Marco Germanotta1, Arianna Cruciani2, Cristiano Pecchioli2, Simona Loreti2,3, Albino Spedicato2, Matteo Meotti2, Rita Mosca2, Gabriele Speranza2, Francesca Cecchi4, Giorgia Giannarelli2, Luca Padua2,5, Irene Aprile2.
Abstract
BACKGROUND: In the last few years, there has been an increasing interest in the use of robotic devices to objectively quantify motor performance of patients after brain damage. Although these robot-derived measures can potentially add meaningful information about the patient's dexterity, as well as be used as outcome measurements after the rehabilitation treatment, they need to be validated before being used in clinical practice. The present work aims to evaluate the reliability, the validity and the discriminant ability of the metrics provided by a novel robotic device for upper limb rehabilitation.Entities:
Keywords: Discriminant ability; Reliability; Robot-mediated evaluation; Stroke; Upper limb; Validity
Mesh:
Year: 2018 PMID: 29769127 PMCID: PMC5956822 DOI: 10.1186/s12984-018-0385-8
Source DB: PubMed Journal: J Neuroeng Rehabil ISSN: 1743-0003 Impact factor: 4.262
Demographic and clinical characteristics of the sample
| Patients with stroke ( | Healthy subjects ( | |
|---|---|---|
| Sex M/F | 33/15 | 26/14 |
| Age, mean ± SD (years) | 64 ± 11 | 65 ± 13 |
| Classification | ||
| Cerebral ischemia (N) | 28 | – |
| Cerebral hemorrhage (N) | 20 | – |
| Time from lesion, mean ± SD (days) | 88 ± 42 | – |
| FMA-UL, mean ± SD | 29 ± 18 | – |
| ARAT, mean ± SD | 15 ± 18 | – |
| BI, mean ± SD | 40 ± 24 | – |
SD Standard Deviation, FMA-UL Fugl-Meyer Assessment for the Upper Limb, ARAT Action Research Arm Test, BI Barthel Index
Fig. 1Patient engaged in a rehabilitation session with MOTORE
Fig. 2The evaluation task of MOTORE. In figure is showed the visual feedback showed to the patients on the screen, together with the position of each target. The white ball indicates the position of the end-effector; the yellow circle indicates the target to be reached. The yellow squares, not showed to the patient during the task, indicate the position of the targets: C is the central target, while the numbers from 1 to 8 indicate the external targets with the sequence of the center-out movements. In addition, the distance of each target from the center is reported
Outcome measures provided by MOTORE
| Index | Description |
|---|---|
| Duration | Time required to complete the task |
| Velocitymean | Average velocity of the device during the test |
| Lengthtot | Global length of the path travelled by the subject during center-out movements; it ranges from 0 (no movement) to 2.808 m (patient can fully perform the entire task) |
| Lengthi | Length of the path travelled by the subject toward the i-th target ( |
| Score | Mean of the ratios between the actual distance covered by the patients and the required distance to be travelled, computed for each required movement. |
| Worktot | Line integral of the force along the path described by the patient |
| Worktan | The amount of total work directed towards the target |
