tip confirmation during the insertion procedure. More and more evidence indicated ECG-guided PICC tip positioning could be as accurate as X-ray method.5,9–11 Walker’s systematic review suggested ECG-based positioning could potentially remove the requirement for postprocedural chest X-ray, especially during PICC line insertion.6 ECG-guided PICC tip location is clarified live during catheter insertion with no need for postprocedural adjustment or repositioning. The PICC can be used straight after placement with no time delay for patient treatment.9 The current standard for PICC tip position is the lower one-third of the superior vena cava (SVC) at the caval–atrial junction (CAJ).10,11 As the PICC tip approaches the sinoatrial node at CAJ, the P wave starts to elevate, reaching its maximum amplitude at the CAJ. When it passes through into the right atrium, the P wave starts to invert, indicating the PICC is inserted too far. The ideal PICC tip position is where the ECG shows maximal P amplitude. Since ECG-guided PICC tip positioning is regarded as a safe, reliable, and replication method, can it be used in patients with atrial fibrillation (AF) who have no P wave in their ECG? Our research team is from the Affiliated Hospital of Qingdao University where about 5,000 PICCs are inserted annually. The research team has been working on the research of PICCs, and during our research, we found that AF patients also had some distinct changes in the f waves during the PICC tip positioning. The team consequently explored this.

Patients and methods

Patients

This prospective cohort study was conducted in the Affiliated Hospital of Qingdao University, a tertiary referral hospital with more than 3,600 beds from June 2015 to May 2017. The study protocol was reviewed and approved by institutional review board of the Affiliated Hospital of Qingdao University (approval no. QDFYLL201422). All the patients enrolled signed written informed consent. The inclusion criteria consisted of the following: 1) patients needing a PICC whose ECG showed AF wave before PICC insertion; 2) over 18 years old; 3) patients tolerable to X-ray examination. Exclusion criteria were as follows: 1) patients suffering from mental illness or skin diseases; 2) patients with a heart pacemaker; 3) patients with any other types of catheter 4) patients allergic to alcohol and iodophor.

PICC insertion procedure

PICC was inserted by professional PICC nurses with ultrasound guidance under standard sterile conditions. Four French (Fr) single-lumen distal valved silicone Bard Groshong® PICC (Bard Access Systems, Inc.) was used in the study. Bard Site Rite 5 Sonogram Machine Ultrasound System (Bard Access Systems, Inc.) was used to measure and evaluated relevant veins at the insertion point. All PICCs were inserted by the modified Seldinger technique using Groshong® NXT ClearVue. All the PICCs were flushed with 10 mL saline after catheterization and the catheter entrance sites were covered with dressing after skin disinfection. Chest radiography was routinely examined to confirm the catheter tip position.10

Tip confirmation by X-ray

According to the Infusion Nurses Society, the recommended tip position is located at the lower third of the SVC close to the entrance to the right atrium.11 It was reported that CVC tip position approximately 4 cm (95% CI: 3.8–4.3 cm) below the carina would result in placement near the CAJ and the average length of SVC was 7.1 cm.12 In this study, we used X-ray method as the “gold standard” to confirm PICC tip position. When taking X-ray examination, all patients took a neutral supine position, arms along the body, and with no forced breathing to avoid possible tip malposition by position or forced inspiration. We utilized carina as an anatomical landmark from which to measure the PICC tip. The optimal position was estimated to be 1.6–4 cm under the carina in our study.12,13 The X-ray data were evaluated separately by 2 radiologists. If there was inconsistent judgment, a third radiologist would further check the X-ray result and confirm a decision.

Intraprocedural verification of tip position by ECG method

The ECG in the study was obtained by so-called “saline technique” which uses a column of saline solution contained in the catheter as an intracavitary electrode.13 Braun® transducer and switch for shifting from surface ECG tracing to intracavitary ECG (IC-ECG) tracing were used in the study. Three surface electrodes (right arm [RA], left arm, and left leg) and lead II were connected. When the catheter tip entered SVC, the catheter was attached to the connector of the transducer, and then a continuous infusion of saline solution through PICC was performed. The ECG in patients with AF shows f wave instead of P wave. As the catheter tip goes deeper, f wave also undergoes certain changes. When the catheter proceeds into SVC, the f wave gets higher which is similar to P wave change, that is, when the catheter gets into the SVC, the amplitude of f wave gradually increases. When the catheter enters the lower 1/3 of SVC, the amplitude of the f wave reaches the maximum, and when the catheter enters the right atrium, the amplitude of the f wave falls.

