Literature DB >> 29596906

Atrial electrogram quality in single-pass defibrillator leads with floating atrial bipole in patients with permanent atrial fibrillation and cardiac resynchronization therapy.

Christian Sticherling1, Dirk Müller2, Beat A Schaer3, Silke Krüger4, Christof Kolb5.   

Abstract

Many patients receiving cardiac resynchronization therapy (CRT) suffer from permanent atrial fibrillation (AF). Knowledge of the atrial rhythm is important to direct pharmacological or interventional treatment as well as maintaining AV-synchronous biventricular pacing if sinus rhythm can be restored. A single pass single-coil defibrillator lead with a floating atrial bipole has been shown to obtain reliable information about the atrial rhythm but has never been employed in a CRT-system. The purpose of this study was to assess the feasibility of implanting a single coil right ventricular ICD lead with a floating atrial bipole and the signal quality of atrial electrograms (AEGM) in CRT-defibrillator recipients with permanent AF. METHODS AND
RESULTS: Seventeen patients (16 males, mean age 73 ± 6 years, mean EF 25 ± 5%) with permanent AF and an indication for CRT-defibrillator placement were implanted with a designated CRT-D system comprising a single pass defibrillator lead with a atrial floating bipole. They were followed-up for 103 ± 22 days using remote monitoring for AEGM transmission. All patients had at last one AEGM suitable for atrial rhythm diagnosis and of 100 AEGM 99% were suitable for visual atrial rhythm assessment. Four patients were discharged in sinus rhythm and one reverted to AF during follow-up.
CONCLUSION: Atrial electrograms retrieved from a single-pass defibrillator lead with a floating atrial bipole can be reliably used for atrial rhythm diagnosis in CRT recipients with permanent AF. Hence, a single pass ventricular defibrillator lead with a floating bipole can be considered in this population.
Copyright © 2018 Indian Heart Rhythm Society. Production and hosting by Elsevier B.V. All rights reserved.

Entities:  

Keywords:  Atrial fibrillation; Atrial sensing; Cardiac resynchronization therapy; Defibrillator; Heart failure; Remote monitoring

Year:  2018        PMID: 29596906      PMCID: PMC6090003          DOI: 10.1016/j.ipej.2018.03.005

Source DB:  PubMed          Journal:  Indian Pacing Electrophysiol J        ISSN: 0972-6292


Background

Cardiac resynchronization therapy (CRT) has been shown to reduce morbidity and mortality in symptomatic heart failure patients on optimal medical therapy with left bundle branch block and a left ventricular ejection fraction ≤35% [1,2]. The bulk of evidence has been generated in patients in sinus rhythm. Up to one third of heart failure patients are in persistent or permanent atrial fibrillation (AF). There is evidence that these patients do benefit from CRT if intrinsic atrioventricular (AV) nodal conduction is completely blocked by AV-node ablation [3,4] and therefore most CRT candidates with permanent AF receive CRT today [5,6]. Since atrial stimulation is not possible in permanent AF, many operators only implant a right and a left ventricular lead, thereby evading the risk of atrial lead placement like dislodgements. Knowledge of the atrial rhythm may be nonetheless important, since some patient will revert to sinus rhythm under CRT [7]. Any sort of atrial sensing then allows for AV-synchronous biventricular pacing. Furthermore, with regular remote monitoring transmission of atrial intracardiac electrograms (AEGM), an early appreciation of sinus rhythm may be facilitated [8]. Recently, a single-pass single-coil defibrillator lead with a floating atrial bipole has been shown to obtain reliable information of the atrial rhythm in patients receiving a single chamber ICD without CRT [9]. The purpose of this study was to assess the signal quality of AEGM in CRT-D recipients with permanent AF implanted with a single coil right ventricular ICD lead with a floating atrial bipole. AEGMs were analysed both as received from the programming device and as transmitted by remote monitoring.

Methods

Study population

Between August 2013 and June 2014 17 patients with permanent atrial fibrillation, heart failure NYHA class III/IV, QRS width ≥120 ms and a mean ventricular rate ≤60 bpm at rest or ≤90 bpm at exercise or pacemaker dependency as a result of planned AV junction ablation [5,10], and left ventricular ejection fraction ≤ 35% were implanted with a CRT-D in 3 sites in Germany and Switzerland. Local ethics review committees at each center approved the study and informed consent was obtained from all participants. The trial was funded by Biotronik SE (Berlin/Germany) and registered at ClinicalTrial.gov number, NCT01930605.

