Literature DB >> 29713783

A high rate of tibial plateau fractures after early experience with patient-specific instrumentation for unicompartmental knee arthroplasties.

A M Leenders1, M G M Schotanus2, R J P Wind3, R A P Borghans4, N P Kort5.   

Abstract

PURPOSE: Patient-specific instrumentation (PSI) for unicompartmental knee arthroplasty (UKA) has been available for a few years. However, limited literature is available on this subject. Hence, the aim of this cohort study is to evaluate the 2 years' results of our first experiences with the use of PSI in UKA. It is hypothesised that there is no advantage in rate of adverse events and in radiological and functional outcomes in comparison to literature on the conventional method.
METHODS: This cohort included 129 knees of 122 patients, operated by one surgeon. Outcome measures were the rate of adverse events (AEs); implant position as determined on radiographs; the accuracy of the default and approved planning of the implant sizes and the patient-reported outcome measures (PROMs) preoperatively, and at 3, 12 and 24 months, postoperatively.
RESULTS: A total of 6 (4.9%) AEs were observed in this study, with 4 (3.3%) tibial fractures being the main complication. The mean postoperative biomechanical axis was 176.4° and in the majority of cases, the radiographic criteria, as determined by the manufacturer, were met. The tibial component showed 20 (16.4%) outliers in the sagittal and 3 (2.5%) outliers in the frontal plane. There were no outliers of the femoral component. For the femoral and tibial components, respectively, in 125 (96.9%) and 79 (61.7%) cases, there was an agreement between approved planning and implanted component size. All PROMs improved significantly after surgery.
CONCLUSION: Tibial fracture was the most common AE, probably related to the transition from cemented to uncemented UKA. Perioperative modifications to the surgical technique were made in order to prevent this AE. Improvements should be made to the operation technique of the uncemented tibial plateau to obtain an adequate placement and at the same time reduce the risk for tibial fracture. The PSI technique was a reliable tool for the placement of the femoral component. Functional outcome was in line with literature on the conventional method. It is strongly recommended that the surgeon approves every preoperative plan, in order to optimise the accuracy during the PSI surgery. LEVEL OF EVIDENCE: III.

Entities:  

Keywords:  Custom fit; Instruments; PSG; PSI; Patient specific; Patient-specific guides; Patient-specific matched instruments; Templating; UKA; Unicompartmental knee arthroplasty; Unicondylar knee arthroplasty

Mesh:

Year:  2018        PMID: 29713783     DOI: 10.1007/s00167-018-4956-9

Source DB:  PubMed          Journal:  Knee Surg Sports Traumatol Arthrosc        ISSN: 0942-2056            Impact factor:   4.342


  29 in total

1.  Three-dimensional mechanics, kinematics, and morphology of the knee viewed in virtual reality.

Authors:  Donald G Eckhoff; Joel M Bach; Victor M Spitzer; Karl D Reinig; Michelle M Bagur; Todd H Baldini; Nicolas M P Flannery
Journal:  J Bone Joint Surg Am       Date:  2005       Impact factor: 5.284

2.  Patient-reported outcome measures versus inertial performance-based outcome measures: A prospective study in patients undergoing primary total knee arthroplasty.

Authors:  S A A N Bolink; B Grimm; I C Heyligers
Journal:  Knee       Date:  2015-05-29       Impact factor: 2.199

3.  Evaluation of implant position and knee alignment after patient-specific unicompartmental knee arthroplasty.

Authors:  Franz Xaver Koeck; Johannes Beckmann; Christian Luring; Bjoern Rath; Joachim Grifka; Erhan Basad
Journal:  Knee       Date:  2010-08-04       Impact factor: 2.199

4.  The Chitranjan Ranawat award: is neutral mechanical alignment normal for all patients? The concept of constitutional varus.

Authors:  Johan Bellemans; William Colyn; Hilde Vandenneucker; Jan Victor
Journal:  Clin Orthop Relat Res       Date:  2012-01       Impact factor: 4.176

5.  The John Insall Award: No Functional Benefit After Unicompartmental Knee Arthroplasty Performed With Patient-specific Instrumentation: A Randomized Trial.

Authors:  Matthieu Ollivier; Sebastien Parratte; Alexandre Lunebourg; Elke Viehweger; Jean-Noel Argenson
Journal:  Clin Orthop Relat Res       Date:  2016-01       Impact factor: 4.176

6.  Unsatisfactory outcomes following unicompartmental knee arthroplasty in patients with partial thickness cartilage loss: a medium-term follow-up.

Authors:  T W Hamilton; H G Pandit; A Inabathula; S J Ostlere; C Jenkins; S J Mellon; C A F Dodd; D W Murray
Journal:  Bone Joint J       Date:  2017-04       Impact factor: 5.082

7.  Initial experience with the oxford unicompartmental knee arthroplasty.

Authors:  Geoffrey F Dervin; Chris Carruthers; Robert J Feibel; Alan A Giachino; Paul R Kim; Peter R Thurston
Journal:  J Arthroplasty       Date:  2010-07-29       Impact factor: 4.757

8.  Improved positioning of the tibial component in unicompartmental knee arthroplasty with patient-specific cutting blocks.

Authors:  M L Dao Trong; C Diezi; G Goerres; N Helmy
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-01-17       Impact factor: 4.342

9.  No radiographic difference between patient-specific guiding and conventional Oxford UKA surgery.

Authors:  Bart Kerens; Martijn G M Schotanus; Bert Boonen; Nanne P Kort
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2014-01-26       Impact factor: 4.342

10.  Inertial sensor motion analysis of gait, sit-stand transfers and step-up transfers: differentiating knee patients from healthy controls.

Authors:  S A A N Bolink; S N van Laarhoven; M Lipperts; I C Heyligers; B Grimm
Journal:  Physiol Meas       Date:  2012-10-31       Impact factor: 2.833

View more
  3 in total

1.  What future in the treatment of osteochondral knee defects?

Authors:  Riccardo D'Ambrosi; Vincenza Ragone; Nicola Ursino
Journal:  Ann Transl Med       Date:  2018-12

Review 2.  [New technologies (robotics, custom-made) in unicondylar knee arthroplasty-pro].

Authors:  Malin Meier; Tilman Calliess; Carsten Tibesku; Johannes Beckmann
Journal:  Orthopade       Date:  2021-02       Impact factor: 1.087

Review 3.  Comparable incidence of periprosthetic tibial fractures in cementless and cemented unicompartmental knee arthroplasty: a systematic review and meta-analysis.

Authors:  Joost A Burger; Tjeerd Jager; Matthew S Dooley; Hendrik A Zuiderbaan; Gino M M J Kerkhoffs; Andrew D Pearle
Journal:  Knee Surg Sports Traumatol Arthrosc       Date:  2021-02-02       Impact factor: 4.342

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.