Efficacy evaluation and data collection

For every PICC insertion, the following data were collected: 1) patient’s data, containing age, sex, diagnosis, and CVC history; 2) PICC insertion data, containing PICC insertion aim, side of insertion arm, vein of puncture, and possible puncture-related complication; 3) intracavitary ECG data, containing the surface ECG amplitude before PICC insertion, peak of f wave while entering the lower 1/3 of SVC, and possible intraoperative complication related to IC-ECG; 4) X-ray data on PICC tip position. All the data were collected by the authors and all PICCs were inserted by PICC specialty nurses with at least 5-year experience in PICC placement. ECG was confirmed by a cardiologist and the authors.

Statistical analysis

The data were recorded and analyzed using SPSS 19.0 (IBM Corporation, Armonk, NY, USA). Continuous variables were presented as mean ± SD. χ2 test was used to evaluate the difference of enumeration data. The sensitivity and specificity of f amplitude in 1/3 SVC compared with X-ray as “gold standard” were calculated. The significance level was set at 0.05.

Results

Study population

From June 2015 to May 2017, 118 AF patients with 118 PICCs who met the eligibility criteria were enrolled in the study and retained for analysis. There were 58 men (49.15%) and 60 women (50.85%), with a mean age of 67.23±10.12 years (range, 50–89 years old). Of the 118 patients with AF, 16 patients were inserted PICCs for parenteral nutrition, 90 for chemotherapy, and 12 with poor peripheral vein condition for long-term infusion. There was no catheterization-related complication. The baseline characteristics of the 118 PICCs are listed in Table 1.
Table 1

Baseline demographics, PICC insertions, and clinical characteristics

VariablesN (%)IC-ECG f amplitude ≥0.5 cm higher N
χ2P-values
YesNo
Sex1.3380.247
 Male58 (49.2)544
 Female60 (50.8)528
Diagnosis0.6670.955
 Hematopoietic malignancies29 (24.6)254
 Gynecologic malignancies19 (16.1)172
 Breast cancer21 (17.8)192
 Gastrointestinal malignancies27 (22.9)252
 Other diseases22 (18.6)202
PICC insertion aim2.1430.342
 Parenteral nutrition16 (13.6)133
 Chemotherapy90 (76.3)837
 Long-term infusion12 (10.2)102
Side of insertion arm0.0001.000
 Right upper extremity92 (78.0)839
 Left upper extremity26 (22.0)242
Vein of puncture1.7900.409
 Basilic vein91 (77.1)838
 Brachial vein24 (20.3)204
 Cephalic vein3 (2.5)30
CVC history0.6430.423
 Yes24 (20.3)204
 No94 (79.7)868

Abbreviations: IC-ECG, intracavitary electrocardiogram; N, number; PICC, peripherally inserted central catheter; CVC, central venous catheter.

Change of f wave amplitude as the tip entered lower 1/3 of SVC

In this study, we found that the f wave amplitude changed as the PICC tip went deeper at SVC. When the catheter entered into SVC, f wave got higher, which was similar to P wave change of the normal surface ECG patients. When the catheter entered the lower 1/3 of SVC, the amplitude of f wave reached the maximum. When the catheter entered the right atrium, the amplitude of the f wave began to fall. Figure 1 shows the surface ECG and the IC-ECG when PICC tip entered the lower 1/3 of SVC in a patient with AF.
Figure 1

Change of f wave as PICC entered the lower 1/3 of SVC, CAJ, and the right atrium.

Notes: The surface ECG before PICC insertion (A), the ECG when PICC tip approached the lower 1/3 of SVC (B), the ECG when PICC tip approached CAJ (C), the ECG when PICC tip entered the right atrium (D). Red circles = f wave.

Abbreviations: CAJ, caval–atrial junction; ECG, electrocardiogram; PICC, peripherally inserted central catheter; SVC, superior vena cava.

When the f wave reached the maximum amplitude, it was recorded as f′. The “f” stood for the average surface ECG f amplitude. The average of surface ECG f wave amplitude, f′, and (f′ - f)/f were calculated. The detailed data are listed in Table 2.
Table 2

Average height of f wave and 95% CI

VariablesNAverage height (cm)95% CI
LowerUpper
f1181.06±0.251.0151.105
f′1181.68±0.340.6191.741
(f′ - f)/f1180.59±0.110.5730.615

Notes: f: f stands for the average surface ECG f amplitude. f′: when the f wave reached the maximum amplitude during PICC insertion, it was recorded as f′.

Abbreviations: ECG, electrocardiogram; N, number; PICC, peripherally inserted central catheter.

