Christian Sticherling1, Dirk Müller2, Beat A Schaer3, Silke Krüger4, Christof Kolb5.
Abstract
Many patients receiving cardiac resynchronization therapy (CRT) suffer from permanent atrial fibrillation (AF). Knowledge of the atrial rhythm is important to direct pharmacological or interventional treatment as well as maintaining AV-synchronous biventricular pacing if sinus rhythm can be restored. A single pass single-coil defibrillator lead with a floating atrial bipole has been shown to obtain reliable information about the atrial rhythm but has never been employed in a CRT-system. The purpose of this study was to assess the feasibility of implanting a single coil right ventricular ICD lead with a floating atrial bipole and the signal quality of atrial electrograms (AEGM) in CRT-defibrillator recipients with permanent AF. METHODS ANDEntities:
Keywords: Atrial fibrillation; Atrial sensing; Cardiac resynchronization therapy; Defibrillator; Heart failure; Remote monitoring
Year: 2018 PMID: 29596906 PMCID: PMC6090003 DOI: 10.1016/j.ipej.2018.03.005
Source DB: PubMed Journal: Indian Pacing Electrophysiol J ISSN: 0972-6292
Patient demographics.
| Patient characteristic | |
|---|---|
| Male | 16 (94%) |
| Age (years) | 73 ± 6 |
| LVEF (%) | 25 ± 5 |
| AF duration (months) | 43 ± 40 |
| Mean heart rate (bpm) | 79 ± 23 |
| QRS duration (ms) | 160 ± 27 |
| Left bundle branch block (n) | 13 (76%) |
| NYHA class II/III/IV | 0/16/1 |
| AV-node ablation | 5 (29%) |
| Ischemic cardiomyopathy | 7 (41%) |
| Primary preventive indication | 17 (100%) |
| Hypertension | 16 (94%) |
| Oral anticoagulation | 15 (88%) |
| Amiodarone | 4 (24%) |
| Beta-Blocker | 14 (82%) |
Fig. 1Examples of three channel EGMs transmitted by Home Monitoring in a patient at 63 days (A) and 83 days (B) after implantation. A: correct atrial sensing of sinus rhythm and AV sequential biventricular pacing, B: The same patient 20 days later, in atrial fibrillation with loss of biventricular pacing and the need for intervention.
Fig. 2Examples of three channel EGMs transmitted by Home Monitoring with AF and correct atrial marker annotation (A) and without correct atrial marker annotation due to undersensing (B).