Test-retest reliability in stroke patients (n = 48)
| Test mean (SD) | Retest mean (SD) | ICC | 95% CI | Paired t test (P) | ||
|---|---|---|---|---|---|---|
| Lower bound | Upper bound | |||||
| Duration (s) | 193.7 (107.30) | 176.8 (111.90) | 0.962 | 0.922 | 0.980 |
|
| Velocitymean (m/s) | 0.05 (0.04) | 0.07 (0.04) | 0.914 | 0.756 | 0.962 |
|
| Lengthtot (m) | 1.80 (0.92) | 1.84 (1.01) | 0.951 | 0.912 | 0.972 | 0.495 |
| Lenght1 (m) | 0.25 (0.16) | 0.25 (0.18) | 0.930 | 0.876 | 0.960 | 0.784 |
| Lenght2 (m) | 0.38 (0.14) | 0.35 (0.17) | 0.804 | 0.652 | 0.890 | 0.149 |
| Lenght3 (m) | 0.13 (0.03) | 0.13 (0.02) | 0.693 | 0.456 | 0.828 | 0.542 |
| Lenght4 (m) | 0.34 (0.18) | 0.36 (0.17) | 0.917 | 0.851 | 0.953 | 0.113 |
| Lenght5 (m) | 0.24 (0.17) | 0.25 (0.18) | 0.907 | 0.834 | 0.948 | 0.551 |
| Lenght6 (m) | 0.20 (0.17) | 0.22 (0.18) | 0.917 | 0.852 | 0.953 | 0.13 |
| Lenght7 (m) | 0.05 (0.04) | 0.06 (0.04) | 0.957 | 0.924 | 0.976 | 0.188 |
| Lenght8 (m) | 0.21 (0.17) | 0.22 (0.18) | 0.934 | 0.883 | 0.963 | 0.467 |
| Score | 7.91 (2.20) | 7.99 (2.45) | 0.972 | 0.949 | 0.984 | 0.477 |
| Worktot (J) | 19.88 (12.75) | 20.79 (15.45) | 0.908 | 0.837 | 0.949 | 0.446 |
| Worktan (J) | 10.22 (8.65) | 11.26 (0.18) | 0.957 | 0.922 | 0.976 | 0.061 |
Intraclass Correlation Coefficient (ICC) with 95% Confidence Interval (CI), and result of the t tests. Bold values indicated statistical significance, with p value less than 0.05
Fig. 3Bland-Altman plots of the robotic indices assessing the whole task
Fig. 4Bland-Altman plots of the robotic indices assessing the path length travelled by stroke patients towards each target
Fig. 5Intra-session reliability analysis in stroke patients: robotic indices assessing the whole task. Blue lines represent the statistical analysis of the first session (test), while green lines represent the statistical analysis of the second session (retest). The symbols *, ** and *** represent a statistically significant difference between repetitions, with a p value less than 0.05, 0.01 and 0.001, respectively
Fig. 6Intra-session reliability analysis in stroke patients: robotic indices assessing the path length travelled towards each target. Blue lines represent the statistical analysis of the first session (test), while green lines represent the statistical analysis of the second session (retest)
Test-retest reliability in healthy subjects (n = 19)
| Test mean (SD) | Retest mean (SD) | ICC | 95% CI | Paired t test (P) | ||
|---|---|---|---|---|---|---|
| Lower bound | Upper bound | |||||
| Duration (s) | 107.3 (56.30) | 83.96 (50.02) | 0.914 | 0.336 | 0.977 |
|
| Velocitymean (m/s) | 0.08 (0.04) | 0.10 (0.06) | 0.81 | 0.437 | 0.931 |
|
| Lengthtot (m) | 2.64 (0.29) | 2.68 (0.21) | 0.593 | −0.064 | 0.844 | 0.484 |
| Lenght1 (m) | 0.39 (0.03) | 0.39 (0.03) | 0.627 | −0.002 | 0.858 | 0.966 |
| Lenght2 (m) | 0.46 (0.08) | 0.47 (0.09) | 0.93 | 0.822 | 0.973 | 0.309 |
| Lenght3 (m) | 0.14 (0.01) | 0.14 (0.00) | § | § | § | 0.331 |