Accuracy of IC-ECG using X-ray as a gold standard

There was no statistical difference between X-ray PICC tip position verification and IC-ECG PICC tip position verification among patients with AF (χ2=1.31, P=0.232). If the average f wave amplitude increased ≥0.5 cm compared that of the surface ECG f wave amplitude, IC-ECG was recorded having positive change. If X-ray showed the tip position of PICC was in the lower 1/3 of SVC, X-ray as the gold standard was recorded positive. Utilizing the cutoff point of f wave change ≥0.5 cm, a sensitivity of 0.94, a specificity of 0.71, a positive predictive value of 0.98, and a negative predictive value of 0.42 were observed. Table 3 shows the detailed data.
Table 3

Accuracy of IC-ECG for tip verification using X-ray as a gold standard

IC-ECGX-ray
Total
PositiveNegative
Positive1060106
Negative7512
Total1135118

Notes: Positive: if the tip of PICC was in the lower 1/3 of SVC, it was recorded as positive. Negative: if the tip of PICC was not in the lower 1/3 of SVC, it was recorded as negative.

Abbreviations: IC-ECG, intracavitary electrocardiogram; PICC, peripherally inserted central catheter; SVC, superior vena cava.

A receiver operating characteristic curve was calculated to verify the relationship between X-ray PICC tip position verification and IC-ECG tip position verification (Figure 2). The area under the curve was 0.909 (95% CI: 0.810–1.000), indicating that the accuracy of the IC-ECG PICC tip verification was good.
Figure 2

Receiver operating characteristic curve for the relationship between X-ray PICC tip position verification and IC-ECG tip position verification.

Abbreviations: IC-ECG, intracavitary electrocardiogram; PICC, peripherally inserted central catheter; ROC, receiver operating characteristic.

Discussion

Since the first report of Hellerstein et al’s8 study in 1949 regarding CVC placement guided by ECG, numerous studies have been performed to evaluate whether this technique was reliable when placing a CVC. But there have been few studies done on using ECG technology specifically for PICC position verification.5 Smith et al’s1 study found that the peak P wave amplitude was highest at the superior vena cava–right atrial (SVC-RA) junction by performing venography while placing the PICCs under ECG control at 1 cm intervals. They concluded that with PICC insertion into the right atrial, P wave amplitude decreased and eventually became negative.1 In Pittiruti et al’s13,14 studies, they found that there were no complications potentially related to the method itself, and they drew the conclusion that the intracavitary ECG method for real-time confirmation of tip position was accurate, safe, feasible in all adult patients, and applicable to any type of short-term or long-term central venous access device. Oliver et al’s9 5-year observation found that using ECG to confirm PICC tip position was cheaper, quicker, and more accurate than postprocedural chest X-ray. Zhou et al’s15,16 studies of the ECG-guided PICC tip placement among neonates found no significant complications occurred in the studied neonates. Their studies suggest ECG-guided tip placement is a promising approach for improving the success rate of tip location when placing a PICC in neonates and the accuracy in the lower extremity was higher than that in the upper extremity among neonates.15,16 Rossetti’s study suggests that the IC-ECG method is safe and accurate in the pediatric patients.17 The studies so far have shown that the ECG-guided method has clear advantages in terms of accuracy and feasibility, does not expose the patient to any radiation, is cheap to run, and provides instant, real-time confirmation of tip position. Furthermore, there is no time delay in commencing the patient on IV therapy.9,13,17,18 However, all the studies above were based on the normal surface ECG because ECG technique relied on interpretation of changes to a patient’s P wave, and patients with rhythms were often in the exclusion criteria from published studies.19,20 Besides Engelhardt et al’s21 study on comparing CVC location control by ECG and by chest X-ray in 40 patients with absolute arrhythmia and AF, there were no other data on patients with abnormal surface ECG such as patients with AF with f wave instead of P wave. To the best of our knowledge, this study is the first study to evaluate the safety and accuracy of ECG-guided PICC tip position verification applied in patients with AF. In this study, we found that the f wave amplitude changed as the PICC tip went deeper at SVC. When the catheter entered SVC, f wave got higher, which was similar to P wave change of the normal surface ECG patients. When the catheter entered the lower 1/3 of SVC, the amplitude of f wave reached the maximum. When the catheter entered the right atrium, the amplitude of the f wave began to fall. In the normal surface ECG, changes of the P wave morphology are used to track the PICC tip as the tip goes along the SVC to the right atrium. As the PICC tip approaches the sinoatrial node at the CAJ, P wave starts to elevate and reaches the maximum height at the CAJ. When it passes through into the RA, the P wave starts to invert, indicating the PICC is inserted too deep. The ideal end position for the PICC is where the ECG shows maximal P.5 Patients with AF have no P wave, instead it is by f wave with frequency of 350–600 times/min. The cardiac electrical principles of changes of f wave during the insertion of PICC observed in this study are not clear at present. When the PICC tip goes close to the right atrium, it can stimulate cardiac myocytes to generate potentials. The f wave is modified by the near field potential and the far field potential. The amplitude of f wave arises along with the deepening of the catheter tip. When the tip of PICC comes to the ideal position (the lower 1/3 of SVC), f wave amplitude is the highest, which is similar to the change of P wave in the normal surface ECG patients. The mechanism may be that when the catheter tip approaches the right atrium, it stimulates the myocardium to generate electric potential. Under the effect of the near field potential and the far field potential, f wave changes. The amplitude of the f wave increases gradually as the tip of the catheter goes deeper. When the tip of the catheter reaches the ideal position, the amplitude of the f wave also reaches the highest, showing a change similar to the P wave.21 However, further study is required to clarify the detailed f wave change regulation and its specific mechanism. Regarding the safety of the ECG-guided method used among patients with AF, our study (118 AF patients with 118 PICCs) did not detect any complication related to the use of an intracavitary electrode. The reported incidence of heart rhythm disturbance by ECG method was very low (0.7%).13 The operator can know at each moment the ECG waveform, so there is no danger of going too deep. Regarding the accuracy, our study found that there was no statistical difference between X-ray method and IC-ECG method among patients with AF. Using the cutoff point of f wave change ≥0.5 cm, a sensitivity of 0.94, a specificity of 0.71, a positive predictive value of 0.98, and a negative predictive value of 0.42 were observed. The area of the receiver operating characteristic curve was 0.909 (95% CI: 0.810–1.000), indicating that there was high accuracy with the ECG method when X-ray was considered as gold standard. The change of f wave was quantized in our study, which provided relatively objective data for the judgment of f wave. As shown in Table 2, the f wave amplitude increased by 0.59±0.11 cm (95% CI: 0.573–0.615) on average when PICC tip came to the lower 1/3 of SVC.