Device specification

Patients received a designated CRT-D device (Biotronik Lumax 640/740 DX, Biotronik SE, Berlin, Germany) with a specific 4-fold preamplifier for the atrial sensing signal and no atrial pacing capability. This device was connected to a single-coil right ventricular defibrillation lead with two floating atrial ring electrodes 15 or 17 cm proximal to the RV tip (Biotronik Linox Smart S DX, Biotronik SE, Berlin, Germany). The devices were equipped with remote monitoring (Biotronik Home Monitoring©) with the ability to transmit one three-channel IEGM per day either scheduled by programming or if a spontaneous episode of arrhythmia occurred. The devices record the “atrial burden” (AB) as the percentage of the time in which the atrial rate is above 200 beats per minute, and transmit the result of the last 24 h by Home Monitoring. AB figures were imported from the HM service center and averaged for the periods from implantation to the 1 month follow-up, and from the 1 month to the 3 months follow-ups.

Implantation and follow-up

Implantations were performed according to institutional standards. All patients received the Linox Smart S DX right ventricular lead and a transvenous left-ventricular lead, but no separate atrial lead. Leads were connected to a designated CRT defibrillator with optimized filter settings for atrial signal processing. The devices were programmed to the VVI or VVIR (for patients with bradycardic AF) pacing mode but atrial IEGM acquisition and transmission was activated in all patients. Study participants were followed-up for three months. AEGMs were retrieved by device interrogation at implantation, discharge, after 1 month and after 3 months. Furthermore, periodic transmission of AEGMs by Home Monitoring was scheduled on days 3 and 20 after discharge from hospital, and every 20 days after the one-month follow-up. AEGMs were also analysed when an event triggered transmission was available (theoretically numerous, but the maximal real number was 3 during the three-month study period). Biotronik Home Monitoring© does allow for the transmission of periodic AEGMs only when the patient is near the transmission station at the time of EGM recording, as no memory possibility is available in the system. This explains why these periodic EGMs are not available in all the patients all of the time points. For all AEGMs, whether retrieved by programmer or transmitted by Home Monitoring, two investigators independently decided whether or not it was possible to judge the atrial rhythm, and which atrial rhythm was present. These investigators were blinded to other potentially available ECG or electrogram tracings of the respective patients.

Endpoints

The first endpoint was the percentage of patients with at least one AEGM suitable for atrial diagnosis within the study period of three months. The second endpoint was the proportion of AEGM suitable for atrial diagnosis and the third the number of patients found in sinus rhythm based on the AEGM. During follow-up visits, the AEGM diagnosis was confirmed by 12-lead ECGs.

Results

The study population consisted of 17 patients, predominantly male (94%), with a mean age of 73 ± 6 years, and a mean EF of 25 ± 5%). They were followed for 103 ± 22 days. The mean duration of AF before implantation was 43 ± 40 months and 5 patients underwent scheduled AV-node ablation (Table 1). During the follow-up no antitachycardia therapies for ventricular tachyarrhythmias were delivered by the ICD.
Table 1

Patient demographics.

Patient characteristic
Male16 (94%)
Age (years)73 ± 6
LVEF (%)25 ± 5
AF duration (months)43 ± 40
Mean heart rate (bpm)79 ± 23
QRS duration (ms)160 ± 27
Left bundle branch block (n)13 (76%)
NYHA class II/III/IV0/16/1
AV-node ablation5 (29%)
Ischemic cardiomyopathy7 (41%)
Primary preventive indication17 (100%)
Hypertension16 (94%)
Oral anticoagulation15 (88%)
Amiodarone4 (24%)
Beta-Blocker14 (82%)
Patient demographics.

Atrial EGM

In the 17 remaining patients, 100 AEGM (2–11 per patient) were analysed. All patients had at least one AEGM that was suitable for atrial diagnosis. 99 out of 100 (99%) AEGM allowed for rhythm diagnosis (Fig. 1, Fig. 2).
Fig. 1

Examples of three channel EGMs transmitted by Home Monitoring in a patient at 63 days (A) and 83 days (B) after implantation. A: correct atrial sensing of sinus rhythm and AV sequential biventricular pacing, B: The same patient 20 days later, in atrial fibrillation with loss of biventricular pacing and the need for intervention.

Fig. 2

Examples of three channel EGMs transmitted by Home Monitoring with AF and correct atrial marker annotation (A) and without correct atrial marker annotation due to undersensing (B).

