| Literature DB >> 29881885 |
Maximilian F Kasparek1, Michael Töpker2, Mathias Lazar2, Michael Weber2, Michael Kasparek3, Thomas Mang2, Paul Apfaltrer2, Bernd Kubista1, Reinhard Windhager1, Helmut Ringl4.
Abstract
PURPOSE: To evaluate the influence of different scan parameters for single-energy CT and dual-energy CT, as well as the impact of different material used in a TKA prosthesis on image quality and the extent of metal artifacts.Entities:
Keywords: Blooming artifacts; Dual-energy CT; Mono-energetic imaging; Single-energy CT; Streak artifacts; Total knee arthroplasty
Mesh:
Substances:
Year: 2018 PMID: 29881885 PMCID: PMC6527539 DOI: 10.1007/s00167-018-5001-8
Source DB: PubMed Journal: Knee Surg Sports Traumatol Arthrosc ISSN: 0942-2056 Impact factor: 4.342
Fig. 1CT scan of a water phantom showing blooming and beam-hardening artifacts (a). The scan shows a section through a plastic tube (1, low attenuation), an aluminum tube (2, medium attenuation), and a steel tube (3, high attenuation). All tubes had the same diameter of 10 mm and a wall thickness of 1 mm. The plastic tube is less dense than water, and was, therefore, visible as a dark ring without virtual thickening of the wall. The aluminum tube is denser, and was, therefore, visualized as a bright ring, with the wall thickness slightly thicker than the actual thickness, representing a slight blooming artifact. The steel tube, however, has the highest attenuation of all three materials and showed, therefore, a pronounced blooming artifact with a wall thickness four times larger than its physical size. The dark streaks and triangles (white arrows) around the steel pipe are typical streak artifacts that arise from very dense materials, such as steel. A comparison between a conventional CoCr steel alloy prosthesis (b) and a titanium prosthesis (c). White arrows show considerable streak artifacts in close proximity to the metal in the steel prosthesis, while the titanium prosthesis offered only a low amount of artifacts. The magnitude of the blooming artifacts for steel is demonstrated by the superimposed calipers (same length) on the condyles of both prostheses
All prosthesis pairs and material characteristics of all vendors
| List of total knee arthroplasty implants (TKA) | ||||
|---|---|---|---|---|
| Pairs | Standard model | Material | “Anti-allergic” model | Material |
| 1 | LCS® | CoCr-alloy | LCS® TiNi | Titanium alloy prosthesis with a titanium nitride ceramic surface |
| 2 | ACS® | CoCrMo-alloy | ACS® Titannitrid (TiN) | CoCr prosthesis coated with titanium nitride |
| 3 | Vanguard™ | CoCr-alloy | Vanguard™ Titan-Niob-Nitrid (TiNbN) | CoCr prosthesis coated with titanium-niobium-nitride |
| 4 | Columbus® | CoCr-alloy | AS Columbus® | CoCr prosthesis ceramic coated with zirconium and zirconium nitride |
| 5 | Legion™ | CoCr-alloy | Legion™ Oxinium® | Zirconium alloy prosthesis with an oxidized ceramic surface |
| 6 | NexGen® LPS | CoCrMo-alloy | NexGen® Tivanium® LPS | Full titanium alloy prosthesis |
| 7 | BPK-S | CoCr-alloy | BPK-S ceramic knee | Ceramic prosthesis |
| 8 | Scorpio® | CoCr-alloy | Scorpio® TiNi | CoCr prosthesis coated with titanium-niobium-oxynitride |
1. DePuy Synthes Inc., Warsaw, IN, USA; 2. Implantcast Gmbh, Buxtehude, Germany; 3. Biomet Inc., Warsaw, IN, USA; 4. Aesculap AG, Tuttlingen, Germany; 5. Smith & Nephew PLC, London, UK; 6. Zimmer Inc., Warsaw, USA; 7. Peter Brehm Gmbh, Weisendorf, Germany/CeramTec Gmbh, Plochingen, Germany; 8. Stryker Corporation, MI, Kalamazoo, USA
Fig. 2Play Doh® prosthetic phantom setting. For simulation of the anatomic conditions, the distal femoral shaft of each TKA was formed with Play Doh® as a surrogate for cancellous bone. Play Doh® allowed the simulation of cancellous bone with comparable attenuation Hounsfield values of around 260 HU. The phantom was placed underwater in an acrylic glass water tank to simulate soft tissue around the knee joint and to avoid inappropriate artifacts
Scan and dose parameters for all protocols, including the subjective ranking of image quality at the bone/prosthesis interface
| Number of scan protocols | Dual (DE)- or single-energy (SE) CT | Tube A (kV) | Tube B (kV) | CARE DOSE | Ref. mAsa | Pitch | Rotation time (s) | Kernel | Iterative reconstruction SAFIRE strength | Dose (CTDIvol) | Blooming artifacts: mean virtual growth mm (± SD) | Streak artifacts ROI: mean image noise HU (± SD) | Subjective image quality rankingb | |
