| Literature DB >> 29850207 |
Shino Sato1, Kaori Ukegawa1, Eri Nitta1, Kazuyuki Hirooka1.
Abstract
PURPOSE: To examine the influence of optic disc size on the diagnostic accuracy of optic nerve head (ONH) parameters determined by Cirrus spectral-domain optical coherence tomography (Cirrus HD-OCT).Entities:
Year: 2018 PMID: 29850207 PMCID: PMC5925212 DOI: 10.1155/2018/5692404
Source DB: PubMed Journal: J Ophthalmol ISSN: 2090-004X Impact factor: 1.909
Clinical characteristics in study participants.
| Normal | Glaucoma |
| |
|---|---|---|---|
|
| 51 | 71 | |
| Age (y) | 61.0 ± 10.1 | 64.5 ± 11.7 | 0.09 |
| Gender (M/F) | 24/27 | 25/46 | 0.19 |
| MD (dB) | −6.85 ± 7.66 | ||
| VFI (%) | 78.4 ± 22.1 | ||
| Disc size (mm2) | 1.89 ± 0.36 | 1.92 ± 0.45 | 0.64 |
| Spherical equivalent (D) | −1.26 ± 1.98 | −1.00 ± 2.30 | 0.51 |
M: male; F: female; MD: mean deviation; VFI: visual field index.
Figure 1Distribution of disease severity, as measured by the VFI in the glaucoma group.
Figure 2Distribution of optic disc size in the entire cohort.
Sensitivity at fixed specificity of optic nerve head measured by Cirrus HD-OCT in control and in glaucoma.
| All discs | Small discs | Large discs | |||||
|---|---|---|---|---|---|---|---|
| Sensitivity at specificity > 90% | Cutoff | Sensitivity at specificity > 90% | Cutoff | Sensitivity at specificity > 90% | Cutoff | Sensitivity CoV (%) | |
| Rim area | 0.92 | 0.96 ∗ | 0.88 | 0.88∗ | 0.92 | 0.96 ∗ | 2.8 |
| Average C/D | 0.68 | 0.74 | 0.71 | 0.69 | 0.76 | 0.76 | 4.9 |
| Vertical C/D | 0.83 | 0.71 | 0.76 | 0.71 | 0.89 | 0.78 | 10.9 |
| Cup volume | 0.37 | 0.45# | 0.5 | 0.33# | 0.24 | 0.73# | 48.9 |
CoV: coefficient of variation of sensitivity across small and large disc groups; ∗: mm2; #: mm3.
Factors associated with the diagnostic accuracy of disc size.
|
| |
|---|---|
| Rim area | 0.17 |
| Average C/D | <0.01 |
| Vertical C/D | 0.02 |
| Cup volume | <0.01 |