Cardiac cine parallel imaging on a 0.7T open system.

Masahiro Takizawa1, Tomohiro Goto, Hiroyuki Mochizuki, Masayuki Nonaka, Shizuka Nagai, Hiroyuki Takeuchi, Yo Taniguchi, Hisaaki Ochi, Tetsuhiko Takahashi.   

Abstract

BACKGROUND: Parallel imaging can be applied to cardiac imaging with a cylindrical MRI (magnetic resonance imaging) apparatus. Studies of open MRI, however, are few. This study sought to achieve cardiac cine parallel imaging (or RAPID, for "rapid acquisition through parallel imaging design") with an open 0.7T MRI apparatus.
MATERIALS AND METHODS: Imaging time was shortened in all slice directions with the use of a dedicated four-channel RF receiving coil comprising solenoid coils and butterfly coils. Coil shape was designed through an RF-coil simulation that considered biological load. The auto-calibration of a 0.7T open MRI apparatus incorporated a modified image-domain reconstruction algorithm. Cine images were obtained with a BASG, or balanced SARGE (steady-state acquisition with rewound gradient echo), sequence. Image quality was evaluated with cylindrical phantoms and five healthy volunteers.
RESULTS: Multi-slice phantom images showed no visible artifacts. Cine images taken under breath-hold with an acceleration factor of two were evaluated carefully. With auto-calibration, the images revealed no visible unfolded artifacts or motion artifacts. RAPID thus improved the acquisition speed, time resolution, and spatial resolution of short-axis, long-axis, and four-chamber images.
CONCLUSION: The use of a dedicated RF coil enabled cardiac cine RAPID to be performed with an open MRI apparatus.

Entities:  

Mesh:

Year:  2004        PMID: 16093619     DOI: 10.2463/mrms.3.45

Source DB:  PubMed          Journal:  Magn Reson Med Sci        ISSN: 1347-3182            Impact factor:   2.471


  2 in total

1.  Cardiac Magnetic Resonance Imaging Using an Open 1.0T MR Platform: A Comparative Study with a 1.5T Tunnel System.

Authors:  Katharina Fischbach; Otrud Kosiek; Björn Friebe; Christian Wybranski; Bernhard Schnackenburg; Alexander Schmeisser; Jan Smid; Jens Ricke; Maciej Pech
Journal:  Pol J Radiol       Date:  2017-09-15

Review 2.  MR Imaging in the 21st Century: Technical Innovation over the First Two Decades.

Authors:  Hiroyuki Kabasawa
Journal:  Magn Reson Med Sci       Date:  2021-04-16       Impact factor: 2.760