Literature DB >> 29866138

Capillary and venous lactate measurements with a handheld device compared to venous blood-gas analysis for emergency patients.

David Stoll1, Erling Englund2, Helene Hillborg2, Stefan Vedin3, Agneta Larsson4.   

Abstract

BACKGROUND/AIM: Early identification of lactate levels may have a large impact on triage classification and assist in identifying critically ill patients. A handheld device provides a rapid and timesaving measurement of lactate levels adapted to work in a prehospital care setting. I.e., the device is small, fast, and easy-to-use. The aim of this study was to evaluate the Accutrend Plus handheld lactate analyzer in comparison to the reference in-hospital method.
METHODS: Patients triaged as minimum yellow according to the RETTS System (Rapid Emergency Triage and Treatment System) and transported to hospital by ambulance were selected and a written consent to participate was obtained prior to inclusion in the study. Capillary (CAP) and venous (VEN) blood were analyzed with Accutrend Plus (AP). Venous blood samples were analyzed at the local hospital laboratory (GEM premier 4000) within 20 min from sampling. All sampling was conducted by two registered nurses specially trained in prehospital care.
RESULTS: 480 lactate measurements were performed in 160 patients. The mean difference between measurements in capillary blood compared with the reference method was 0.7 mmol/L and for venous blood 0.9 mmol/L. The limits of agreement from the Bland-Altman plot was - 0.9 to + 2.5 mmol/L and and - 0.1 to + 1.9 mmol/L, for CAP and VEN compared with GEM.
CONCLUSION: Our results shows low accuracy and low precision with VEN / CAP measurements of lactate compared to reference GEM.

Entities:  

Keywords:  Handheld lactate analyzer; Lactate; Point of care (POC); Prehospital emergency care

Mesh:

Substances:

Year:  2018        PMID: 29866138      PMCID: PMC5987430          DOI: 10.1186/s13049-018-0510-5

Source DB:  PubMed          Journal:  Scand J Trauma Resusc Emerg Med        ISSN: 1757-7241            Impact factor:   2.953


  10 in total

1.  Point-of-care testing at triage decreases time to lactate level in septic patients.

Authors:  Munish Goyal; Jesse M Pines; Byron C Drumheller; David F Gaieski
Journal:  J Emerg Med       Date:  2008-07-09       Impact factor: 1.484

2.  Correlation of arterial, central venous and capillary lactate levels in septic shock patients.

Authors:  Pattharawin Pattharanitima; Surat Tongyoo; Ranistha Ratanarat; Warakarn Wilachone; Aekarin Poompichet; Chairat Permpikul
Journal:  J Med Assoc Thai       Date:  2011-02

3.  Sample size and optimal designs for reliability studies.

Authors:  S D Walter; M Eliasziw; A Donner
Journal:  Stat Med       Date:  1998-01-15       Impact factor: 2.373

4.  Evaluation of the Lactate Pro blood lactate analyser.

Authors:  D B Pyne; T Boston; D T Martin; A Logan
Journal:  Eur J Appl Physiol       Date:  2000-05       Impact factor: 3.078

5.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

6.  Capillary and Venous Lactate Agreement: a pilot prospective observational study.

Authors:  Deepankar Datta; Julia Grahamslaw; Alasdair James Gray; Catriona Graham
Journal:  Emerg Med J       Date:  2016-09-21       Impact factor: 2.740

7.  Medical Emergency Triage and Treatment System (METTS): a new protocol in primary triage and secondary priority decision in emergency medicine.

Authors:  Bengt R Widgren; Majid Jourak
Journal:  J Emerg Med       Date:  2008-10-18       Impact factor: 1.484

8.  Validation of a hand-held lactate device in determination of blood lactate in critically injured patients.

Authors:  B M Slomovitz; R F Lavery; B J Tortella; J H Siegel; B L Bachl; A Ciccone
Journal:  Crit Care Med       Date:  1998-09       Impact factor: 7.598

9.  Accuracy, reliability, linearity of Accutrend and Lactate Pro versus EBIO plus analyzer.

Authors:  Carlo Baldari; Valerio Bonavolontà; Gian Pietro Emerenziani; Maria Chiara Gallotta; Antonio José Silva; Laura Guidetti
Journal:  Eur J Appl Physiol       Date:  2009-06-14       Impact factor: 3.078

10.  The prognostic value of blood lactate levels relative to that of vital signs in the pre-hospital setting: a pilot study.

Authors:  Tim C Jansen; Jasper van Bommel; Paul G Mulder; Johannes H Rommes; Selma J M Schieveld; Jan Bakker
Journal:  Crit Care       Date:  2008-12-17       Impact factor: 9.097

  10 in total
  2 in total

1.  Point of Care Testing of Serum Electrolytes and Lactate in Sick Children.

Authors:  Aashima Dabas; Shipra Agrawal; Vernika Tyagi; Shikha Sharma; Vandana Rastogi; Urmila Jhamb; Pradeep Kumar Dabla
Journal:  EJIFCC       Date:  2021-06-29

2.  Recommendations for hemodynamic monitoring for critically ill children-expert consensus statement issued by the cardiovascular dynamics section of the European Society of Paediatric and Neonatal Intensive Care (ESPNIC).

Authors:  Yogen Singh; Javier Urbano Villaescusa; Eduardo M da Cruz; Shane M Tibby; Gabriella Bottari; Rohit Saxena; Marga Guillén; Jesus Lopez Herce; Matteo Di Nardo; Corrado Cecchetti; Joe Brierley; Willem de Boode; Joris Lemson
Journal:  Crit Care       Date:  2020-10-22       Impact factor: 9.097

  2 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.