Examples of three channel EGMs transmitted by Home Monitoring in a patient at 63 days (A) and 83 days (B) after implantation. A: correct atrial sensing of sinus rhythm and AV sequential biventricular pacing, B: The same patient 20 days later, in atrial fibrillation with loss of biventricular pacing and the need for intervention. Examples of three channel EGMs transmitted by Home Monitoring with AF and correct atrial marker annotation (A) and without correct atrial marker annotation due to undersensing (B). Four patients (patients1/3/4/15, 23%) were discharged in sinus rhythm after cardioversion during defibrillation threshold testing. All of them had sinus rhythm confirmed in a first scheduled Home Monitoring transmission 3 days after discharge. After one month, 3 of these patients were still in sinus rhythm and another converted spontaneously to sinus rhythm. The rhythm diagnosis of the AEGM was congruent with the rhythm found during FU using a 12 lead ECG in all cases. However, signal amplitudes in some patients were too low and variable to be used for automated algorithms. This explains why the burden is not 100% in certain patients who were always in AF on every single documented ECG or AEGM.

Discussion

The main finding of this study is that a single-pass defibrillator lead with a floating atrial bipole can be safely implanted and used for visual diagnosis of the atrial rhythm in patients receiving a CRT defibrillator system. AEGM quality allowed for adequate rhythm diagnosis in 99% of episodes. The study for the first time collected data on atrial sensing capabilities of single-pass defibrillator leads connected to a dedicated CRT defibrillator. Previously, it has been shown that atrial rhythm detection via a floating atrial bipole can be safely employed in patients receiving a VDD defibrillator [9,11]. Furthermore, it was demonstrated that the sensitivity and specificity for detection of ventricular tachyarrhythmias and discrimination of supraventricular tachyarrhythmias were not different from a conventional dual chamber ICD [9]. The present study investigated its performance in a CRT heart failure population with pre-existing permanent AF. In our study, the AEGM retrieved from the atrium were used for visual diagnosis of the atrial rhythm only. It is important to point out that the atrial EGM could not be reliably used for correct annotation in the device's marker channel in the studied cohort of patients with advanced atrial remodelling due to long-lasting AF. This may be due to the fact that changing activation wave fronts during atrial fibrillation together with the slightly different location of the floating atrial bipole during cardiac contraction and different filling states cause greater differences in the atrial signal and accounts for more over- or undersensing. Whether the signals from the floating atrial bipole can be used for automated diagnosis or for detection enhancement algorithms with modified signal processing capabilities, needs to be further studied. Currently, the atrial channel should not be used for automated detection enhancement algorithms. Many patients who require CRT are in long-standing AF and a significant proportion later requires ablation of the atrioventricular node in order to achieve sufficient biventricular pacing [12]. Many operators will not implant an atrial lead since there is a risk of dislodgement and atrial pacing or AV-synchronization is impossible in this population. Our study shows that the use of the studied lead allows for atrial rhythm diagnosis without an additional atrial lead. In fact, after a one month follow-up 4 of the patients with perceived permanent AF were in sinus rhythm and AV-synchronous pacing based on the atrial signal in sinus rhythm was possible. In these patients, remote monitoring may be particularly helpful as recurrence of atrial fibrillation may prevent effective biventricular stimulation and mandate antiarrhythmic therapy or AV-node ablation [8]. It is important to point out that the system used in this study contains an amplifier for the atrial channel to compensate for the smaller signal amplitudes of floating ring leads. We did not test whether atrial sensing would be possible without the amplifier, but experience from VDD pacemakers suggests that a more sensitive input stage is essential. AV sequential pacing with floating ring VDD-electrodes has become less popular in the field of bradycardia pacing since atrial pacing is not possible in case of sinus node disease. However, it has been shown, that in a population with different degrees of AV-block less than 5% need to receive an additional atrial lead after a mean follow-up of 3.8 years [13]. Many operators will not implant any atrial lead in patients receiving a CRT-D system who are in permanent atrial fibrillation. Since the single-pass “DX” lead is not thicker than a comparable standard lead and single coil defibrillator leads are the current gold-standard, there is no obvious medical disadvantage compared to a conventional system without atrial lead. On the other hand it gives the few patients who may revert back to sinus rhythm the benefit of AV-sequential biventricular pacing. Atrial pacing will not be possible with the investigated system. However, atrial pacing may be disadvantageous and is suggested to be avoided in the CRT population since it increases intraatrial and consequently AV dyssynchrony [14].

