| Literature DB >> 35600463 |
Giuseppe Gullo1,2, Salah D Qanadli1,2.
Abstract
Intracavitary electrocardiography is an accurate guidance technique for peripherally inserted central catheters (PICC) tip location that is spreading widely among providers using non x-ray-based facilities. The principle behind this technology relies on the transmission of the electrocardiographic signal at the tip of the catheter and its use as an internal mobile electrode, allowing the system to identify the cavo-atrial junction (CAJ) through internal P-wave amplitude modulations. The gain in popularity of intracavitary electrography and its large diffusion have led manufacturers to offer various devices with heterogeneous properties, among which clinician who place PICCs have to choose. It is therefore important to understand differences between available techniques and devices. The potential impact might not only affect availability and costs but also the clinical impact through advantages and limitations regarding electric signal transmission PICC selection. Current perspectives on intracavitary electrocardiography will also be discussed, to give the reader a global view of the management of electrocardiographically guided PICCs, especially in an environment without x-ray support.Entities:
Keywords: ECG; PICC; central venous access; central venous catheters; intracavitary electrocardiogram; tip locating device; tip location
Year: 2022 PMID: 35600463 PMCID: PMC9120917 DOI: 10.3389/fcvm.2022.765935
Source DB: PubMed Journal: Front Cardiovasc Med ISSN: 2297-055X
FIGURE 1(A) Chest X-ray displaying CAJ localization. Catheter tip in T1 (tip within 1 cm of the CAJ) is optimal. Catheter tip in T2 (tip within 1–3 cm of the CAJ) is suboptimal. Catheter tip in T3 (tip more than 3 cm below the CAJ or not in the SVC) is inadequate, needing repositioning. (B) Intracavitary electrocardiogram with overlay of the QRS complex zone, T-wave zone, and P-wave zone. A catheter tip in T3 that is not in the SVC will feature this trace, equivalent to the superficial ECG trace. (C) Intracavitary electrocardiogram with increasing P-wave (equivalent to the T wave). This is representative of a catheter tip approaching the CAJ in T2. (D) Intracavitary electrocardiogram with maximal P-wave without initial negative deflection. The tip is optimally positioned at the CAJ in the T1 zone (within 1 cm of the CAJ). (E) Intracavitary electrocardiogram with decreasing P-wave and initial negative deflection. This is representative of a catheter tip beyond the CAJ, in the T2–T3 zone. CAJ, cavo-atrial junction; SVC, superior vena cava.
FIGURE 2(A) PICC with a free-end-side trim point (technique A). Valved reverse-tapered device with solid-wire signal conduction. PUR material with high-flow injection properties. (B) PICC with a hub-side (outer part) trim point (technique B). Centrally valved straight device with solid-wire signal conduction and silicone material without high-flow injection properties. (C) PICC with a free-end-side trim point (technique A). Non-valved reverse-tapered device with liquid (saline) signal conduction. PUR material with high-flow injection properties. (D) PICC with a hub-side (outer part) trim point (technique B). Non-valved straight device with liquid (saline) signal conduction. PUR material with high-flow injection properties. PICC, peripherally inserted central catheter; PUR, polyurethane.
Summary of the different advantages and limits of PICC devices classified by technique and transmission medium.
| Technique A | Technique B | |||
| Wire | Precision | ✓ | Precision | ✓ |
| Clot occlusion of the PICC | ✓ | Clot occlusion of the PICC | ✓ | |
| Conduction | ✓✓ | Conduction | ✓✓ | |
| Insertion site repercussions | – | Insertion site repercussions | ✓ | |
| 5 ml/s power injection | ✓ | 5 ml/s power injection | – | |
| Cost | ✓ | Cost | – | |
| Saline | Precision | ✓ | Precision | ✓ |
| Clot occlusion of the PICC | – | Clot occlusion of the PICC | – | |
| Conduction | ✓ | Conduction | ✓ | |
| Insertion site repercussions | – | Insertion site repercussions | ✓ | |
| 5 ml/s power injection | ✓ | 5 ml/s power injection | ✓ | |
| Cost | ✓ | Cost | ✓ | |
Summary of different commercially available PICC devices classified by technique and valve presence.
| Technique A | Technique B | |
| Valved | • Valved Pro-PICC®; Medical Components, Inc., Harleysville, PA, United States | • PowerGroshong™ PICC Catheter; Becton Dickinson, Franklin Lakes, NJ, United States |
| Non-valved | • Synergy CT PICC™; Health Line Medical Products, Salt Lake City, UT, United States | • CT PICC® Easy; Vygon, Ecouen, France |
All device may be ECG guided through homemade positioning system using connectors and a standard ECG machine with saline or wire transmission medium.
*Device with brand catheter tip positioning system, wired transmission medium.
**Device with brand catheter tip positioning system, saline transmission medium.