Abstract
BACKGROUND: Yttrium-90 (90Y) radioembolization involves the intra-arterial delivery of radioactive microspheres to treat hepatic malignancies. Though this therapy involves careful pre-treatment planning and imaging, little is known about the precise location of the microspheres once they are administered. Recently, there has been growing interest post-radioembolization imaging using positron-emission tomography (PET) for quantitative dosimetry and identifying lesions that may benefit from additional salvage therapy. In this study, we aim to measure the inter-center variability of 90Y PET measurements as measured on PET/MRI in preparation for a multi-institutional prospective phase I/II clinical trial. Eight institutions participated in this study and followed a standardized phantom filling and imaging protocol. The NEMA NU2-2012 body phantom was filled with 3 GBq of 90Y chloride solution. The phantom was imaged for 30 min in listmode on a Siemens Biograph mMR non-TOF PET/MRI scanner at five time points across 10 days (0.3-3.0 GBq). Raw PET data were sent to a central site for image reconstruction and data analysis. Images were reconstructed with optimal parameters determined from a previous study. Volumes of interest (VOIs) matching the known sphere diameters were drawn on the vendor-provided attenuation map and propagated to the PET images. Recovery coefficients (RCs) and coefficient of variation of the RCs (COV) were calculated from these VOIs for each sphere size and activity level.Entities:
Keywords: PET; PET/MRI; Radioembolization; Yttrium-90
Year: 2018 PMID: 29616365 PMCID: PMC5882483 DOI: 10.1186/s40658-018-0206-y
Source DB: PubMed Journal: EJNMMI Phys ISSN: 2197-7364
Dose calibrator and settings used at each site to measure 90Y activity and verify against that reported on shipping label
| Site number | Dose calibrator manufacturer | Model | Setting for 90Y |
|---|---|---|---|
| 1 | Capintec | CRC-12 | Channel 53 × 10 |
| 2 | Capintec | CRC-55 T PET | Channel 010 |
| 3 | Capintec | CRC-15R | Channel 48 × 10 |
| 4 | Capintec | CRC-15R | Channel 56 × 10 |
| 5 | Capintec | 55 t W | Channel 48 × 10 |
| 6 | Veenstra Instruments | VIK 2025051 | Factor 902, scale 100 |
| 7 | MED | ISOMED 2010 | Factor 2.3 (geometry: 3 ml syringe/1.4 ml solution) |
| 8 | Capintec | CRC-15R | Channel 48 × 10 |
Fig. 1Attenuation map of NEMA 2007/IEC 2008 body phantom with VOIs drawn around each of the six fillable spheres, 72 VOIs drawn in the background compartment, and a 28-mm diameter VOI drawn in the cold insert (center, yellow)
Quantities for volume of interest (VOI) analysis
| VOI and size | Measured quantity | Purpose |
|---|---|---|
| Field of view (FOV) | Total activity (GBq) | Accuracy in total injected activity measurement |
| Fillable spheres: 37, 28, 22, 17, 13, 10 mm | Recovery coefficient of mean activity concentration (MBq/ml) | Accuracy in VOI measurements from partial volume effects |
| Background spheres: 12 of 37-mm diameters | Recovery coefficient of mean activity concentration (MBq/ml) | Accuracy in warm background VOI measurements |
| Background spheres: 12 each of 37-, 28-, 22-, 17-, 13-, 10-mm diameters | Standard deviation of mean activity concentration (MBq/ml) | Variability in warm background measurements (noise) |
| Insert: 28-mm diameter, 160-mm length | Mean activity concentration (MBq/ml) | Background counts |
Fig. 2PET images from first day of imaging (~ 3.0 GBq) for each of the eight sites
Fig. 3a Total measured activity within the FOV compared to the true activity, averaged across the eight sites. The dashed line represents the identity (i.e., where measured equals true). b Error in measured activity within the FOV. Each point represents the mean measured activity across the eight sites. Error bars represent one standard deviation across the eight sites’ measurements. The shaded region represents ± 10% error. c Mean activity concentration of the 37 mm VOI within the warm background compartment of the phantom compared to the true activity concentration, averaged across the eight sites. The dashed line represents the identity (i.e., where measured equals true). d Deviation of measured activity concentration within warm background. Each point represents the mean measured activity concentration of the 37 mm VOI across the eight sites. Error bars represent one standard deviation across the eight sites’ measurements. The shaded region represents ± 10% error