Limitations

One limitation of our study is the small sample size of only 17 patients. In terms of the actual implantation procedure there were no signs of specific problems with the novel system. Furthermore, all atrial EGM where analysed visually and were not used for automated annotation. This carries a certain risk of subjective interpretation of the electrograms.

Conclusion

Atrial electrograms retrieved from a single-pass defibrillator lead combined with a dedicated CRT defibrillator can be reliably used for atrial rhythm diagnosis, e.g. after transmission through a remote monitoring system. The system can be used for monitoring the atrial rhythm in CRT recipients with long-standing atrial fibrillation without the need for an atrial lead.

Funding

The study was funded by Biotronik, SE, Berlin, Germany.

Conflicts of interest

Dr. Sticherling has received lecture honoraria/travel support from Biotronik, Boston Scientific, Medtronic, LivaNova, and St. Jude Medical; has been a consultant to Biotronik, Boston Scientific, and Medtronic; and has performed clinical studies supported by Biotronik, Medtronic, Boston Scientific, and St. Jude Medical. Dr. Schaer has received lecture honoraria/travel support Medtronic, and LivaNova, and has performed clinical studies supported by Biotronik, Medtronic, Boston Scientific, and St. Jude Medical. Dr. Kolb has received lecture honoraria/travel support from Biotronik, Boston Scientific, Medtronic, Sorin Group, Spectranetics, and St. Jude Medical; has been a consultant to Biotronik, Boston Scientific, and Sorin Group; and has performed clinical studies supported by Biotronik, Medtronic, Boston Scientific, Sorin Group, and St. Jude Medical.
  14 in total

1.  Cardiac-resynchronization therapy for mild-to-moderate heart failure.

Authors:  Anthony S L Tang; George A Wells; Mario Talajic; Malcolm O Arnold; Robert Sheldon; Stuart Connolly; Stefan H Hohnloser; Graham Nichol; David H Birnie; John L Sapp; Raymond Yee; Jeffrey S Healey; Jean L Rouleau
Journal:  N Engl J Med       Date:  2010-11-14       Impact factor: 91.245

2.  Right atrial pacing impairs cardiac function during resynchronization therapy: acute effects of DDD pacing compared to VDD pacing.

Authors:  Alain Bernheim; Peter Ammann; Christian Sticherling; Peter Burger; Beat Schaer; Hans Peter Brunner-La Rocca; Jens Eckstein; Stephanie Kiencke; Christoph Kaiser; Andre Linka; Peter Buser; Matthias Pfisterer; Stefan Osswald
Journal:  J Am Coll Cardiol       Date:  2005-05-03       Impact factor: 24.094

3.  2013 ESC Guidelines on cardiac pacing and cardiac resynchronization therapy: the Task Force on cardiac pacing and resynchronization therapy of the European Society of Cardiology (ESC). Developed in collaboration with the European Heart Rhythm Association (EHRA).

Authors:  Michele Brignole; Angelo Auricchio; Gonzalo Baron-Esquivias; Pierre Bordachar; Giuseppe Boriani; Ole-A Breithardt; John Cleland; Jean-Claude Deharo; Victoria Delgado; Perry M Elliott; Bulent Gorenek; Carsten W Israel; Christophe Leclercq; Cecilia Linde; Lluís Mont; Luigi Padeletti; Richard Sutton; Panos E Vardas; Jose Luis Zamorano; Stephan Achenbach; Helmut Baumgartner; Jeroen J Bax; Héctor Bueno; Veronica Dean; Christi Deaton; Cetin Erol; Robert Fagard; Roberto Ferrari; David Hasdai; Arno W Hoes; Paulus Kirchhof; Juhani Knuuti; Philippe Kolh; Patrizio Lancellotti; Ales Linhart; Petros Nihoyannopoulos; Massimo F Piepoli; Piotr Ponikowski; Per Anton Sirnes; Juan Luis Tamargo; Michal Tendera; Adam Torbicki; William Wijns; Stephan Windecker; Paulus Kirchhof; Carina Blomstrom-Lundqvist; Luigi P Badano; Farid Aliyev; Dietmar Bänsch; Helmut Baumgartner; Walid Bsata; Peter Buser; Philippe Charron; Jean-Claude Daubert; Dan Dobreanu; Svein Faerestrand; David Hasdai; Arno W Hoes; Jean-Yves Le Heuzey; Hercules Mavrakis; Theresa McDonagh; Jose Luis Merino; Mostapha M Nawar; Jens Cosedis Nielsen; Burkert Pieske; Lidija Poposka; Frank Ruschitzka; Michal Tendera; Isabelle C Van Gelder; Carol M Wilson
Journal:  Eur Heart J       Date:  2013-06-24       Impact factor: 29.983

4.  2012 EHRA/HRS expert consensus statement on cardiac resynchronization therapy in heart failure: implant and follow-up recommendations and management.

