The Accuracy of Cone-Beam Computed Tomography for Evaluating Bone Density and Cortical Bone Thickness at the Implant Site: Micro-Computed Tomography and Histologic Analysis.

Srisurang Suttapreyasri1, Pradipat Suapear, Narit Leepong.   

Abstract

The aim of this study was to evaluate the accuracy of cone-beam computed tomography (CBCT) for determining cortical thickness and the gray value, investigating its correlation with micro-computed tomography (CT) and histology analysis. Sixty-two bone samples from 4 anatomic regions of the jaw were analyzed. A radiographic surgical stent was used during CBCT and bone sample harvesting. The cortical thickness and gray value of the planned implant were evaluated by CBCT. Bone volumetric fractions, bone mineral density, and % porosity assessed by micro-CT and mineralized material by histology analysis from harvested bone samples were analyzed and assessed for the association with the CBCT using Pearson correlation. A correlation between cortical thickness measured from the CBCT and Micro-CT (r = 0.933, P < 0.01) was identified. There was no difference between gray values measured from the CBCT among regions, while bone density parameters from micro-CT and histologic analysis showed significant difference (P < 0.01) among regions. Bone density parameters from micro-CT and histologic analysis showed correlation with cortical thickness but not with the gray value. In conclusion, CBCT is highly accurate in linear measurements and demonstrated correlation with genuine bone density. However, the gray value could not demonstrate the true bone density according to a low correlation to bone density variable measured from micro-CT and histologic analysis.

Mesh:

Substances:

Year:  2018        PMID: 29894463     DOI: 10.1097/SCS.0000000000004672

Source DB:  PubMed          Journal:  J Craniofac Surg        ISSN: 1049-2275            Impact factor:   1.046


  5 in total

1.  Definition of the Region of Interest for the Assessment of Alveolar Bone Repair Using Micro-computed Tomography.

Authors:  Juliana Simeão Borges; Vitor Cardoso Costa; Milena Suemi Irie; Gabriella Lopes de Rezende Barbosa; Rubens Spin-Neto; Priscilla Barbosa Ferreira Soares
Journal:  J Digit Imaging       Date:  2022-09-07       Impact factor: 4.903

2.  Effect of micro-computed tomography reconstruction protocols on bone fractal dimension analysis.

Authors:  Hugo Gaêta-Araujo; Nicolly Oliveira-Santos; Danieli Moura Brasil; Eduarda Helena Leandro do Nascimento; Daniela Verardi Madlum; Francisco Haiter-Neto; Christiano de Oliveira-Santos
Journal:  Dentomaxillofac Radiol       Date:  2019-08-13       Impact factor: 2.419

3.  Influence of reconstruction parameters of micro-computed tomography on the analysis of bone mineral density.

Authors:  Hugo Gaêta-Araujo; Eduarda Helena Leandro Nascimento; Danieli Moura Brasil; Daniela Verardi Madlum; Francisco Haiter-Neto; Christiano Oliveira-Santos
Journal:  Imaging Sci Dent       Date:  2020-06-18

4.  Evaluation of the effect of long-term use of antidepressants in the SSRI group on bone density with dental volumetric tomography.

Authors:  Kamil Serkan Agacayak; Rıdvan Guler; Bekir Ilyasov
Journal:  Drug Des Devel Ther       Date:  2019-10-03       Impact factor: 4.162

5.  Bone quality affects stability of orthodontic miniscrews.

Authors:  Wan-Ping Yu; Ming-Tzu Tsai; Jian-Hong Yu; Heng-Li Huang; Jui-Ting Hsu
Journal:  Sci Rep       Date:  2022-02-18       Impact factor: 4.379

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.