| Literature DB >> 29732338 |
B Barati1, M Zabihzadeh1, M J Tahmasebi Birgani1,2, N Chegini1, J Fatahiasl1, I Mirr3.
Abstract
OBJECTIVE: The use of miniature X-ray source in electronic brachytherapy is on the rise so there is an urgent need to acquire more knowledge on X-ray spectrum production and distribution by a dose. The aim of this research was to investigate the influence of target thickness and geometry at the source of miniature X-ray tube on tube output.Entities:
Keywords: Electronic Brachytherapy ; Energy Spectrum ; Miniature Source; Target Optimization ; Monte Carlo
Year: 2018 PMID: 29732338 PMCID: PMC5928308
Source DB: PubMed Journal: J Biomed Phys Eng ISSN: 2251-7200
Figure1A view of miniature X-ray sources simulated for conducting the study
Figure2A view of the targets that had a size range of optimum thickness of sources
Figure3A view of the form used to determine the optimum thickness by measuring the dose
Optimum thicknesses obtained according to Figures 4, 5, and comparisons with other available references
| Energy | 20Kev | 30Kev | 40Kev | 50Kev |
|---|---|---|---|---|
| H-C a | 0.3 | 0.6 | 0.9 | 1.1 |
| Cb | 0.5 | 0.8 | 1.2 | 1.7 |
| Hc | 0.4 | 0.6 | 1.1 | 1.3 |
| T-Cd | 0.5 | 0.8 | 1.2 | 1.7 |
| Pe | 0.4 | 0.6 | 1.1 | 1.5 |
| Khajehf and etal 2015 | - | 0.65 | 0.85 | 1.45 |
| Safigholig and etal 2012 | 0.18 | 0.64 | 1.1 | 1.56 |
Optimal Thickness (µm) for a. Hemispherical -conical, b.conical, c.Hemispherical, d.Truncated -conical, e. Planer, f, g)
Figure4Diagrams obtained based on electron penetration , that show optimized thickness for the sources (first method)
Figure4
Figure4
Figure4
Figure5Diagrams of relative dose of the results of section of dose calculations to determine the optimum thickness and to validate the first
Figure5
Figure5
Figure5
Figure5
Figure6the diagram of the relationship between electron range with the release of energy within the target until it is stopped so that, the optimum thickness is less than the electron ranges and sources with larger optimum thickness, the electron range within them is far greater than the other sources
Figure7Shows spectra of 5 miniature X-ray source and suggests that the intensity of the planer source was higher than the rest and that the spectrum intensity declined sharply according to the increased distance from the source
Figure8Comparison of axial and transverse dose distribution of a miniature X-ray source