Characterization of a soft X-ray source for intravascular radiation therapy.

D P Gierga1, R E Shefer.   

Abstract

PURPOSE: A soft X-ray device for intravascular radiation therapy of restenosis is characterized in terms of dose delivery for several artery configurations, including arteries with implanted stents, calcified plaque, and noncentered sources. METHODS AND MATERIALS: The Monte Carlo code MCNP4B was used to determine the X-ray fluence and energy spectra for 15, 20, and 30-kV X-ray source generating voltages. Dose as a function of distance was calculated under a variety of artery conditions.
RESULTS: Calculated depth-dose profiles for the X-ray sources are within presumed artery dose tolerance limits for the range of generating voltages considered. Treatment times to deliver 8 Gy to the adventitia range from 2.7 minutes to 6.7 minutes for the 20-kV generating voltage and a 3-cm-long lesion, depending on the diameter of the artery. The does perturbation due to stent wires or calcified plaque is found to be more severe for the X-ray sources than for the radioactive sources. The effects of noncentering are found to be similar for radioactive sources and X-ray sources with generating voltages of 20 kV or higher.
CONCLUSION: The results of this study indicate that soft X-ray sources are suitable candidates for intravascular radiation therapy over a wide range of artery sizes, tissue compositions, and stent configurations.

Mesh:

Year:  2001        PMID: 11172969     DOI: 10.1016/s0360-3016(00)01510-8

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Effect of early external X-ray radiation on arterial restenosis post percutaneous transluminal angioplasty.

Authors:  Jun Cheng; Bujin Liu; Huarong Yu; Qining Fu; Fenghe Li; Yu Zhao
Journal:  Int J Clin Exp Med       Date:  2015-09-15

2.  The effect of early external X-ray radiation on arterial restenosis post percutaneous transluminal angioplasty.

Authors:  Jun Cheng; Bujin Liu; Huarong Yu; Qining Fu; Fenghe Li; Yu Zhao
Journal:  Int J Clin Exp Med       Date:  2015-07-15

3.  Assessment of two hemispherical and hemispherical-conical miniature sources used in electronic brachytherapy using Monte Carlo Simulation.

Authors:  Barat Barati; Mansour Zabihzadeh; Mohammad Javad Tahmasebi Birgani; Nahid Chegini; Mojtaba Hoseini Ghahfarokhi; Jafar Fatahiasl
Journal:  Electron Physician       Date:  2017-02-25

4.  Evaluation of the Effect of Source Geometry on the Output of Miniature X-ray Tube for Electronic Brachytherapy through Simulation.

Authors:  B Barati; M Zabihzadeh; M J Tahmasebi Birgani; N Chegini; J Fatahiasl; I Mirr
Journal:  J Biomed Phys Eng       Date:  2018-03-01
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.