| Literature DB >> 29775657 |
Nathan K Itoga1, Ga-Young Suh1, Christopher P Cheng2.
Abstract
The Nellix device uses polymer-filled endobags to stabilize the abdominal aortic aneurysm (AAA) sac which is described as endovascular aneurysm sealing (EVAS). We analyzed cardiac-gated computed tomography angiography scans of repaired AAA with EVAS in 4 patients to evaluate the geometry and cardiac pulsatility-induced deformation. Graft translation and aortic curvature changes were found to be minimal during the cardiac cycle. The mean ± standard deviation changes in renal-aorta angles (1.0 ± 0.9°) were less than the changes in the superior mesenteric artery-aorta angle (4.0 ± 2.1°) (P < 0.01), during the cardiac cycle, demonstrating greater stabilization of the visceral branches closer to the device. These findings confirm stabilization of the abdominal aorta during the cardiac cycle using EVAS.Entities:
Mesh:
Year: 2018 PMID: 29775657 PMCID: PMC6131040 DOI: 10.1016/j.avsg.2018.02.039