| Literature DB >> 29608197 |
Christine F Carson1, Timothy Jj Inglis1,2,3.
Abstract
This study investigated aerosolized viable bacteria in a university research laboratory during operation of an acoustic-assisted flow cytometer for antimicrobial susceptibility testing by sampling room air before, during and after flow cytometer use. The aim was to assess the risk associated with use of an acoustic-assisted flow cytometer analyzing unfixed bacterial suspensions. Air sampling in a nearby clinical laboratory was conducted during the same period to provide context for the existing background of microorganisms that would be detected in the air. The three species of bacteria undergoing analysis by flow cytometer in the research laboratory were Klebsiella pneumoniae, Burkholderia thailandensis and Streptococcus pneumoniae. None of these was detected from multiple 1000 L air samples acquired in the research laboratory environment. The main cultured bacteria in both locations were skin commensal and environmental bacteria, presumed to have been disturbed or dispersed in laboratory air by personnel movements during routine laboratory activities. The concentrations of bacteria detected in research laboratory air samples were reduced after interventional cleaning measures were introduced and were lower than those in the diagnostic clinical microbiology laboratory. We conclude that our flow cytometric analyses of unfixed suspensions of K. pneumoniae, B. thailandensis and S. pneumoniae do not pose a risk to cytometer operators or other personnel in the laboratory but caution against extrapolation of our results to other bacteria and/or different flow cytometric experimental procedures.Entities:
Keywords: bacteria; biosafety; flow cytometer; laboratory biocontainment; risk assessment
Year: 2018 PMID: 29608197 PMCID: PMC5873458 DOI: 10.12688/gatesopenres.12759.2
Source DB: PubMed Journal: Gates Open Res ISSN: 2572-4754
Identities of the commonest bacterial species isolated on blood agar † from 1000 L air samples, arranged by laboratory location.
†sampling onto HBA frequently yielded plates crowded with microbial growth. Only the six commonest colony types were identified plus any that resembled the three target bacteria ‡results from both sampling sites in the clinical laboratory were pooled for this table
| Bacteria identified from blood agar | ||||
|---|---|---|---|---|
| Laboratory
| Sampling site | Gram positive
| Gram negative
| Gram positive
|
| Research laboratory | Sample introduction port
|
|
|
|
| Preparation bench |
|
|
| |
| Clinical
| Open benches
[ |
| (not present among
| (not present
|
Incidence of isolation of all Gram negative bacterial species sampled onto MacConkey agar and identified in 1000 L air samples according to laboratory location.
‡results from both sampling sites in the clinical laboratory were pooled for this table
| Laboratory location and sampling site | |||
|---|---|---|---|
| Research laboratory | Clinical laboratory | ||
| Gram negative bacteria
| Acoustic-assisted
| Preparation
| Open benches
[ |
|
| 0 | 1 | 3 |
|
| 4 | 3 | 0 |
| Unidentifiable | 1 | 0 | 3 |
| No bacterial growth | 17 | 14 | 13 |
Figure 1. Airborne bacterial count detected on blood agar increases in the research laboratory over the course of the day, both beside the acoustic-assisted flow cytometer (AFC) and at the preparation bench (PB) (respective regression lines shown).
Bacterial counts (CFU/ 1000 L air) recovered from air sampled onto blood agar in the Research Laboratory or the Clinical Laboratory.
†‘In use’ refers to cytometer in operation for bacterial analysis. ‡results from both sampling sites in the clinical laboratory were pooled for this table
| Laboratory
| Sampling site | Number
| Number of CFU/1000 L air on
| ||
|---|---|---|---|---|---|
| Lower
| Median | Upper
| |||
| Research
| Acoustic flow cytometer,
| 9 | 0 | 5 | 85 |
| Acoustic flow cytometer,
| 10 | 10 | 54 | 99 | |
| Acoustic flow cytometer,
| 4 | 1 | 15.5 | 21 | |
| Preparation bench | 20 | 0 | 15.5 | 83 | |
| Clinical
| Open benches
[ | 14 | 0 | 33.5 | 189 |
| All locations | All sites | 57 | 0 | 21 | 189 |
Figure 2. Airborne bacterial count close to acoustic flow cytometer sample introduction port during use, before (pre clean) and after research laboratory cleaning (post clean), compared with samples from the clinical microbiology laboratory (CML) where no interventional cleaning was undertaken.
Pre clean research laboratory and post-clean median counts in close proximity to flow cytometer sample introduction port = 54 and 15.5 CFU/ 1000 L, U = 2.5, **p = 0.0090.