Literature DB >> 29741488

Long-term durability and haemodynamic performance of a self-expanding transcatheter heart valve beyond five years after implantation: a prospective observational study applying the standardised definitions of structural deterioration and valve failure.

Erik W Holy1, Julia Kebernik, Mohammad Abdelghani, Simon F Stämpfli, Jens Hellermann, Abdelhakim Allali, Mohamed El-Mawardy, Susanne Sachse, Thomas F Lüscher, Felix C Tanner, Gert Richardt, Mohamed Abdel-Wahab.   

Abstract

AIMS: Long-term results of transcatheter aortic valve implantation (TAVI), in particular the incidence of bioprosthetic valve failure (BVF), are uncertain. This study presents data derived from a long-term, structured follow-up programme of the self-expanding CoreValve device utilising standardised definitions and core lab adjudication of valve performance. METHODS AND
RESULTS: The study prospectively included all 152 patients who had undergone TAVI with the self-expanding CoreValve up to December 2011 at the Heart Center, Bad Segeberg, Germany. Late BVF (>30 days) was defined as either: 1) severe structural valve deterioration (transprosthetic mean pressure gradient ≥40 mmHg and/or ≥20 mmHg rise from baseline OR severe intraprosthetic aortic regurgitation), OR 2) bioprosthetic valve dysfunction leading to death or reintervention. Echocardiographic follow-up at 6.3±1.0 years (range: 5.0-8.9 years) was 88% complete (60 out of 68 survivors beyond five years) and all echocardiograms were analysed by an independent core laboratory. The all-cause mortality rate at 1, 2, 5, 6, 7 and 8 years was 14%, 20%, 50%, 60%, 65%, and 73%, respectively. Among survivors beyond five years, effective orifice area was 1.60±0.46 cm2, and transvalvular mean pressure gradient was 6.7±3.1 mmHg; no cases showed evidence of structural valve deterioration. Five patients (3.3%) had undergone redo TAVI (n=4) or surgery (n=1) 0.6 to 5.2 years after the index procedure, all due to paravalvular leakage. The estimated rate of BVF at eight years was 7.9% for the actuarial and 4.5% for the actual analysis.
CONCLUSIONS: Long-term follow-up up to 8.9 years after TAVI documents favourable performance of the self-expanding CoreValve with low rates of BVF.

Entities:  

Mesh:

Year:  2018        PMID: 29741488     DOI: 10.4244/EIJ-D-18-00041

Source DB:  PubMed          Journal:  EuroIntervention        ISSN: 1774-024X            Impact factor:   6.534


  9 in total

Review 1.  Valve durability after transcatheter aortic valve implantation.

Authors:  Akash Kataruka; Catherine M Otto
Journal:  J Thorac Dis       Date:  2018-11       Impact factor: 2.895

2.  10-Year Impact of Transcatheter Aortic Valve Replacement Leaflet Design (Intra- Versus Supra-Annular) in Mortality and Hemodynamic Performance.

Authors:  Andrea Scotti; Luca Nai Fovino; Augustin Coisne; Tommaso Fabris; Francesco Cardaioli; Mauro Massussi; Giulio Rodinò; Alberto Barolo; Mauro Boiago; Saverio Continisio; Carolina Montonati; Tommaso Sciarretta; Vittorio Zuccarelli; Valentina Bernardini; Giulia Masiero; Massimo Napodano; Chiara Fraccaro; Alfredo Marchese; Giovanni Esposito; Juan F Granada; Azeem Latib; Sabino Iliceto; Giuseppe Tarantini
Journal:  Front Cardiovasc Med       Date:  2022-06-08

3.  Long-term Follow-up After Transcatheter Aortic Valve Replacement.

Authors:  Stephan Haussig; Constantin Pleissner; Norman Mangner; Felix Woitek; Marion Zimmer; Philipp Kiefer; Florian Schlotter; Georg Stachel; Sergey Leontyev; David Holzhey; Michael A Borger; Axel Linke
Journal:  CJC Open       Date:  2021-02-01

Review 4.  TAVI for Pure Native Aortic Regurgitation: Are We There Yet?

Authors:  Eduardo A Arias; Amit Bhan; Zhan Y Lim; Michael Mullen
Journal:  Interv Cardiol       Date:  2019-02

Review 5.  Long-term Transcatheter Aortic Valve Durability.

Authors:  Giuliano Costa; Enrico Criscione; Denise Todaro; Corrado Tamburino; Marco Barbanti
Journal:  Interv Cardiol       Date:  2019-05-21

Review 6.  Transcatheter Bioprosthetic Aortic Valve Dysfunction: What We Know So Far.

Authors:  Fadi Sawaya; Troels H Jørgensen; Lars Søndergaard; Ole De Backer
Journal:  Front Cardiovasc Med       Date:  2019-10-04

7.  Structural Valve Deterioration after Transcatheter Aortic Valve Implantation Using J-Valve: A Long-Term Follow-Up.

Authors:  Fei Li; Xu Wang; Yuetang Wang; Fei Xu; Xin Wang; Xuan Li; Wei Wang
Journal:  Ann Thorac Cardiovasc Surg       Date:  2020-04-03       Impact factor: 1.520

8.  Degeneration of Bioprosthetic Heart Valves: Update 2020.

Authors:  Alexander E Kostyunin; Arseniy E Yuzhalin; Maria A Rezvova; Evgeniy A Ovcharenko; Tatiana V Glushkova; Anton G Kutikhin
Journal:  J Am Heart Assoc       Date:  2020-09-21       Impact factor: 5.501

9.  European heart health survey 2019.

Authors:  Luise Gaede; Marta Sitges; Johnson Neil; Eleonara Selvi; William Woan; Richard Derks; Helge Möllmann
Journal:  Clin Cardiol       Date:  2020-10-28       Impact factor: 2.882

  9 in total