| Literature DB >> 29875727 |
Kirsten A Dalrymple1, Marie D Manner2, Katherine A Harmelink3, Elayne P Teska4, Jed T Elison1,5.
Abstract
The quantitative assessment of eye tracking data quality is critical for ensuring accuracy and precision of gaze position measurements. However, researchers often report the eye tracker's optimal manufacturer's specifications rather than empirical data about the accuracy and precision of the eye tracking data being presented. Indeed, a recent report indicates that less than half of eye tracking researchers surveyed take the eye tracker's accuracy into account when determining areas of interest for analysis, an oversight that could impact the validity of reported results and conclusions. Accordingly, we designed a calibration verification protocol to augment independent quality assessment of eye tracking data and examined whether accuracy and precision varied between three age groups of participants. We also examined the degree to which our externally quantified quality assurance metrics aligned with those reported by the manufacturer. We collected data in standard laboratory conditions to demonstrate our method, to illustrate how data quality can vary with participant age, and to give a simple example of the degree to which data quality can differ from manufacturer reported values. In the sample data we collected, accuracy for adults was within the range advertised by the manufacturer, but for school-aged children, accuracy and precision measures were outside this range. Data from toddlers were less accurate and less precise than data from adults. Based on an a priori inclusion criterion, we determined that we could exclude approximately 20% of toddler participants for poor calibration quality quantified using our calibration assessment protocol. We recommend implementing and reporting quality assessment protocols for any eye tracking tasks with participants of any age or developmental ability. We conclude with general observations about our data, recommendations for what factors to consider when establishing data inclusion criteria, and suggestions for stimulus design that can help accommodate variability in calibration. The methods outlined here may be particularly useful for developmental psychologists who use eye tracking as a tool, but who are not experts in eye tracking per se. The calibration verification stimuli and data processing scripts that we developed, along with step-by-step instructions, are freely available for other researchers.Entities:
Keywords: calibration; development; eye tracking; methods; quality assessment; toddlers
Year: 2018 PMID: 29875727 PMCID: PMC5974590 DOI: 10.3389/fpsyg.2018.00803
Source DB: PubMed Journal: Front Psychol ISSN: 1664-1078
Selected list of infant studies that provide accuracy and/or precision data.
| Study | Tracker | Method |
|---|---|---|
| SensoMotoric Instruments IView, 60 Hz | 6-, 9-, and 12-month-old infants at high genetic risk for ASD and low risk controls. | |
| Tobii ET – 17 | Excluded data from infants (3-, 6-, and 9-months-old) with average point of gaze >2° from central fixation point. | |
| Tobii T60XL | 4-point offline calibration to verify and correct 3–30 month old infants and toddlers. | |
| Tobii TX300 | Explicitly tested accuracy and precision of eye tracking in 10-month-olds under various testing conditions. | |
| Tobii T60XL | 9-point procedure evaluating spatial and temporal accuracy in 3–18 month olds. | |
| TX300 | Explicitly tested accuracy and precision of eye tracking in 9-, 12-, 15-month-olds. | |
| Various | Explore and discuss the effect of eye tracking data quality on fixation duration estimates in infants and adults. Discuss unique challenges associated with infant eye tracking data. |
Accuracy and precision (degrees of visual angle) for adults, school-age children, and toddlers, by target location.
| Target Location | |||||||
|---|---|---|---|---|---|---|---|
| Group | Measure | Upper Left | Upper Right | Middle | Lower Left | Lower Right | Mean |
| Adults | Accuracy (range) | 0.82 (0.37–1.41) | 0.73 (0.32–1.64) | 0.87 (0.16–3.16) | 0.78 (0.17–1.35) | 0.69 (0.28–1.30) | 0.78 (0.35–1.52) |
| Precision SD | 0.15, 0.17 | 0.10, 0.19 | 0.09, 0.16 | 0.12, 0.19 | 0.11, 0.19 | 0.11, 0.18 | |
| Precision RMS | 0.11, 0.13 | 0.11, 0.13 | 0.10, 0.15 | 0.13, 0.20 | 0.12, 0.22 | 0.11, 0.17 | |
| School-age | Accuracy (range) | 0.84 (0.20–1.93) | 1.37 (0.30–4.47) | 0.48 (0.07–0.84) | 0.63 (0.02–2.40) | 0.54 (0.11–1.41) | 0.93 (0.37–2.70) |
| Precision SD | 0.15, 0.18 | 0.13, 0.20 | 0.12, 0.17 | 0.15, 0.20 | 0.11, 0.18 | 0.14, 0.19 | |
| Precision RMS | 0.15, 0.17 | 0.14, 0.18 | 0.11, 0.15 | 0.16, 0.21 | 0.11, 0.21 | 0.15, 0.19 | |
| Toddlers 18 months | Accuracy (range) | 1.09 (0.16–2.04) | 1.26 (0.18–2.11) | 1.01 (0.19 –1.78) | 1.93 (0.44–11.37) | 2.22 (0.52–12.16) | 1.31 (0.18–3.85) |
| Run 1 | Precision SD | 0.19, 0.22 | 0.19, 0.21 | 0.19, 0.19 | 0.19, 0.21 | 0.19, 0.22 | 0.20, 0.21 |
| Precision RMS | 0.25, 0.24 | 0.23, 0.23 | 0.23, 0.22 | 0.24, 0.29 | 0.24, 0.26 | 0.25, 0.25 | |
| Run 2 | Accuracy (range) | 1.46 (0.34–4.26) | 1.34 (0.50–2.49) | 1.31 (0.40–2.04) | 0.85 (0.24–2.09) | 1.03 (0.22–2.21) | 1.28 (0.56–2.32) |
| Precision SD | 0.20, 0.21 | 0.18, 0.20 | 0.18, 0.19 | 0.17, 0.21 | 0.20, 0.20 | 0.20, 0.21 | |
| Precision RMS | 0.24, 0.23 | 0.23, 0.24 | 0.22, 0.22 | 0.24, 0.27 | 0.22, 0.26 | 0.25, 0.26 | |
| Toddlers 30 months | Accuracy (range) | 1.32 (0.35–3.79) | 1.33 (0.06–3.27) | 1.19 (0.30–2.23) | 1.24 (0.22–1.93) | 1.18 (0.41–2.78) | 1.29 (0.67–2.33) |
| Run 1 | Precision SD | 0.18, 0.21 | 0.19, 0.21 | 0.17, 0.21 | 0.20, 0.22 | 0.18, 0.22 | 0.19, 0.21 |
| Precision RMS | 0.22, 0.23 | 0.22, 0.25 | 0.22, 0.27 | 0.21, 0.27 | 0.22, 0.26 | 0.23, 0.27 | |
| Run 2 | Accuracy (range) | 1.27 (0.64–2.08) | 1.54 (0.61–2.44) | 1.78 (0.24–6.51) | 2.06 (0.41–6.41) | 2.14 (1.18–6.66) | 1.77 (0.81–2.27) |
| Precision SD | 0.18, 0.22 | 0.19, 0.19 | 0.15, 0.19 | 0.16, 0.22 | 0.17, 0.19 | 0.17, 0.20 | |
| Precision RMS | 0.22, 0.23 | 0.23, 0.22 | 0.19, 0.21 | 0.19, 0.27 | 0.18, 0.24 | 0.21, 0.24 | |
| Tobii TX300 | Accuracy | 0.4–0.9♦ | |||||
| Manufacturer specifications | Precision RMS | 0.04–0.15• | |||||