| Lenght4 (m) | 0.46 (0.10) | 0.48 (0.01) | 0.01 | −1.511 | 0.615 | 0.273 |
| Lenght5 (m) | 0.38 (0.06) | 0.39 (0.03) | 0.087 | −1.446 | 0.652 | 0.458 |
| Lenght6 (m) | 0.35 (0.07) | 0.35 (0.09) | 0.722 | 0.260 | 0.894 | 0.870 |
| Lenght7 (m) | 0.09 (0.01) | 0.09 (0.00) | § | § | § | 0.358 |
| Lenght8 (m) | 0.38 (0.04) | 0.38 (0.06) | 0.814 | 0.512 | 0.929 | 0.729 |
| Score | 9.68 (0.58) | 9.78 (0.37) | 0.522 | −0.249 | 0.816 | 0.453 |
| Worktot (J) | 17.62 (10.29) | 19.84 (8.21) | 0.695 | 0.227 | 0.882 | 0.296 |
| Worktan (J) | 13.17 (7.38) | 15.42 (6.61) | 0.868 | 0.632 | 0.950 |
|
Intraclass Correlation Coefficient (ICC) with 95% Confidence Interval (CI), and result of the t tests. Bold values indicated statistical significance, with p value less than 0.05. The symbol § indicate null variance in the data
Validity
| FMA-UL | ARAT | BI | |
|---|---|---|---|
| Duration | −0.8507*** | −0.7716*** | − 0.6738*** |
| Velocitymean | 0.8227*** | 0.7587*** | 0.6340*** |
| Lengthtot | 0.8551*** | 0.7268*** | 0.7139*** |
| Lenght1 | 0.7700*** | 0.6273*** | 0.6423*** |
| Lenght2 | 0.6259*** | 0.5904*** | 0.4419** |
| Lenght3 | 0.5093*** | 0.4613*** | 0.4195** |
| Lenght4 | 0.7172*** | 0.5689*** | 0.6021*** |
| Lenght5 | 0.8047*** | 0.6666*** | 0.6638*** |
| Lenght6 | 0.8026*** | 0.7075*** | 0.6914*** |
| Lenght7 | 0.8267*** | 0.6979*** | 0.6404*** |
| Lenght8 | 0.8584*** | 0.7196*** | 0.6241*** |
| Score | 0.8443*** | 0.7384*** | 0.7417*** |
| Worktot | 0.3472** | 0.3672** | 0.1700 |
| Worktan | 0.8188*** | 0.6942*** | 0.6564*** |
Spearman’s correlation coefficient between the robotic indices and the following clinical scale: Upper limb subscale of the Fugl-Meyer Assessment (FMA-UL), Action Research Arm Test (ARAT), and Barthel Index (BI). The symbols **, and *** indicate a p value less than 0.01 and 0.001 respectively
Discriminant ability
| Robotic indices | Patients with stroke ( | Healthy subjects ( | Unpaired t test (P) | Effect Size |
|---|---|---|---|---|
| Duration (s) | 193.7 (107.30) | 93.30 (53.50) | < 0.001 | 1.18 |
| Velocitymean (m/s) | 0.05 (0.04) | 0.09 (0.05) | < 0.001 | 0.86 |
| Lengthtot (m) | 1.80 (0.92) | 2.67 (0.24) | < 0.001 | 1.30 |
| Lenght1 (m) | 0.25 (0.16) | 0.39 (0.03) | < 0.001 | 1.17 |
| Lenght2 (m) | 0.38 (0.14) | 0.46 (0.08) | 0.001 | 0.75 |
| Lenght3 (m) | 0.13 (0.03) | 0.14 (0.01) | 0.004 | 0.62 |
| Lenght4 (m) | 0.34 (0.18) | 0.47 (0.07) | < 0.001 | 0.98 |
| Lenght5 (m) | 0.24 (0.17) | 0.39 (0.04) | < 0.001 | 1.20 |
| Lenght6 (m) | 0.20 (0.17) | 0.35 (0.08) | < 0.001 | 1.17 |
| Lenght7 (m) | 0.05 (0.04) | 0.09 (0.00) | < 0.001 | 1.31 |
| Lenght8 (m) | 0.21 (0.17) | 0.38 (0.05) | < 0.001 | 1.33 |
| Score | 7.91 (2.20) | 9.74 (0.47) | < 0.001 | 1.16 |
| Worktot (J) | 19.88 (12.75) | 19.03 (9.11) | 0.717 | – |
| Worktan (J) | 10.22 (8.65) | 14.62 (6.98) | 0.010 | 0.56 |
Descriptive statistics for the robotic indices in patients with stroke (N = 48) and healthy subjects (N = 40). Comparison is assessed by means of t tests. For significant differences, the Effect Size is also reported