Design and nonlinear modeling of a sensitive sensor for the measurement of flow in mice.

Samer Bou Jawde1, Bradford J Smith, Adam Sonnenberg, Jason H T Bates, Béla Suki.   

Abstract

OBJECTIVE: While many studies rely on flow and pressure measurements in small animal models of respiratory disease, such measurements can however be inaccurate and difficult to obtain. Thus, the goal of this study was to design and implement an easy-to-manufacture and accurate sensor capable of monitoring flow. APPROACH: We designed and 3D printed a flowmeter and utilized parametric (resistance and inertance) and nonparametric (polynomial and Volterra series) system identification to characterize the device. The sensor was tested in a closed system for apparent flow using the common mode rejection ratio (CMRR). The sensor properly measured tidal volumes and respiratory rates in spontaneously breathing mice. The device was used to evaluate a ventilator's ability to deliver a prescribed volume before and after lung injury. MAIN
RESULTS: The parametric and polynomial models provided a reasonable prediction of the independently measured flow (Adjusted coefficient of determination [Formula: see text]  =  0.9591 and 0.9147 respectively), but the Volterra series of the 1st, 2nd, and 3rd order with a memory of six time points provided better fits ([Formula: see text]  =  0.9775, 0.9787, and 0.9954, respectively). At and below the mouse breathing frequency (1-5 Hz), CMRR was higher than 40 dB. Following lung injury, the sensor revealed a significant drop in delivered tidal volume. SIGNIFICANCE: We demonstrate that the application of nonparametric nonlinear Volterra series modeling in combination with 3D printing technology allows the inexpensive and rapid fabrication of an accurate flow sensor for continuously measuring small flows in various physiological conditions.

Entities:  

Mesh:

Year:  2018        PMID: 29877866      PMCID: PMC6067907          DOI: 10.1088/1361-6579/aacb1b

Source DB:  PubMed          Journal:  Physiol Meas        ISSN: 0967-3334            Impact factor:   2.833


  18 in total

1.  Pseudorandom signals to estimate apparent transfer and coherence functions of nonlinear systems: applications to respiratory mechanics.

Authors:  B Suki; K R Lutchen
Journal:  IEEE Trans Biomed Eng       Date:  1992-11       Impact factor: 4.538

2.  Non-invasive intraoperative monitoring of blood pressure and arterial pCO2 during surgical anesthesia in mice.

Authors:  Serge C Thal; Nikolaus Plesnila
Journal:  J Neurosci Methods       Date:  2006-09-01       Impact factor: 2.390

3.  Evaluation of respiratory system mechanics in mice using the forced oscillation technique.

Authors:  Toby K McGovern; Annette Robichaud; Liah Fereydoonzad; Thomas F Schuessler; James G Martin
Journal:  J Vis Exp       Date:  2013-05-15       Impact factor: 1.355

Review 4.  The identification of nonlinear biological systems: Wiener kernel approaches.

Authors:  M J Korenberg; I W Hunter
Journal:  Ann Biomed Eng       Date:  1990       Impact factor: 3.934

5.  Design and calibration of unicapillary pneumotachographs.

Authors:  A Giannella-Neto; C Bellido; R B Barbosa; M F Melo
Journal:  J Appl Physiol (1985)       Date:  1998-01

6.  A computer-controlled research ventilator for small animals: design and evaluation.

Authors:  T F Schuessler; J H Bates
Journal:  IEEE Trans Biomed Eng       Date:  1995-09       Impact factor: 4.538

7.  In-phase rejection requirements for measuring respiratory input impedance.

Authors:  R Peslin; P Jardin; C Duvivier; P Begin
Journal:  J Appl Physiol Respir Environ Exerc Physiol       Date:  1984-03

8.  Calibration of pneumotachographs using a calibrated syringe.

Authors:  Yongquan Tang; Martin J Turner; Johnny S Yem; A Barry Baker
Journal:  J Appl Physiol (1985)       Date:  2003-04-18

9.  A simple and novel method to monitor breathing and heart rate in awake and urethane-anesthetized newborn rodents.

Authors:  Christoph M Zehendner; Heiko J Luhmann; Jenq-Wei Yang
Journal:  PLoS One       Date:  2013-05-03       Impact factor: 3.240

Review 10.  Measuring the lung function in the mouse: the challenge of size.

Authors:  Charles G Irvin; Jason H T Bates
Journal:  Respir Res       Date:  2003-05-15
View more
  2 in total

1.  Tracking respiratory mechanics around natural breathing rates via variable ventilation.

Authors:  Samer Bou Jawde; Allan J Walkey; Arnab Majumdar; George T O'Connor; Bradford J Smith; Jason H T Bates; Kenneth R Lutchen; Béla Suki
Journal:  Sci Rep       Date:  2020-04-21       Impact factor: 4.379

2.  A Damaged-Informed Lung Ventilator Model for Ventilator Waveforms.

Authors:  Deepak K Agrawal; Bradford J Smith; Peter D Sottile; David J Albers
Journal:  Front Physiol       Date:  2021-10-01       Impact factor: 4.566

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.