Limitations

This study has several limitations. First, the number of samples is relatively less. In clinical application, the f wave morphology can change by the influence of various factors, such as disease and cardiac function, and so the reliability and practicability of ECG method for patients with AF need to be further explored and verified. Second, we only detected the f amplitude changes during PICC placement, and detailed f waveform changes need to be recorded in a further study. Third, this is only a preliminary observation study, and a random control design will be needed to compare the effectiveness between ECG-guided PICC tip confirmation and X-ray.

Conclusion

Recent national guidelines (National Institute for Health and Care Excellence, 2015) have advocated the ECG method as a valid alternative to radiological verification of CVC tip placement.22 This study suggests ECG guidance is a safe and effective way of correctly confirming PICC tip position for patients with AF and can potentially remove the requirement for postprocedural chest X-ray among the patients with AF. Further studies are required to evaluate the detailed f waveform changes and to compare the effectiveness between ECG-guided PICC tip confirmation and X-ray.
  19 in total

Review 1.  The electrocardiographic method for positioning the tip of central venous catheters.

Authors:  Mauro Pittiruti; Antonio La Greca; Giancarlo Scoppettuolo
Journal:  J Vasc Access       Date:  2011 Oct-Dec       Impact factor: 2.283

2.  Intravenous electrocardiographic guidance for placement of peripherally inserted central catheters.

Authors:  Brigham Smith; Renée M Neuharth; Mary Ann Hendrix; Daniel McDonnall; Andrew D Michaels
Journal:  J Electrocardiol       Date:  2010-03-06       Impact factor: 1.438

3.  ECG or X-ray as the 'gold standard' for establishing PICC-tip location?

Authors:  Gemma Oliver; Matt Jones
Journal:  Br J Nurs       Date:  2014-10-22

4.  ECG-based PICC tip verification system: an evaluation 5 years on.

Authors:  Gemma Oliver; Matt Jones
Journal:  Br J Nurs       Date:  2016-10-27

5.  [Evaluation of tip location reliability of intraprocedural ECG vs. chest Xrays in PICC placement].

Authors:  Stefano Elli; Luigi Cannizzo; Alessia Marini; Silvia Porcarelli; Filippo Azzarone; Roberto Fumagalli; Giuseppe Foti; Alberto Lucchini
Journal:  Assist Inferm Ric       Date:  2016 Oct-Dec       Impact factor: 0.804

6.  The intracavitary ECG method for positioning the tip of central venous access devices in pediatric patients: results of an Italian multicenter study.

