Literature DB >> 11423248

A review of two alternative ultrasound quality assurance programmes.

N J Dudley1, K Griffith, G Houldsworth, M Holloway, M A Dunn.   

Abstract

OBJECTIVE: This study reviewed the results of B-mode Quality Assurance (QA) performance tests on 17 real-time ultrasound scanners, performed over a period of 3 years, in order to assess their value. Following this review we revised and simplified our testing schedules to include two tests for noise and sensitivity. The value of the new schedules was assessed.
METHODS: Initially, testing schedules were similar to those recommended by two professional bodies. Results were reviewed to determine whether the tests predicted or confirmed faults. We then introduced a simplified testing programme using alternative measurements, attempting to demonstrate or predict noise related faults that affect the image, but were not demonstrated by current tests. These new tests have been performed on 24 ultrasound machines for up to 18 months.
RESULTS: A review of results has shown that measurements occasionally fall outside tolerance due to chance, and that faults that significantly affect the image, e.g. probe faults and noise, are reported by the users without predictive or concomitant changes in test results using our original schedules. Faults occur that do not immediately affect image quality and are not reported by the users. Inappropriate settings, e.g. monitors, are frequently reset at QA, particularly where there are potentially untrained users. The additional tests showed consistent changes in noise (four) or sensitivity (one) on five machines.
CONCLUSION: Our earlier tests were inadequate in demonstrating deterioration in the clinical performance of ultrasound imaging equipment. Introduction of a revised testing schedule has resulted in changes in equipment performance being detected and rectified.

Mesh:

Year:  2001        PMID: 11423248     DOI: 10.1016/s0929-8266(00)00119-1

Source DB:  PubMed          Journal:  Eur J Ultrasound        ISSN: 0929-8266


  11 in total

1.  The feasibility of implementation of ultrasound equipment standards set by United Kingdom professional bodies.

Authors:  Nicholas J Dudley; Debbie Harries; Jessica Wardle
Journal:  Ultrasound       Date:  2017-01-23

Review 2.  Quality assurance of ultrasound systems: current status and review of literature.

Authors:  H Grazhdani; E David; O Ventura Spagnolo; F Buemi; A Perri; N Orsogna; S Gigli; R Chimenz
Journal:  J Ultrasound       Date:  2018-06-14

3.  Early experience with automated B-mode quality assurance tests.

Authors:  N J Dudley; N M Gibson
Journal:  Ultrasound       Date:  2013-12-19

4.  The analysis of in-air reverberation patterns from medical ultrasound transducers.

Authors:  T Quinn; P K Verma
Journal:  Ultrasound       Date:  2014-02-06

5.  Automated performance assessment of ultrasound systems using a dynamic phantom.

Authors:  J Joy; F Riedel; A A Valente; S Cochran; G A Corner
Journal:  Ultrasound       Date:  2014-08-25

6.  The BMUS guidelines for regular quality assurance testing of ultrasound scanners.

Authors:  Nick Dudley; Stephen Russell; Barry Ward; Peter Hoskins
Journal:  Ultrasound       Date:  2014-02-06

7.  Critical US visibility with tissue harmonic imaging of subcutaneous nodules().

Authors:  S M Stella; B Ciampi; D Melchiorre; E Benedetti; E Orsitto; P V Lippolis
Journal:  J Ultrasound       Date:  2011-06-29

8.  Phantom-based quality assurance measurements in B-mode ultrasound.

Authors:  Vilma Mannila; Outi Sipilä
Journal:  Acta Radiol Short Rep       Date:  2013-12-11

9.  Method for automatic detection of defective ultrasound linear array transducers based on uniformity assessment of clinical images - A case study.

Authors:  Robert Lorentsson; Nasser Hosseini; Jan-Olof Johansson; Wiebke Rosenberg; Benny Stenborg; Lars Gunnar Månsson; Magnus Båth
Journal:  J Appl Clin Med Phys       Date:  2018-01-11       Impact factor: 2.102

10.  Intra- and interobserver reliability of gray scale/dynamic range evaluation of ultrasonography using a standardized phantom.

Authors:  Song Lee; Joon-Il Choi; Michael Yong Park; Dong Myung Yeo; Jae Young Byun; Seung Eun Jung; Sung Eun Rha; Soon Nam Oh; Young Joon Lee
Journal:  Ultrasonography       Date:  2014-02-26
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.