Enrico Di Girolamo1, Nanda Furia1, Massimiliano Faustino1, Marianna Appignani1, Giampiero Arcari2, Alessandro Angelini2, Loris Spadoni2.
Abstract
Subcutaneous implantable cardioverter defibrillator (S-ICD) is easy to implant, with poor risks of the patient. However, fat is a poor current conductor and increases defibrillation threshold. As shock impedance alone should not be considered a good efficacy predictor of an S-ICD system, an X-ray latero-lateral view for lead position should be achieved.Entities:
Keywords: shock efficacy; shock impedance; subcutaneous defibrillator
Year: 2018 PMID: 29636916 PMCID: PMC5889266 DOI: 10.1002/ccr3.1408
Source DB: PubMed Journal: Clin Case Rep ISSN: 2050-0904
Figure 1First and second failed induction test.
Figure 2Lead position before and after replacement in LL view.
Figure 3Third (final) induction test.