Clinical evaluation of a new correction algorithm for dynamic Scheimpflug analyzer tonometry before and after laser in situ keratomileusis and small-incision lenticule extraction.

Kai-Jung Chen1, Akram Joda1, Riccardo Vinciguerra1, Ashkan Eliasy2, Shervin Mir Mohi Sefat1, Daniel Kook1, Brendan Geraghty1, Cynthia J Roberts1, Ahmed Elsheikh1.   

Abstract

PURPOSE: To compare a biomechanically corrected intraocular pressure (bIOP) algorithm provided by the dynamic Scheimpflug analyzer (Corvis ST) with Goldmann applanation tonometry IOP (Goldmann IOP) and standard dynamic Scheimpflug analyzer IOP measurements before and after laser in situ keratomileusis (LASIK) and refractive lenticule extraction small-incision lenticule extraction (SMILE) surgeries.
SETTING: Smile Eye Clinic, Munich, Germany, and University of Liverpool, Liverpool, United Kingdom.
DESIGN: Retrospective case series.
METHODS: Patients scheduled for LASIK and patients scheduled for small-incision lenticule extraction for myopia or myopic astigmatism were included. The preoperative and postoperative evaluations included Goldmann, Scheimpflug tomography, and dynamic Scheimpflug analyzer IOP measurements.
RESULTS: The study comprised 14 patients in the LASIK group and 22 patients in the small-incision lenticule extraction group. Preoperative Goldmann IOP and Scheimpflug analyzer IOP values showed significant positive correlation with central corneal thickness (CCT) (P = .05 for LASIK; P = .003 for small-incision lenticule extraction). No significant correlation was found between bIOP and CCT (P > .05). After both surgeries, there were significant decreases in Goldmann IOP (-3.2 mm Hg ± 3.4 [SD] and -3.2 ± 2.1 mm Hg, respectively; both P < .001) and Scheimpflug analyzer IOP (-3.7 ± 2.1 mm Hg and -3.3 ± 2.0 mm Hg, respectively, both P < .001) compared with preoperative readings, whereas bIOP did not differ significantly (0.1 ± 2.1 mm Hg and 0.8 ± 1.8 mm Hg, respectively; P = .80 and P = .273, respectively).
CONCLUSIONS: The bIOP readings before and after LASIK and small-incision lenticule extraction were neither significantly different nor correlated with CCT. In contrast, both Goldmann IOP and Scheimpflug analyzer IOP had significant reductions postoperatively and showed significant correlation with CCT preoperatively.
Copyright © 2018 ASCRS and ESCRS. Published by Elsevier Inc. All rights reserved.

Entities:  

Mesh:

Year:  2018        PMID: 29685776     DOI: 10.1016/j.jcrs.2018.01.023

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  5 in total

1.  Comparison of Three Tonometers in Measuring Intraocular Pressure in Eyes That Underwent Myopic Laser in situ Keratomileusis and Photorefractive Keratectomy.

Authors:  Robert Edward T Ang; Neiman Vincent R Bargas; Gladness Henna A Martinez; George Michael N Sosuan; Maria Isabel Nabor-Umali
Journal:  Clin Ophthalmol       Date:  2022-05-27

2.  A new method for detecting the outer corneal contour in images from an ultra-fast Scheimpflug camera.

Authors:  Magdalena Jędzierowska; Robert Koprowski; Sławomir Wilczyński; Katarzyna Krysik
Journal:  Biomed Eng Online       Date:  2019-12-03       Impact factor: 2.819

3.  Biomechanically-Corrected Intraocular Pressure Compared To Pressure Measured With Commonly Used Tonometers In Normal Subjects.

Authors:  Mohammad-Reza Sedaghat; Hamed Momeni-Moghaddam; AbbasAli Yekta; Ahmed Elsheikh; Mehdi Khabazkhoob; Renato Ambrósio; Nasim Maddah; Zeynab Danesh
Journal:  Clin Optom (Auckl)       Date:  2019-10-17

4.  In Vivo Corneal Biomechanical Properties in a Selected Chinese Population, Measured Using the Corneal Visualization Scheimpflug Technology.

Authors:  Yuxin Li; Lei Tian; Li-Li Guo; Yiran Hao; Ying Jie
Journal:  Front Bioeng Biotechnol       Date:  2022-04-13

5.  Intraocular pressure changes and corneal biomechanics after hyperopic small-incision lenticule extraction.

Authors:  Dan Fu; Meiyan Li; Michael C Knorz; Shengsheng Wei; Jianmin Shang; Xingtao Zhou
Journal:  BMC Ophthalmol       Date:  2020-04-05       Impact factor: 2.209