Gonzalo de la Iglesia1, André Walter1, Fernando de la Iglesia1, Heinz Winsauer2, Andreu Puigdollers1.
Abstract
INTRODUCTION: The force applied to the teeth by fixed orthopaedic expanders has previously been studied, but not the force applied to the orthodontic mini-implant (OMI) used to expand the maxilla with Hyrax hybrid expanders (HHE).Entities:
Mesh:
Year: 2018 PMID: 29791684 PMCID: PMC5962246 DOI: 10.1590/2177-6709.23.1.037-045.oar
Source DB: PubMed Journal: Dental Press J Orthod ISSN: 2176-9451
OMIs dimensions and abutment fixations types.
| Outer diameter | Inner diameter | Total length | Abutment type | Abutment fixation | |
| Jeil & Tiger | 2.5 mm | 1.8 mm | 14.5 mm | collar | Bonded / resin |
| Microdent | 1.6 mm | 1.1 mm | 15.0 mm | cap | Screwed-on |
| Ortholox | 2.2 mm | 1.5 mm | 14.0 mm | Push-button clipped cap | Snap-in coupling system |
Figure 2Illustration of the FTS.
Figure 3X-ray sequence (initial, 2 mm, 3 mm and 6 mm) of Jeil & Tiger, Microdent and Ortholox (left to the right) with double fastening wire.
Figure 4D1 and D2 distances; angle 1 and angle 2.
Loads, deformations and displacements of the 3 FTS.
| P values * | P values ** | P values ** | P values ** | |||||
| J & T | Microdent | Ortholox | J & T vs. Md vs. Orthx | J & T vs. Md | J & T vs. Orthx | Md vs. Orthx | ||
| 1-mm load (n=10) | Newton | 183 ± 48 | 67 ± 13 | 76 ± 5 | 0 | 0 | 0 | 0.533 |
| 3-mm load (n=10) | Newton | 323 ± 32 | 160 ± 14 | 109 ± 4 | 0 | 0 | 0.001 | 0.001 |
| 6-mm load (n=10) | Newton | 378 ± 22 | 201 ± 18 | 102 ±10 | 0 | 0 | 0 | 0 |
| Distance 1 (n=5) | mm | -1 ± 0.6 | 0 ± 0 | -0.5 ± 0.4 | 0.014 | 0.008 | 1 | 0.032 |
| Distance 2 (n=5) | mm | 1.96 ± 0.2 | 1.56 ± 0.2 | 1.68 ± 0.1 | 0.013 | 0.016 | 0.032 | 0.31 |
| α1 angle (n=5) | degrees | 21 ± 1.6 | 15 ± 2.2 | 10 ± 1.8 | 0.002 | 0.008 | 0.008 | 0.016 |
| α2 angle (n=5) | degrees | 8.4 ± 1.7 | 4.2 ± 1.5 | 0 ± 0 | 0.002 | 0.011 | 0.005 | 0.005 |
J & T = Jeil Medical and Tiger Dental, Md = Microdent, Orthx = Ortholox.
* For Kruskal-Wallis test (J & T vs. Md vs. Othx), p < 0.05 was statistically significant.
** For Mann-Whitney U test with Bonferroni correction, p < 0.017 was statistically significant.
Figure 5Tests performed for the three FTS: A) Jeil Medical & Tiger Dental, B) Microdent and C) Ortholox.