Katharina Fischbach1, Otrud Kosiek1, Björn Friebe1, Christian Wybranski1, Bernhard Schnackenburg2, Alexander Schmeisser3, Jan Smid3, Jens Ricke1, Maciej Pech1.
Abstract
BACKGROUND: Cardiac magnetic resonance imaging (cMRI) has become the non-invasive reference standard for the evaluation of cardiac function and viability. The introduction of open, high-field, 1.0T (HFO) MR scanners offers advantages for examinations of obese, claustrophobic and paediatric patients.The aim of our study was to compare standard cMRI sequences from an HFO scanner and those from a cylindrical, 1.5T MR system. MATERIAL/Entities:
Keywords: B-SSFP; Black Blood Sequences; Cardiac MRI; Cylindrical 1.5T MR System; Open 1.0T MR System
Year: 2017 PMID: 29662579 PMCID: PMC5894020 DOI: 10.12659/PJR.899822
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
Figure 1A volunteer placed in the supine position in the 8-channel body coil for diagnostic cardiac MR imaging in an open 1.0T MR system.
Sequence parameters for the open 1.0T scanner and the 1.5T scanner.
| Parameter | B-SSFP (1.0/1.5T) | T1-wTSE (1.0/1.5T) | T1-w BB GRE (1.0/1.5T) | T2-w BB TSE (1.0/1.5T) |
|---|---|---|---|---|
| TR (ms) | 3.3/2.7 | 1000/1000 | 4.7/4.3 | 2000/2000 |
| TE (ms) | 1.64/1.35 | 15/15 | 2.3/2.1 | 80/80 |
| Flip angle á | 55/50 | 90/90 | 40/40 | 90/90 |
| FOV (mm) | 370/370 | 370/370 | 370/370 | 370/370 |
| Matrix | 384/384 | 384/384 | 256/256 | 512/512 |
| Slice thickness (mm) | 8/8 | 8/8 | 8/8 | 8/8 |
| Scan duration (sec) | 6/5 | 15/15 | 8/8 | 14/14 |
Per slice at a heart rate of approximately 65 bpm; TR – repetition time; TE – echo time; FOV – field of view.
SNR values of standard unenhanced cine and tissue characterization sequences both at an HFO and a cylindrical 1.5T MR system.
| SNR | Open 1.0T (HFO) MR system (mean ±SD, [range]) | Cylindrical 1.5T MR system (mean ±SD, [range]) |
|---|---|---|
| B-SSFP | 12±1.5 (9–15) | 32±10 (23–59) |
| T1-wTSE | 67±18 (25–88) | 205±32 (162–271) |
| T1-w BB GRE | 32±6 (24–43) | 79±16 (48–104) |
| T2-w BB TSE | 33±13 (17–65) | 85±23 (54–140) |
Overall image quality (1 = non-diagnostic; 4 excellent image quality) of standard unenhanced cine and tissue characterization sequences both at an HFO and a cylindrical 1.5T MR system.
| Overall image quality | Open 1.0T (HFO) MR system (mean ±SD, [range]) | Cylindrical 1.5T MR system (mean ±SD, [range]) |
|---|---|---|
| B-SSFP | 3.8±0.41 (3–4) | 3.6±0.51 (3–4) |
| T1-wTSE | 3.3±0.59 (2–4) | 3.6±0.51 (3–4) |
| T1-w BB GRE | 3.5±0.64 (2–4) | 3.3±0.49 (3–4) |
| T2-w BB TSE | 3.5±0.64 (2–4) | 3.4±0.51 (3–4) |
Visual assessment of artifacts (0 = none; 4 = non-diagnostic) of standard unenhanced cine and tissue characterization sequences both at an HFO and a cylindrical 1.5T MR system.
| Visual assessment of artifacts | Open 1.0T (HFO) MR system (mean ±SD, [range]) | Cylindrical 1.5T MR system (mean ±SD, [range]) |
|---|---|---|
| B-SSFP | 0.3±0.46 (0–1) | 0.8±1.01 (0–3) |
| T1-w TSE | 1.3±0.72 (0–3) | 1.2±1.01 (0–3) |
| T1-w BB GRE | 0.6±0.91 (0–3) | 1.1±1.06 (0–3) |
| T2-w BB TSE | 1.1±1.23 (0–3) | 1.2±1.26 (0–3) |
Figure 2Distribution of non-diagnostic segments in a bull’s eye plot according to the 16-segment model of the AHA. Numbers represent the total non-diagnostic segment count in the 1.0T/1.5T scanner for the T2-weighted, black blood, turbo spin echo (A: T2 BB) and for the T1-weighted, black blood, gradient echo (B: T1 BB) sequences. Most nondiagnostic segments occurred in the basal and lateral segments.
Figure 5Balanced steady-state free precession (A+E), T1-weighted turbo spin echo (B+F), T1-weighted black blood gradient echo (C+G) and T2-weighted black blood turbo spin echo (D+H) in the open 1.0T MR scanner (A–D) and in the cylindrical 1.5T MR system (E–H) in a healthy volunteer in whom a non-diagnostic, basal inferoseptal segment was imaged (arrow).