Ewa Fabiszewska1, Katarzyna Pasicz1, Iwona Grabska1, Witold Skrzyński1, Wioletta Ślusarczyk-Kacprzyk1, Wojciech Bulski1.
Abstract
BACKGROUND: Regular quality control is required in Poland only for those methods of medical imaging which involve the use of ionizing radiation but not for ultrasonography. It is known that the quality of ultrasound images may be affected by the wearing down or malfunctioning of equipment. MATERIAL/Entities:
Keywords: Phantoms, Imaging; Quality Control; Ultrasonography
Year: 2017 PMID: 29657644 PMCID: PMC5894034 DOI: 10.12659/PJR.904135
Source DB: PubMed Journal: Pol J Radiol ISSN: 1733-134X
List of evaluated ultrasound scanners and transducers.
| Manufacturer | Type | Year of manufacture |
|---|---|---|
| Siemens | Acuson S2000 | 2013 |
| ACUSON Aspen | 2002 | |
| Philips | HDI5000SonoCT | 2005 |
| IU22 | 2010 | |
| Sparq | 2013 | |
| IU22 | 2007 | |
| HD15 | 2014 | |
| HDI 4000 | 2003 | |
| HD 15 Pure Wave | 2014 | |
| Hitachi | HI Vision Preirus | 2014 |
| Prosound Alpha 7 | 2011 | |
| HI Vision Preirus | 2014 | |
| HI Vision Preirus | 2015 | |
| HI Vision 900 | 2007 | |
| HI Vision 900 | 2007 | |
| EUB 7500 | 2008 | |
| ATL | Ultra Mark 4 Plus | 1995 |
| HDI 3000 | 1997 | |
| GE Healthcare | Voluson 730 EXP | 2002 |
| OLYMPUS | EU-ME2 | 2015 |
| Supersonic Imagine | AIXPLORE | 2012 |
| BK Medical | Ultra View 800 Pro Focus | 2012 |
Figure 1Scheme of the phantom used for evaluation of ultrasound scanners (source: [20]). During measurements, the transducer is in contact with the top surface of the phantom.
Score scale for visibility of anechoic structures.
| Score | Shape | Edge | Interior |
|---|---|---|---|
| 1 | Round | Sharp | Black (anechoic) |
| 2 | Elliptical | Blurred | Single bright pixels |
| 3 | >20% difference between width and height | Hard to define | Filled with bright pixels |
Criteria for evaluation of test results.
| Test | Suggested action level | Suggested defect level |
|---|---|---|
| Depth of penetration | Change from baseline > 0,6 cm | Change from baseline >1.0 cm |
| Dead zone | 7 mm for f <3 MHz | 10 mm for f <3 MHz |
| Distance accuracy - vertical | >1.5 mm or 1.5% | >2 mm or 2% |
| Distance accuracy - horizontal | >2% or 2 mm | >3% or 3 mm |
| Axial resolution | Change from baseline | Change from baseline |
| Lateral resolution | Change from baseline | Change from baseline |
| Uniformity | Change from baseline | Change from baseline |
| Visibility of anechoic structures | Change from baseline | Change from baseline |
| Visibility of structures with various contrast (grayscale) | Change from baseline | Change from baseline |
Results of the depth of penetration test.
| Attenuation 0.5 dB/(cm·MHz) | Attenuation 0.7 dB/(cm·MHz) | ||||||
|---|---|---|---|---|---|---|---|
| Minimum [cm] | Maximum [cm] | Median [cm] | Minimum [cm] | Maximum [cm] | Median [cm] | ||
| Transducer type | Linear | 3.0 | 8.0 | 5.0 | 3.1 | 5.5 | 4.0 |
| Convex | 5.4 | 17.0 | 12.2 | 5.5 | 12.0 | 9.6 | |
| Sector | 12.3 | 16.0 | 13.0 | 9.0 | 13.0 | 10.6 | |
Figure 2Results of the depth of penetration test.
Figure 3Results of the dead zone test (zero is the best result).
Results of the test of distance accuracy in two directions.
| Maximum difference between nominal and measure distance [mm] | |||
|---|---|---|---|
| Attenuation 0.5 dB/(cm·MHz) | Attenuation 0.7 dB/(cm·MHz) | ||
| Horizontal direction | |||
| Transducer type | Linear | 1.7 | 1.1 |
| Convex | 2.6 | 2.0 | |
| Sector | 1.6 | 1.5 | |
| Vertical Direction | |||
| Transducer type | Linear | 1.2 | 1.7 |
| Convex | 2.0 | 2.0 | |
| Sector | 1.0 | 1.4 | |
Figure 4Difference between nominal and measured distance in the horizontal direction.
Figure 5Difference between nominal and measured distance in the vertical direction.
Resolution in the axial direction.
| Resolution in axial direction [mm] | |||||
|---|---|---|---|---|---|
| Attenuation 0.5 dB/(cm·MHz) | Attenuation 0.7 dB/(cm·MHz) | ||||
| Minimum | Maximum | Minimum | Maximum | ||
| Location of resolution pattern (depth): 3.0 cm | |||||
| Transducer type | Linear | 0.25 | 1.0 | 0.25 | 1.0 |
| Convex | 1.0 | 2.0 | 0.25 | 1.0 | |
| Sector | 0.5 | 4.0 | 0.5 | 2.0 | |
| Location of resolution pattern (depth): 6.5 cm | |||||
| Transducer type | Linear | 0.5 | 1.0 | 1.0 | x |
| Convex | 0.5 | 4.0 | 1.0 | 2.0 | |
| Sector | 0.5 | 4.0 | 1.0 | 4.0 | |
| Location of resolution pattern (depth): 10.5 cm | |||||
| Transducer type | Linear | x | x | x | x |
| Convex | 1.0 | 4.0 | 1.0 | 5.0 | |
| Sector | 2.0 | 4.0 | 2.0 | x | |
Resolution in the lateral direction.
| Resolution in lateral direction [mm] | |||||
|---|---|---|---|---|---|
| Attenuation 0.5 dB/(cm·MHz) | Attenuation 0.7 dB/(cm·MHz) | ||||
| Minimum | Maximum | Minimum | Maximum | ||
| Location of resolution pattern (depth): 3.0 cm | |||||
| Transducer type | Linear | 0.5 | 3.0 | 0.5 | 3.0 |
| Convex | 0.5 | 4.0 | 2.0 | 3.0 | |
| Sector | 1.0 | 4.0 | 2.0 | 3.0 | |
| Location of resolution pattern (depth): 6.5 cm | |||||
| Transducer type | Linear | 1.0 | 2.0 | 2.0 | x |
| Convex | 1.0 | 4.0 | 2.0 | 4.0 | |
| Sector | 2.0 | 4.0 | 3.0 | 4.0 | |
| Location of resolution pattern (depth): 10.5 cm | |||||
| Transducer type | Linear | x | x | x | x |
| Convex | 2.0 | 5.0 | 2.0 | 5.0 | |
| Sector | 4.0 | 5.0 | 4.0 | 5.0 | |
Figure 6Visibility of anechoic structures. A score of 3 points means no distortions, a higher score represents suboptimal image quality.
Figure 7Number of visible structures with various contrasts (largest result is best).