Authors:  Jean-Claude Daubert; Leslie Saxon; Philip B Adamson; Angelo Auricchio; Ronald D Berger; John F Beshai; Ole Breithard; Michele Brignole; John Cleland; David B Delurgio; Kenneth Dickstein; Derek V Exner; Michael Gold; Richard A Grimm; David L Hayes; Carsten Israel; Christophe Leclercq; Cecilia Linde; Joann Lindenfeld; Bela Merkely; Lluis Mont; Francis Murgatroyd; Frits Prinzen; Samir F Saba; Jerold S Shinbane; Jagmeet Singh; Anthony S Tang; Panos E Vardas; Bruce L Wilkoff; Jose Luis Zamorano
Journal:  Heart Rhythm       Date:  2012-09       Impact factor: 6.343

Review 5.  Is cardiac resynchronization therapy an antiarrhythmic therapy for atrial fibrillation? A systematic review and meta-analysis.

Authors:  Paul L Hess; Kevin P Jackson; Vic Hasselblad; Sana M Al-Khatib
Journal:  Curr Cardiol Rep       Date:  2013-02       Impact factor: 2.931

6.  The effect of cardiac resynchronization on morbidity and mortality in heart failure.

Authors:  John G F Cleland; Jean-Claude Daubert; Erland Erdmann; Nick Freemantle; Daniel Gras; Lukas Kappenberger; Luigi Tavazzi
Journal:  N Engl J Med       Date:  2005-03-07       Impact factor: 91.245

7.  Biventricular pacing for atrioventricular block and systolic dysfunction.

Authors:  Anne B Curtis; Seth J Worley; Philip B Adamson; Eugene S Chung; Imran Niazi; Lou Sherfesee; Timothy Shinn; Martin St John Sutton
Journal:  N Engl J Med       Date:  2013-04-25       Impact factor: 91.245

8.  Cardiac resynchronization therapy in patients with atrial fibrillation: the CERTIFY study (Cardiac Resynchronization Therapy in Atrial Fibrillation Patients Multinational Registry).

Authors:  Maurizio Gasparini; Christophe Leclercq; Maurizio Lunati; Maurizio Landolina; Angelo Auricchio; Massimo Santini; Giuseppe Boriani; Barbara Lamp; Alessandro Proclemer; Antonio Curnis; Catherine Klersy; Francisco Leyva
Journal:  JACC Heart Fail       Date:  2013-10-23       Impact factor: 12.035

9.  2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC)Developed with the special contribution of the Heart Failure Association (HFA) of the ESC.

Authors:  Piotr Ponikowski; Adriaan A Voors; Stefan D Anker; Héctor Bueno; John G F Cleland; Andrew J S Coats; Volkmar Falk; José Ramón González-Juanatey; Veli-Pekka Harjola; Ewa A Jankowska; Mariell Jessup; Cecilia Linde; Petros Nihoyannopoulos; John T Parissis; Burkert Pieske; Jillian P Riley; Giuseppe M C Rosano; Luis M Ruilope; Frank Ruschitzka; Frans H Rutten; Peter van der Meer
Journal:  Eur Heart J       Date:  2016-05-20       Impact factor: 29.983

10.  Long-term survival in patients undergoing cardiac resynchronization therapy: the importance of performing atrio-ventricular junction ablation in patients with permanent atrial fibrillation.

Authors:  Maurizio Gasparini; Angelo Auricchio; Marco Metra; François Regoli; Cecilia Fantoni; Barbara Lamp; Antonio Curnis; Juergen Vogt; Catherine Klersy
Journal:  Eur Heart J       Date:  2008-04-04       Impact factor: 29.983

View more
  1 in total

1.  Novel two-lead cardiac resynchronization therapy system provides equivalent CRT responses with less complications than a conventional three-lead system: Results from the QP ExCELs lead registry.

Authors:  Naushad A Shaik; Michael Drucker; Christopher Pierce; Gabor Z Duray; Shane Gillett; Crystal Miller; Camden Harrell; George Thomas
Journal:  J Cardiovasc Electrophysiol       Date:  2020-06-01
  1 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.