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
| 1 | DE | 100 | 140 | On | 200 | 155 | 0.8 | 0.5 | Q40f | 2 | 4.0 | 3.4 (± 1.6) | 258.0 (± 139.1) | 5 |
| 2 | DE | 100 | 140 | On | 300 | 232 | 0.6 | 0.5 | Q40f | 2 | 6.0 | 3.4 (± 1.6) | 242.8 (± 135.4) | 4 |
| 3 | DE | 100 | 140 | On | 100 | 77 | 0.6 | 0.5 | Q40f | 2 | 2.0 | 3.8 (± 1.9) | 296.3 (± 127.9) | 6 |
| 4 | DE | 80 | 140 | On | 401 | 155 | 0.8 | 0.5 | Q40f | 2 | 3.7 | 3.4 (± 1.7) | 256.9 (± 134.6) | 7 |
| 5 | DE | 100 | 140 | Off | 150 | 116 | 0.4 | 0.5 | Q40f | 2 | 11.8 | 3.3 (± 1.6) | 237.7 (± 138.4) | 3 |
| 6 | DE | 100 | 140 | Off | 200 | 155 | 0.4 | 0.5 | Q40f | 2 | 15.7 | 3.4 (± 1.6) | 225.1 (± 141.4) | 2 |
| 7 | DE | 100 | 140 | Off | 250 | 193 | 0.4 | 0.5 | Q40f | 2 | 19.6 | 3.4 (± 1.6) | 214.8 (± 139.0) | 1 |
| 8 | SE | 120 | – | Off | 210 | – | 0.6 | 0.5 | I40f | 2 | 14.2 | 5.5 (± 2.1) | 497.2 (± 208.1) | 8 |
| 9 | SE | 140 | – | Off | 210 | – | 0.6 | 0.5 | I40f | 2 | 20.6 | 4.6 (± 1.9) | 392.4 (± 172.3) | 9 |
Blooming artifacts are given as virtual growth (the difference between the actual diameter and the CT measurement) as the mean of all measured prostheses; streak artifacts are given as mean image noise for the mean of all prostheses. For both blooming and streak measurements, higher numbers represent more artifacts, and, therefore, lower image quality
mm millimeter, HU Hounsfield unit
aRef. mAs (quality reference milliampere × second) represents the value suitable for an average-sized patient that is automatically adjusted to the required value by the scanner according to the actual size of the patient
bConsensus ranking (1–9, 1 indicating the best result)
Blooming artifacts (virtual growth) and standard deviation of HU (streak artifacts), with the mean of the standard deviation of all measurements near each prosthesis compared for each single pair
| Measurement | Blooming artifacts | Streak artifacts | Image quality | |||||
|---|---|---|---|---|---|---|---|---|
| Prosthesis | Real diametera (mm) | Mean CT diametera (mm) | Blooming: mean virtual growth (mm) | Mean standard deviation (HU) | Difference | Median rankingb | ||
| LCS® | 8.7 | 13.3 (± 0.9) | 4.6 (± 0.9) | < 0.001 | 334.3 (± 88.9) | 257.1 | < 0.001 | 3 |
| LCS® TiNi | 8.7 | 9.3 (± 0.4) | 0.6 (± 0.4) | 77.2 (± 40.7) | 1 | |||
| ACS® | 7.5 | 12.5 (± 1.3) | 5.0 (± 1.3) | 0.019 | 310.3 (± 93.9) | 39.4 | 0.001 | 2 |
| ACS® Titannitrid (TiN) | 7.5 | 12.2 (± 1.1) | 4.7 (± 1.1) | 270.9 (± 79.5) | 3 | |||
| Vanguard™ | 9 | 12.8 (± 1.0) | 3.8 (± 1.0) | n.s. | 319.1 (± 113.0) | 9.5 | n.s. | 2 |
| Vanguard™ Titan-Niob-Nitrid (TiNbN) | 9 | 12.9 (± 0.8) | 3.9 (± 0.8) | 309.6 (± 98.3) | 2.5 | |||
| Columbus® | 7.4 | 10.4 (± 0.9) | 3.0 (± 0.9) | n.s. | 279.1 (± 99.2) | 17.7 | n.s. | 2 |
| AS Columbus® | 7.4 | 10.1 (± 0.5) | 2.7 (± 0.5) | 296.8 (± 124.0) | 3 | |||
| Legion™ | 11.6 | 19.0 (± 0.7) | 7.4 (± 0.7) | < 0.001 | 315.3 (± 71.9) | 416.0 | < 0.001 | 4 |
| Legion™ Oxinium® | 11.6 | 17.7 (± 0.3) | 6.1 (± 0.3) | 731.3 (± 88.5) | 5 | |||
| NexGen® LPS | 11.6 | 14.5 (± 1.0) | 2.9 (± 1.0) | < 0.001 | 288.3 (± 73.9) | 208.8 | < 0.001 | 5 |
| NexGen® Tivanium® LPS | 11.6 | 12.8 (± 0.3) | 1.2 (± 0.3) | 79.5 (± 27.7) | 1 | |||
| BPK-S | 6.1 | 9.7 (± 0.4) | 3.6 (± 0.4) | < 0.001 | 277.2 (± 121.3) | 191.2 | < 0.001 | 4 |
| BPK-S ceramic knee | 6.1 | 8.3 (± 0.2) | 2.2 (± 0.2) | 86.0 (± 34.5) | 1 | |||
| Scorpio® | 7.7 | 12.1 (± 1.0) | 4.4 (± 1.0) | 0.001 | 358.5 (± 136.7) | 31.9 | 0.003 | 3 |
| Scorpio® Titan | 7.7 | 12.5 (± 1.1) | 4.8 (± 1.1) | 326.6 (± 155.2) | 3.5 | |||
mm millimeter, HU Hounsfield unit
aDiameter, maximum thickness of the posterior condyles
bFive-point Likert scale (1–5, 1 indicating the best result)
Fig. 3CT images of all vendor prostheses pairs showing the extent of the artifacts. For all images in this gallery, scan protocol 7 (dual-energy CT with mono-energetic imaging at 105 keV) was used. It provided the lowest amount of streak artifacts of all protocols
Fig. 4Streak artifacts for all nine scan protocols as a mean over all prostheses: the fewest streak artifacts were observed for scan protocol 7. Dual-energy CT scan protocols (DE 1–7) offered fewer streak artifacts compared to single-energy CT protocols (SE 8 and 9)