Fig. 4a Variability in measurements between 12 × 8 background VOIs for each sphere size. b Misplaced counts in cold lung insert reported as percent of true background activity concentration. Each point represents the mean. Error bars represent one standard deviation
Fig. 5a–e Recovery coefficients as a function of hot sphere size from five different hot sphere activity concentrations across all sites. Each point represents the mean RC and error bars represent one standard deviation across the eight sites. The COV for each sphere size and hot sphere activity concentrations are shown in Table 3. Measurements from one site’s 18F measurements (54.4 MBq) are shown in black
Inter-center coefficient of variation (COV) of recovery coefficients (RCs) for each sphere size and activity concentration
| COV (%) (range) | |||||
|---|---|---|---|---|---|
| Sphere diameter (mm) | 2.16 MBq/ml | 1.00 MBq/ml | 0.60 MBq/ml | 0.36 MBq/ml | 0.16 MBq/ml |
| 10 | 36.2 (14.0–48.6) | 86.6 (6.18–99.8) | 75.1 (7.79–51.7) | 49.5 (9.83–45.1) | 64.1 (2.63–26.9) |
| 13 | 32.2 (23.5–59.8) | 55.1 (18.0–68.3) | 92.3 (12.5–112) | 44.9 (13.6–49.6) | 77.5 (3.47–65.5) |
| 17 | 29.7 (18.9–69.1) | 30.4 (30.3–78.0) | 36.4 (27.5–75.8) | 47.7 (20.5–75.0) | 78.8 (14.7–118) |
| 22 | 22.7 (41.2–84.1) | 22.8 (29.6–63.4) | 25.0 (32.6–70.0) | 23.7 (30.5–56.1) | 37.9 (10.4–49.4) |
| 28 | 24.1 (42.9–88.1) | 21.5 (41.7–80.3) | 17.8 (38.0–65.3) | 26.8 (26.8–69.6) | 15.8 (26.8–46.9) |
| 37 | 18.0 (56.0–95.0) | 22.2 (52.0–103) | 18.3 (54.9–92.2) | 25.3 (46.4–91.4) | 21.1 (45.2–81.3) |
Values in parentheses represent the range of RCs (min–max)
Fig. 6a–e Recovery coefficients as a function of hot sphere size from three consecutive scans at a single site. Each point represents the mean RC, and error bars represent one standard deviation. The COV for each sphere size and hot sphere activity concentrations are shown in Table 4. f Measurements from raw 18F data (black line), 18F data with similar 90Y true count statistics (light blue square), and 18F data with similar true count statistics and randoms fraction to 90Y. Again, each point represents the mean RC and error bars represent one standard deviation across the ten 3-s 18F datasets
Intra-center coefficient of variation (COV) of recovery coefficients (RCs) for each sphere size and activity concentration
| COV (%) (range) | |||||
|---|---|---|---|---|---|
| Sphere diameter (mm) | 2.32 MBq/ml | 1.06 MBq/ml | 0.65 MBq/ml | 0.38 MBq/ml | 0.17 MBq/ml |
| 10 | 4.49 (29.7–32.3) | 40.2 (20.8–40.2) | 59.4 (8.71–24.3) | 67.6 (14.4–56.6) | 108 (4.42–40.9) |
| 13 | 7.58 (44.1–51.0) | 47.1 (22.6–50.2) | 39.7 (27.2–63.3) | 32.9 (16.6–30.9) | 45.3 (29.1–79.6) |
| 17 | 12.4 (48.8–60.6) | 9.79 (57.5–68.1) | 13.4 (40.4–53.0) | 34.5 (21.6–43.5) | 49.6 (15.6–45.7) |
| 22 | 5.06 (55.6–61.5) | 8.35 (50.2–59.2) | 18.2 (43.4–62.0) | 36.8 (27.8–57.7) | 50.8 (14.5–47.1) |
| 28 | 6.48 (56.2–63.7) | 6.69 (49.9–57.0) | 4.89 (58.9–64.3) | 23.9 (34.9–56.9) | 35.0 (27.2–57.4) |
| 37 | 1.43 (77.0–79.0) | 4.79 (70.1–77.1) | 7.52 (66.6–77.0) | 16.0 (56.6–77.4) | 20.0 (53.9–77.7) |
Values in parentheses represent the range of RCs (min–max)
Intra-center coefficient of variation (COV) between three consecutive scans of recovery coefficients (RCs) of total activity and background activity concentration and of counts in the cold lung insert (C) on each imaging day
| COV (%) (range) | |||||
|---|---|---|---|---|---|
| VOI | 3.0 GBq | 1.4 GBq | 0.8 GBq | 0.5 GBq | 0.2 GBq |
| Total activity | 0.80 (96.7–98.2) | 0.47 (102–102) | 0.78 (106–107) | 0.79 (114–115) | 0.80 (139–141) |
| Background | 16.8 (46.5–147) | 26.9 (33.6–209) | 27.9 (32.8–179) | 34.6 (32.8–254) | 50.4 (30.5–423) |
| Lung insert | 4.51 (31.5–34.5) | 5.99 (39.6–44.6) | 2.82 (37.4–39.1) | 5.05 (39.9–44.5) | 10.2 (61.1–74.1) |
Numbers in parentheses represent the range of RCs for total activity and background activity concentration and C for the lung insert (min–max)