Authors:  Francesca Rossetti; Mauro Pittiruti; Massimo Lamperti; Ugo Graziano; Davide Celentano; Giuseppe Capozzoli
Journal:  J Vasc Access       Date:  2014-09-02       Impact factor: 2.283

7.  Accurate nonfluoroscopic guidance and tip location of peripherally inserted central catheters using a conductance guidewire system.

Authors:  Mark C Svendsen; David Birrer; Benjamin Jansen; Shawn D Teague; Bill Combs; Gregory J Schears; Ghassan S Kassab
Journal:  J Vasc Surg Venous Lymphat Disord       Date:  2013-03-14

8.  Effectiveness of Intracavitary Electrocardiogram Guidance in Peripherally Inserted Central Catheter Tip Placement in Neonates.

Authors:  Lianjuan Zhou; Hongzhen Xu; Jianfeng Liang; Meifang Xu; Jun Yu
Journal:  J Perinat Neonatal Nurs       Date:  2017 Oct/Dec       Impact factor: 1.638

Review 9.  Effectiveness of electrocardiographic guidance in CVAD tip placement.

Authors:  Graham Walker; Raymond J Chan; Evan Alexandrou; Joan Webster; Claire Rickard
Journal:  Br J Nurs       Date:  2015 Jul 23-Aug 12

10.  An Accuracy Study of the Intracavitary Electrocardiogram (IC-ECG) Guided Peripherally Inserted Central Catheter Tip Placement among Neonates.

Authors:  Lian-Juan Zhou; Hong-Zhen Xua; Mei-Fang Xu; Yan Hu; Xiao-Fang Lou
Journal:  Open Med (Wars)       Date:  2017-05-22
View more
  8 in total

1.  Diagnostic accuracy of ultrasound for localising peripherally inserted central catheter tips in infants in the neonatal intensive care unit: a systematic review and meta-analysis.

Authors:  Shauna C Doyle; Niamh M Bergin; Rena Young; Andrew England; Mark F McEntee
Journal:  Pediatr Radiol       Date:  2022-05-05

Review 2.  ECG-Based Techniques to Optimize Peripherally Inserted Central Catheters: Rationale for Tip Positioning and Practical Use.

Authors:  Giuseppe Gullo; Salah D Qanadli
Journal:  Front Cardiovasc Med       Date:  2022-05-06

3.  Magnetic Tracking and Electrocardiography-Guided Tip Confirmation System Versus Fluoroscopy for Placement of Peripherally Inserted Central Catheters: A Randomized, Noninferiority Comparison.

Authors:  V Mack; D Nißler; D Kasikci; A Malouhi; R Aschenbach; U Teichgräber
Journal:  Cardiovasc Intervent Radiol       Date:  2020-06-17       Impact factor: 2.740

4.  A clinical study on the tip localization of peripherally inserted central catheter (PICC) guided by intracavitary electrocardiography in newborns: a randomised trial.

Authors:  Li-Bo Zhu; Ling Liu; Tie-Song Zhang; Yu-Ting Zheng; Chun-Yan Lu; Kun Lu; Shu-Xian Zhang; Liu-Yan Duan; Mei-Lin Yang
Journal:  Transl Pediatr       Date:  2021-10

5.  ECG method for positioning the tip of peripherally inserted central catheters in patients with atrial fibrillation.

Authors:  Conghui Zhao; Yuxin Zhu; Xinxin Yin; Congcong Zhang; Yingxia He; Jingfang Gao
Journal:  Ann Noninvasive Electrocardiol       Date:  2022-02-17       Impact factor: 1.485

6.  Effects of different concentrations of intraluminal sodium chloride solution on intracavitary ECG used for arm infusion port implantation.

Authors:  Lei Dong; Chen-Yang Guan; Ying Zhang; Ai-Xia Wang; Ming-Hua Liu; Chen Guo; Xiao-Li Hao; Qi Zhang
Journal:  Sci Rep       Date:  2022-08-15       Impact factor: 4.996

7.  Effects of Peripherally Inserted Central Catheter (PICC) Catheterization Nursing on Bloodstream Infection in Peripheral Central Venous Catheters in Lung Cancer: A Single-Center, Retrospective Study.

Authors:  Qiu Hu; YanHong Su; Li Yan
Journal:  Comput Math Methods Med       Date:  2022-09-15       Impact factor: 2.809

8.  Acute ischaemia of the upper limb following peripherally inserted central catheter-a venous to arterial complication cascade.

Authors:  Khaleel A Hamdulay; Rene van den Bosch
Journal:  J Surg Case Rep       Date:  2021-07-09
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.