Vesna Antunović1, Slavna Tešanović2, Danica Perušković2, Nikola Stevanović2, Rada Baošić2, Snežana Mandić2, Aleksandar Lolić2.
Abstract
This work presents the development of a flow injection system for differential pulse amperometry (DPA) for diazepam determination in the presence of oxygen. The thin flow cell consisted of the bare glassy carbon electrode, reference silver/silver chloride, and stainless steel as the auxiliary electrode. Electrochemical reduction of diazepam (DZP) was characterised by cyclic voltammetry. Azomethine reduction peak was used for DZP quantification. The detector response was linear in the range 20-250 µmol/dm3 of diazepam, with a calculated detection limit of 3.83 µg/cm3. Intraday and interday precision were 1.53 and 10.8%, respectively. The method was applied on three beverage samples, energetic drink, and two different beer samples, and obtained recoveries were from 93.65 up to 104.96%. The throughoutput of the method was up to 90 analyses per hour.Entities:
Year: 2018 PMID: 29744233 PMCID: PMC5883924 DOI: 10.1155/2018/6121489
Source DB: PubMed Journal: J Anal Methods Chem ISSN: 2090-8873 Impact factor: 2.193
Figure 1Cyclic voltammograms of 0.5 mol/dm3 diazepam on glassy carbon electrode as a function of a scan rate (20–150 mV s−1).
Figure 2Cyclic voltammograms of 100 µmol/dm3 diazepam as function of pH in HCl/KCl buffer (pH 1) and Britton–Robinson buffer (pH 3–8) at 100 mV s−1 scan rate.
Sensitivities at different E1 and E2 values for FIA-DPA.
|
|
| Slope (nA dm3 mol−1) |
|---|---|---|
| −0.6 | −0.8 | 35 |
| −0.65 | −0.8 | 42 |
| −0.7 | −0.8 | 31.3 |
| −0.75 | −0.8 | 14.2 |
| −0.75 | −0.9 | 38.8 |
| −0.75 | −0.95 | 92.7 |
Figure 3Schematic presentation of DPA waveform applied for diazepam determination. The squares are used for current measurement.
Figure 4Linear response for FIA-DPA system for 20–250 µmol/dm3 diazepam. Inset: six consecutive injections of 100 µmol/dm3 of diazepam.
Analytical characteristics of the proposed methods.
| Characteristics | FIA-DPA |
|---|---|
| LR, | 20–250 |
| Sensitivity, nA dm3 | 63 |
| Correlation coefficient, | 0.997 |
| LOD, | 13.4 |
| Intraday RSD ( | 1.53% |
| Interday RSD ( | 10.8% |
| Throughoutput, 1 h−1 | 90 |
Results of interference study.
| Interferent | Relative signal decrease (%) |
|---|---|
| Citric acid | 4 |
| Lactose | 5 |
| Ascorbic acid | 8 |
| Glucose | 10 |
Analysis results of the recovery for spiked samples (n=3).
| Sample | Spiked concentration ( | Results (mean ± SD) ( | Recovery ratio (%) |
|---|---|---|---|
| Guarana | 50 | 52.48 ± 0.80 | 104.96 |
| 100 | 101.25 ± 1.55 | 101.25 | |
| 150 | 152.05 ± 2.33 | 101.37 | |
| Beer I | 50 | 47.51 ± 0.73 | 95.02 |
| 100 | 94.31 ± 1.44 | 94.31 | |
| 150 | 139.48 ± 2.13 | 92.99 | |
| Beer II | 50 | 49.12 ± 0.75 | 98.24 |
| 100 | 96.24 ± 1.47 | 96.24 | |
| 150 | 141.66 ± 2.17 | 94.44 |